Geometric accuracy of a newly developed cone-beam device for maxillofacial imaging

被引:146
作者
Mischkowski, Robert A.
Pulsfort, Reinhard
Ritter, Lutz
Neugebauer, Joerg
Brochhagen, Hans Georg
Keeve, Erwin
Zoeller, Joachim E.
机构
[1] Univ Cologne, Dept Craniomaxillofacial & Plast Surg, D-50937 Cologne, Germany
[2] Ctr Adv European Studies & Res, Surg Syst Lab, Cologne, Germany
[3] Inst Radiol & Nucl Med Koeln Suelz, Bonn, Germany
来源
ORAL SURGERY ORAL MEDICINE ORAL PATHOLOGY ORAL RADIOLOGY AND ENDODONTOLOGY | 2007年 / 104卷 / 04期
关键词
D O I
10.1016/j.tripleo.2007.02.021
中图分类号
R78 [口腔科学];
学科分类号
1003 ;
摘要
Objective. The aim of this study was to determine the geometric accuracy of scans obtained with a newly developed cone- beam computed tomography ( CBCT) device in comparison with a multidetector row computed tomography ( MDCT) scanner. Study design. Cone- beam scans were obtained with the preretail version of a newly developed compact size device with a scan volume of 15 x 15 x 15 cm. Conventional CT scans for comparison were performed with a 6- detector row CT scanner. To determine distance accuracy, 100 measurements were performed on radiopaque markers on a dry human skull. To determine volume accuracy, 25 measurements were carried out on a geometric phantom. Commercially availables software was used for three- dimensional visualization and measurements on imaging data. Results. Mean absolute measurement error ( AME) for linear distances was 0.26 mm ( +/- 0.18 mm) for the CBCT device and 0.18 mm ( +/- 0.17 mm) for the MDCT device ( P =.196 in paired t test). The average absolute percentage error ( APE) was 0.98% ( +/- 0.73%) and 1.26% ( +/- 1.50%), respectively ( P = .485 in paired t test). Linear regression analysis showed a positive correlation between AME and distance length ( R = .0.628; P =.004) for CBCT- based measurements. Average AME in volume measurements was 1.78 mL ( +/- 0.99 mL) for the CBCT device and 1.23 mL ( +/- 0.93 mL) for the MDCT device. The average APE was 6.01% ( +/- 1.49%) and 4.42% ( +/- 1.99%), respectively. Conclusions. The results indicate that the evaluated cone- beam device provides satisfactory information about linear distances and volumes. Multidetector row computed tomography scans proved slightly more accurate in both measurement categories. The difference may be considered as not relevant for the majority of clinical applications.
引用
收藏
页码:551 / 559
页数:9
相关论文
共 28 条
[1]   Development of a compact computed tomographic apparatus for dental use [J].
Arai, Y ;
Tammisalo, E ;
Iwai, K ;
Hashimoto, K ;
Shinoda, K .
DENTOMAXILLOFACIAL RADIOLOGY, 1999, 28 (04) :245-248
[2]   Using a flat-panel detector in high resolution cone beam CT for dental imaging [J].
Baba, R ;
Ueda, K ;
Okabe, M .
DENTOMAXILLOFACIAL RADIOLOGY, 2004, 33 (05) :285-290
[3]   Automated volumetry at CT colonography: A phantom study [J].
Blake, ME ;
Soto, JA ;
Hayes, RA ;
Ferrucci, JT .
ACADEMIC RADIOLOGY, 2005, 12 (05) :608-613
[4]   ACCURACY OF VOLUME MEASUREMENTS OF COMPUTED-TOMOGRAPHY AND MAGNETIC-RESONANCE-IMAGING PHANTOMS BY 3-DIMENSIONAL RECONSTRUCTION AND PRELIMINARY CLINICAL-APPLICATION [J].
DISLER, DG ;
MARR, DS ;
ROSENTHAL, DI .
INVESTIGATIVE RADIOLOGY, 1994, 29 (08) :739-745
[5]   Volumetric measurement of synthetic lung nodules with multi-detector row CT: Effect of various image reconstruction parameters and segmentation thresholds on measurement accuracy [J].
Goo, JM ;
Tongdee, T ;
Tongdee, R ;
Yeo, K ;
Hildebolt, CF ;
Bae, KT .
RADIOLOGY, 2005, 235 (03) :850-856
[6]   State-of-the-art on cone beam CT imaging for preoperative planning of implant placement [J].
Guerrero, ME ;
Jacobs, R ;
Loubele, M ;
Schutyser, F ;
Suetens, P ;
van Steenberghe, D .
CLINICAL ORAL INVESTIGATIONS, 2006, 10 (01) :1-7
[7]  
Hatcher David C, 2003, J Calif Dent Assoc, V31, P825
[8]   Postoperative imaging of zygomaticomaxillary complex fractures using digital volume tomography [J].
Heiland, M ;
Schulze, DK ;
Rother, U ;
Schmelzle, R .
JOURNAL OF ORAL AND MAXILLOFACIAL SURGERY, 2004, 62 (11) :1387-1391
[9]   Intraoperative 3D imaging of the facial skeleton using the SIREMOBIL Iso-C3D [J].
Heiland, M ;
Schmelzle, R ;
Hebecker, A ;
Schulze, D .
DENTOMAXILLOFACIAL RADIOLOGY, 2004, 33 (02) :130-132
[10]   Erweiterte Diagnostik im Rahmen der operativen Weisheitszahnentfernung mittels digitaler VolumentomographieDigital volume tomography – an extension to the diagnostic procedures available for application before surgical removal of third molars [J].
Th. Heurich ;
C. Ziegler ;
H. Steveling ;
R. Wörtche ;
J. Mühling ;
St. Hassfeld .
Mund-, Kiefer- und Gesichtschirurgie, 2002, 6 (6) :427-432