Modeling of needle-tissue interaction using ultrasound-based motion estimation

被引:0
作者
Dehghan, Ehsan [1 ]
Wen, Xu [1 ]
Zahiri-Azar, Reza [1 ]
Marchal, Maud [1 ,2 ]
Salcudean, Septimiu E. [1 ]
机构
[1] Univ British Columbia, Dept Elect & Comp Engn, Vancouver, BC V5Z 1M9, Canada
[2] TIMC GMCAO Lab, Grenoble, France
来源
MEDICAL IMAGE COMPUTING AND COMPUTER-ASSISTED INTERVENTION - MICCAI 2007, PT 1, PROCEEDINGS | 2007年 / 4791卷
关键词
D O I
暂无
中图分类号
TP301 [理论、方法];
学科分类号
081202 ;
摘要
A needle-tissue interaction model is an essential part of every needle insertion simulator. In this paper, a new experimental method for the modeling of needle-tissue interaction is presented. The method consists of measuring needle and tissue displacements with ultrasound, measuring needle base forces, and using a deformation simulation model to identify the parameters of a needle-tissue interaction model. The feasibility of this non-invasive approach was demonstrated in an experiment in which a brachytherapy needle was inserted into a prostate phantom. Ultrasound radio-frequency data and the time-domain cross-correlation method, often used in ultrasound elastography, were used to generate the tissue displacement field during needle insertion. A three-parameter force density model was assumed for the needle-tissue interaction. With the needle displacement, tissue displacement and needle base forces as input data, finite element simulations were carried out to adjust the model parameters to achieve a good fit between simulated and measured data.
引用
收藏
页码:709 / +
页数:2
相关论文
共 13 条
  • [1] Needle insertion and radioactive seed implantation in human tissues: Simulation and sensitivity analysis
    Alterovitz, R
    Goldberg, K
    Pouliot, J
    Taschereau, R
    Hsu, IC
    [J]. 2003 IEEE INTERNATIONAL CONFERENCE ON ROBOTICS AND AUTOMATION, VOLS 1-3, PROCEEDINGS, 2003, : 1793 - 1799
  • [2] Crouch JR, 2005, LECT NOTES COMPUT SC, V3750, P624, DOI 10.1007/11566489_77
  • [3] DEHGHAN E, 2006, P WORKSH COMP BIOM M, P117
  • [4] Needle insertion modeling and simulation
    DiMaio, SP
    Salcudean, SE
    [J]. IEEE TRANSACTIONS ON ROBOTICS AND AUTOMATION, 2003, 19 (05): : 864 - 875
  • [5] Goksel Orcun, 2006, Comput Aided Surg, V11, P279, DOI 10.1080/10929080601089997
  • [6] Reality-based needle insertion simulation for haptic feedback in prostate brachytherapy
    Hing, James T.
    Brooks, Ari. D.
    Desai, Jaydev P.
    [J]. 2006 IEEE INTERNATIONAL CONFERENCE ON ROBOTICS AND AUTOMATION (ICRA), VOLS 1-10, 2006, : 619 - +
  • [7] Kataoka H, 2002, LECT NOTES COMPUT SC, V2488, P216
  • [8] Measurement of prostate rotation during insertion of needles for brachytherapy
    Lagerburg, V
    Moerland, MA
    Lagendijk, JJW
    Battermann, JJ
    [J]. RADIOTHERAPY AND ONCOLOGY, 2005, 77 (03) : 318 - 323
  • [9] Force modeling for needle insertion into soft tissue
    Okamura, AM
    Simone, C
    O'Leary, MD
    [J]. IEEE TRANSACTIONS ON BIOMEDICAL ENGINEERING, 2004, 51 (10) : 1707 - 1716
  • [10] Podder T K, 2006, Conf Proc IEEE Eng Med Biol Soc, V2006, P555