Feasibility Study of a Novel Thoraco-abdominal Aortic Hybrid Device (SPIDER-graft) in a Translational Pig Model

被引:9
作者
Debus, Eike S. [1 ]
Koelbel, Tilo [1 ]
Dupree, Anna [2 ]
Daum, Guenter [1 ]
Sandhu, Harleen L. [3 ]
Manzoni, Daniel [1 ]
Wipper, Sabine H. [1 ]
机构
[1] Univ Hosp Eppendorf, Univ Heart Ctr, Dept Vasc Med, Hamburg, Germany
[2] Univ Hosp Hamburg Eppendorf, Dept Gen & Visceral Surg, Hamburg, Germany
[3] McGovern Med Sch UTHlth, Dept Cardiothorac & Vasc Surg, Houston, TX USA
关键词
SPIDER-graft; TAAA; Experimental study; Hybrid repair; Stentgraft; Thoracoabdominal aorta; ANEURYSM REPAIR; OUTCOMES; DISSECTION; PERFUSION;
D O I
10.1016/j.ejvs.2017.11.018
中图分类号
R61 [外科手术学];
学科分类号
摘要
Background: The hybrid SPIDER-graft consists of a proximal descending aortic stent graft and a conventional six branched Dacron graft for open abdominal aortic repair. Technical feasibility with regard to avoiding thoracotomy and extracorporeal circulation (ECC) during thoraco-abdominal aortic hybrid repair and peri-procedural safety of this novel device are unknown. Material and methods: This was a feasibility and safety study in domestic pigs (75-85 kg). The abdominal aorta including iliac bifurcation, left renal artery, and visceral arteries were exposed via retroperitoneal access. The right iliac branch was first temporarily anastomosed end to side to the distal aorta via partial clamping. During inflow reduction and infra-coeliac cross-clamping, the coeliac trunk (CT) was divided and the proximal stent graft portion of the SPIDER-graft was deployed into the descending aorta via the CT ostium. Retrograde visceral and antegrade aorto-iliac blood flow was maintained via the iliac side branch. The visceral, renal, and iliac arteries were sequentially anastomosed, finally replacing the first iliac end to side anastomosis. Technical success, blood flow, periods of ischaemia, and peri-procedural complications were evaluated after intra-operative completion angiography and post-operative computed tomography angiography. Results: Six animals underwent successful thoracic stent graft deployment and distal open reconstruction without peri-operative death. The median thoracic graft implantation time was 4.5 min, and the median ischaemia times before reperfusion were 10 min for the CT, 8 min for the superior mesenteric artery, 13 min for the right renal artery, and 22 min for the left renal artery. Angiography demonstrated appropriate graft implantation and blood flow measurements confirmed sufficient blood flow through all side branches. Conclusion: In this translational pig model, thoraco-abdominal hybrid repair using the novel SPIDER-graft was successful in avoiding thoracotomy and ECC. Technical feasibility and safety appear promising, but need to be reassessed in humans. (C) 2017 Published by Elsevier Ltd on behalf of European Society for Vascular Surgery.
引用
收藏
页码:196 / 205
页数:10
相关论文
共 18 条
[1]   Open repair of intact thoracoabdominal aortic aneurysms in the American College of Surgeons National Surgical Quality Improvement Program [J].
Bensley, Rodney P. ;
Curran, Thomas ;
Hurks, Rob ;
Lo, Ruby C. ;
Wyers, Mark C. ;
Hamdan, Allen D. ;
Chaikof, Elliot L. ;
Schermerhorn, Marc L. .
JOURNAL OF VASCULAR SURGERY, 2013, 58 (04) :894-900
[2]   Clinical outcomes of single versus staged hybrid repair for thoracoabdominal aortic aneurysm [J].
Canaud, Ludovic ;
Karthikesalingam, Alan ;
Jackson, Dan ;
Cresswell, Lynne ;
Cliff, Michael ;
Markar, S. Sheraz ;
Maytham, Gary ;
Black, Steven ;
Thompson, Matt .
JOURNAL OF VASCULAR SURGERY, 2013, 58 (05) :1192-1200
[3]   Hybrid approach to thoracoabdominal aortic aneurysms in patients with prior aortic surgery [J].
Chiesa, Roberto ;
Tshomba, Yamume ;
Melissano, Germano ;
Marone, Enrico M. ;
Bertoglio, Luca ;
Setacci, Francesco ;
Calliari, Fabio M. .
JOURNAL OF VASCULAR SURGERY, 2007, 45 (06) :1128-1135
[4]   Extent II repair of thoracoabdominal aortic aneurysm secondary to chronic dissection [J].
Coselli, Joseph S. ;
LeMaire, Scott A. ;
Weldon, Scott A. .
ANNALS OF CARDIOTHORACIC SURGERY, 2012, 1 (03) :394-397
[5]   Reversed Frozen Elephant Trunk Technique to Treat a Type II Thoracoabdominal Aortic Aneurysm [J].
Debus, E. Sebastian ;
Koelbel, Tilo ;
Wipper, Sabine ;
Diener, Holger ;
Reiter, Beate ;
Detter, Christian ;
Tsilimparis, Nikolaos .
JOURNAL OF ENDOVASCULAR THERAPY, 2017, 24 (02) :277-280
[6]   Outcomes of Open Repair for Chronic Descending Thoracic Aortic Dissection [J].
Estrera, Anthony L. ;
Jan, Azam ;
Sandhu, Harleen ;
Shalhub, Sherene ;
Medina-Castro, Mario ;
Nguyen, Tom C. ;
Azizzadeh, Ali ;
Charlton-Ouw, Kristofer ;
Miller, Charles C. ;
Safi, Hazim J. .
ANNALS OF THORACIC SURGERY, 2015, 99 (03) :786-794
[7]  
ETHEREDGE SN, 1955, SURGERY, V38, P1071
[8]   Preoperative percutaneous coronary intervention in patients undergoing open thoracoabdominal and descending thoracic aneurysm repair [J].
Girardi, Leonard N. ;
Rabotnikov, Yury ;
Avgerinos, Dimitrios V. .
JOURNAL OF THORACIC AND CARDIOVASCULAR SURGERY, 2014, 147 (01) :163-168
[9]   Thoracoabdominal aortic aneurysm: hybrid repair outcomes [J].
Hughes, G. Chad ;
Andersen, Nicholas D. ;
Hanna, Jennifer M. ;
McCann, Richard L. .
ANNALS OF CARDIOTHORACIC SURGERY, 2012, 1 (03) :311-319
[10]   Outcomes after thoracoabdominal aortic aneurysm repair using hypothermic circulatory arrest [J].
Kouchoukos, Nicholas T. ;
Kulik, Alexander ;
Castner, Catherine F. .
JOURNAL OF THORACIC AND CARDIOVASCULAR SURGERY, 2013, 145 (03) :S139-S141