Preliminary Single-Center Experience with the Bolton Treovance Endograft in the Treatment of Abdominal Aortic Aneurysms

被引:9
作者
Georgakarakos, Efstratios [1 ]
Georgiadis, George S. [1 ]
Argyriou, Christos [1 ]
Schoretsanitis, Nikolaos [1 ]
Antoniou, George A. [2 ]
Lazarides, Miltos K. [1 ]
机构
[1] Democritus Univ Thrace, Univ Hosp Alexandroupolis, Dept Vasc Surg, Alexandroupolis, Greece
[2] Pennine Acute Hosp NHS Trust, Royal Oldham Hosp, Dept Vasc & Endovascular Surg, Manchester, Lancs, England
关键词
ENDOVASCULAR AAA REPAIR; ENDURANT STENT-GRAFT; LIMB OCCLUSION; EVAR; ANGULATION; OUTCOMES; DEVICE; ANGIOGRAPHY; TORTUOSITY; DEPLOYMENT;
D O I
10.1016/j.avsg.2015.12.026
中图分类号
R61 [外科手术学];
学科分类号
摘要
Background: Our aim was to present our preliminary experience with the recently introduced Treovance aortic stent-graft device ( Bolton Medical, Barcelona, Spain) in the treatment of abdominal aortic aneurysm ( AAA). Methods: Eight patients underwent treatment of an infrarenal AAA ( mean maximum diameter, 56.4 +/- 6.8 mm) with the Treovance device. Iliac tortuosity was considered mild, moderate, or severe when >= 1 angulation of 45-90 degrees, 1 angulation >= 90 degrees, or >= 2 angulations >= 90 degrees, respectively, were present. Results: Mild angulation of the infrarenal neck ( 10-45 degrees) was present in 7 patients, whereas the remaining patient had severe infrarenal neck angulation ( 65 degrees). Three patients had severe iliac tortuosity. Primary technical success was achieved in all but 1 patient in whom a type Ia endoleak was identified on completion angiogram. The endoleak was successfully treated with a proximal aortic cuff. A femoral access complication occurred in 1 patient. Mean follow-up was 6.8 months ( range, 1-12). No device-related serious adverse events or rupture occurred during the given follow-up period. The only type II endoleak identified resolved spontaneously within 12 months. Conclusions: The Treovance abdominal stent-graft system seems to guarantee an accurate, safe, and effective deployment in AAA even through angulated and tortuous iliac vessels. Although our preliminary results are promising, follow-up data are needed to establish the durability of this new-generation endovascular device in standard or challenging anatomies.
引用
收藏
页码:68 / 74
页数:7
相关论文
共 32 条
[1]  
Broeders IAMJ, 1997, J ENDOVASC SURG, V4, P252, DOI 10.1583/1074-6218(1997)004<0252:PSOGFT>2.0.CO
[2]  
2
[3]   Impact of exclusion criteria on patient selection for endovascular abdominal aortic aneurysm repair [J].
Carpenter, JP ;
Baum, RA ;
Barker, CF ;
Golden, MA ;
Mitchell, ME ;
Velazquez, OC ;
Fairman, RM .
JOURNAL OF VASCULAR SURGERY, 2001, 34 (06) :1050-1054
[4]   Predicting iliac limb occlusions after bifurcated aortic stent grafting: Anatomic and device-related causes [J].
Carroccio, A ;
Faries, PL ;
Morrissey, NJ ;
Teodorescu, V ;
Burks, JA ;
Gravereaux, EC ;
Hollier, LH ;
Marin, ML .
JOURNAL OF VASCULAR SURGERY, 2002, 36 (04) :679-684
[5]   Current Status of Investigational Devices for EVAR: Similarities and Differences [J].
Cayne, Neal S. ;
Adelman, Mark A. ;
Veith, Frank J. .
SEMINARS IN VASCULAR SURGERY, 2009, 22 (03) :127-131
[6]   Identifying and grading factors that modify the outcome of endovascular aortic aneurysm repair [J].
Chaikof, EL ;
Fillinger, MF ;
Matsumura, JS ;
Rutherford, RB ;
White, GH ;
Blankensteijn, JD ;
Bernhard, VM ;
Harris, PL ;
Kent, KC ;
May, J ;
Veith, FJ ;
Zarins, CK .
JOURNAL OF VASCULAR SURGERY, 2002, 35 (05) :1061-1066
[7]   Reporting standards for endovascular aortic aneurysm repair [J].
Chaikof, EL ;
Blankensteijn, JD ;
Harris, PL ;
White, GH ;
Zarins, CK ;
Bernhard, VM ;
Matsumura, JS ;
May, J ;
Veith, FJ ;
Fillinger, MF ;
Rutherford, RB ;
Kent, KC .
JOURNAL OF VASCULAR SURGERY, 2002, 35 (05) :1048-1060
[8]  
Chiesa R, 2012, J CARDIOVASC SURG, V53, P595
[9]   Limb graft occlusion following EVAR: Clinical pattern, outcomes and predictive factors of occurrence [J].
Cochennec, F. ;
Becquemin, J. P. ;
Desgranges, P. ;
Allaire, E. ;
Kobeiter, H. ;
Roudot-Thoraval, F. .
EUROPEAN JOURNAL OF VASCULAR AND ENDOVASCULAR SURGERY, 2007, 34 (01) :59-65
[10]   Stent-Graft Limb Deployment in the External Iliac Artery Increases the Risk of Limb Occlusion Following Endovascular AAA Repair [J].
Conway, Allan M. ;
Modarai, Bijan ;
Taylor, Peter R. ;
Carrell, Tom W. G. ;
Waltham, Matthew ;
Salter, Richard ;
Bell, Rachel E. .
JOURNAL OF ENDOVASCULAR THERAPY, 2012, 19 (01) :79-85