Personalized stent design for congenital heart defects using pulsatile blood flow simulations

被引:12
|
作者
Gundelwein, L. [1 ,2 ,4 ]
Miro, J. [3 ]
Barlatay, F. Gonzalez [3 ]
Lapierre, C. [3 ]
Rohr, K. [1 ,2 ]
Duong, L. [4 ]
机构
[1] Heidelberg Univ, BioQuant, IPMB, D-69120 Heidelberg, Germany
[2] DKFZ Heidelberg, Biomed Comp Vis Grp, D-69120 Heidelberg, Germany
[3] Ctr Hosp Univ St Justine, 3175 Chemin Cote St Catherine, Montreal, PQ H3T 1C5, Canada
[4] Ecole Technol Super, 1100 Notre Dame St W, Montreal, PQ H3C 1K3, Canada
基金
加拿大自然科学与工程研究理事会;
关键词
Congenital heart defects; Mathematical models; Computational fluid dynamics; Pulmonary artery; Stent; OUTFLOW BOUNDARY-CONDITIONS; PULMONARY; HEMODYNAMICS; DEPLOYMENT; DISEASE; MRI;
D O I
10.1016/j.jbiomech.2018.09.013
中图分类号
Q6 [生物物理学];
学科分类号
071011 ;
摘要
Stent size selection and placement are among the most challenging tasks in the treatment of pulmonary artery stenosis in congenital heart defects (CHD). Patient-specific 3D model from CT or MR improves the understanding of the patient's anatomy and information about the hemodynamics aid in patient risk assessment and treatment planning. This work presents a new approach for personalized stent design in pulmonary artery interventions combining personalized patient geometry and hemodynamic simulations. First, the stent position is initialized using a geometric approach. Second, the stent and artery expansion, including the foreshortening behavior of the stent is simulated. Two stent designs are considered, a regular stent and a Y-stent for bifurcations. Computational fluid dynamics (CFD) simulations of the blood flow in the initial and expanded artery models are performed using patient-specific boundary conditions in form of a pulsatile inflow waveform, 3-element Windkessel outflow conditions, and deformable vessel walls. The simulations have been applied to 16 patient cases with a large variability of anatomies. Finally, the simulations have been clinically validated using retrospective imaging from angiography and pressure measurements. The simulated pressure, volume flow and flow velocity values were on the same order of magnitude as the reference values obtained from clinical measurements, and the simulated stent placement showed a positive impact on the hemodynamic values. Simulation of geometric changes combined with CFD simulations offers the possibility to optimize stent type, size, and position by evaluating different configurations before the intervention, and eventually allow to test customized stent geometries and new deployment techniques in CHD. (C) 2018 Elsevier Ltd. All rights reserved.
引用
收藏
页码:68 / 75
页数:8
相关论文
共 50 条
  • [21] Pathophysiology of coronary blood flow in congenital heart disease
    Aburawi, Elhadi H.
    Pesonen, Erkki
    INTERNATIONAL JOURNAL OF CARDIOLOGY, 2011, 151 (03) : 273 - 277
  • [22] Novel Bioresorbable Stent Design and Fabrication: Congenital Heart Disease Applications
    Welch T.R.
    Eberhart R.C.
    Reddy S.V.
    Wang J.
    Nugent A.
    Forbess J.
    Cardiovascular Engineering and Technology, 2013, 4 (2) : 171 - 182
  • [23] Blood flow patterns underlie developmental heart defects
    Midgett, Madeline
    Thornburg, Kent
    Rugonyi, Sandra
    AMERICAN JOURNAL OF PHYSIOLOGY-HEART AND CIRCULATORY PHYSIOLOGY, 2017, 312 (03): : H632 - H642
  • [24] Human Heart Blood Flow Numerical Modelling and Simulations
    Aksenov, Andrey
    Zhluktov, Sergey
    Zietak, Wojciech
    Cotton, Ross
    Vucinic, Dean
    ADVANCES IN VISUALIZATION AND OPTIMIZATION TECHNIQUES FOR MULTIDISCIPLINARY RESEARCH: TRENDS IN MODELLING AND SIMULATIONS FOR ENGINEERING APPLICATIONS, 2020, : 237 - 263
  • [25] Simulation of pulsatile blood flow through various cardiac defects and quantitative measurements of shunted blood volume
    Sonetha, Vaibhavi A.
    Bellare, Jayesh R.
    2ND INTERNATIONAL CONFERENCE ON NANOMATERIALS AND TECHNOLOGIES (CNT 2014), 2015, 10 : 706 - 713
  • [26] Diffuse Optical Measurements of Blood Oxygenation and Flow for Monitoring CMRO2 in Neonates with Congenital Heart Defects
    Minkoff, D. L.
    Busch, D. R.
    Buckley, E. M.
    Durduran, T.
    Licht, D. J.
    Yodh, A. G.
    MEDICAL PHYSICS, 2010, 37 (06) : 3141 - +
  • [27] The role of fetal nuchal translucency (NT) and ductus venosus blood flow (DV) in the detection of congenital heart defects
    Bas-Budecka, Elzbieta
    Perenc, Malgorzata
    Sieroszewski, Piotr
    GINEKOLOGIA POLSKA, 2010, 81 (04) : 272 - 276
  • [28] Investigation of blood under pulsatile flow using ultrasound imaging
    Paeng, DG
    Choi, MJ
    Shung, KK
    STATE OF THE ART: ULTRASONICS IN MEDICINE, 2004, 1274 : 99 - 108
  • [29] The effects of pulsatile blood flow on stent-induced thromboembolism in a bovine in-vitro model
    Rosa, G
    Sukavaneshvar, S
    Zheng, Y
    Solen, K
    THROMBOSIS AND HAEMOSTASIS, 1999, : 718 - 719
  • [30] STUDIES OF OXYGEN-TRANSPORT IN BLOOD OF CHILDREN WITH CONGENITAL HEART DEFECTS
    MANZKE, H
    ESTERS, W
    DORNER, K
    MALERCZYK, V
    LANGE, P
    ZEITSCHRIFT FUR KARDIOLOGIE, 1975, 64 (06): : 562 - 573