Classification of Sleep Apnea Severity by Electrocardiogram Monitoring Using a Novel Wearable Device

被引:35
作者
Baty, Florent [1 ]
Boesch, Maximilian [1 ]
Widmer, Sandra [1 ]
Annaheim, Simon [2 ]
Fontana, Piero [2 ]
Camenzind, Martin [2 ]
Rossi, Rene M. [2 ]
Schoch, Otto D. [1 ]
Brutsche, Martin H. [1 ]
机构
[1] Cantonal Hosp St Gallen, Lung Ctr, Rorschacherstr 95, CH-9007 St Gallen, Switzerland
[2] Empa, Lab Biomimet Membranes & Text, Lerchenfeldstr 5, CH-9014 St Gallen, Switzerland
关键词
sleep apnea; classification algorithms; ECG signal; wearable acquisition device; heart rate variability analysis; support vector machine; HEART-RATE-VARIABILITY; RISK-FACTOR; DIAGNOSIS; POLYSOMNOGRAPHY; VALIDATION; ALGORITHMS;
D O I
10.3390/s20010286
中图分类号
O65 [分析化学];
学科分类号
070302 ; 081704 ;
摘要
Sleep apnea (SA) is a prevalent disorder diagnosed by polysomnography (PSG) based on the number of apnea-hypopnea events per hour of sleep (apnea-hypopnea index, AHI). PSG is expensive and technically complex; therefore, its use is rather limited to the initial diagnostic phase and simpler devices are required for long-term follow-up. The validity of single-parameter wearable devices for the assessment of sleep apnea severity is still debated. In this context, a wearable electrocardiogram (ECG) acquisition system (ECG belt) was developed and its suitability for the classification of sleep apnea severity was investigated using heart rate variability analysis with or without data pre-filtering. Several classification algorithms were compared and support vector machine was preferred due to its simplicity and overall performance. Whole-night ECG signals from 241 patients with a suspicion of sleep apnea were recorded using both the ECG belt and patched ECG during PSG recordings. 65% of patients had an obstructive sleep apnea and the median AHI was 21 [IQR: 7-40] h(-1). The classification accuracy obtained from the ECG belt (accuracy: 72%, sensitivity: 70%, specificity: 74%) was comparable to the patched ECG (accuracy: 74%, sensitivity: 88%, specificity: 61%). The highest classification accuracy was obtained for the discrimination between individuals with no or mild SA vs. moderate to severe SA. In conclusion, the ECG belt provided signals comparable to patched ECG and could be used for the assessment of sleep apnea severity, especially during follow-up.
引用
收藏
页数:12
相关论文
共 38 条
[1]   DETERMINATION OF OPTIMAL THERMAL CONDITIONS FOR GROWTH OF CLAM (VENERUPIS-PULLASTRA) SEED [J].
ALBENTOSA, M ;
BEIRAS, R ;
CAMACHO, AP .
AQUACULTURE, 1994, 126 (3-4) :315-328
[2]   Reproducibility and predictors of the apnea hypopnea index across multiple nights [J].
Alshaer, Hisham ;
Ryan, Clodagh ;
Fernie, Geoff R. ;
Bradley, T. Douglas .
SLEEP SCIENCE, 2018, 11 (01) :28-33
[3]  
[Anonymous], 2007, The AASM Manual for the Scoring of Sleep and Associated Events: Rules, Terminology and Technical Specifications
[4]  
[Anonymous], J ANESTH CLIN RES
[5]  
[Anonymous], P INT C ISMAC COMP V
[6]   A novel approach using actigraphy to quantify the level of disruption of sleep by in-home polysomnography: the MrOS Sleep Study Sleep disruption by polysomnography [J].
Blackwell, Terri ;
Paudel, Misti ;
Redline, Susan ;
Ancoli-Israel, Sonia ;
Stone, Katie L. .
SLEEP MEDICINE, 2017, 32 :97-104
[7]   Evaluation of a portable recording device (ApneaLink™) for case selection of obstructive sleep apnea [J].
Chen, Hui ;
Lowe, Alan A. ;
Bai, Yuxing ;
Hamilton, Peter ;
Fleetham, John A. ;
Almeida, Fernanda R. .
SLEEP AND BREATHING, 2009, 13 (03) :213-219
[8]  
Collop NA, 2007, J CLIN SLEEP MED, V3, P737
[9]   Automated processing of the single-lead electrocardiogram for the detection of obstructive sleep apnoea [J].
de Chazal, P ;
Heneghan, C ;
Sheridan, E ;
Reilly, R ;
Nolan, P ;
O'Malley, M .
IEEE TRANSACTIONS ON BIOMEDICAL ENGINEERING, 2003, 50 (06) :686-696
[10]  
de Chazal P, 2015, IEEE ENG MED BIO, P7671, DOI 10.1109/EMBC.2015.7320169