Assessing the accuracy and precision of musculoskeletal motion tracking using cine-PC MRI on a 3.0T platform

被引:47
作者
Behnam, Abrahm J. [1 ,2 ]
Herzka, Daniel A. [2 ]
Sheehan, Frances T. [1 ]
机构
[1] NIH, Dept Rehabil Med, Bethesda, MD 20892 USA
[2] Johns Hopkins Sch Med, Dept Biomed Engn, Baltimore, MD USA
基金
美国国家卫生研究院;
关键词
Patella; Knee; Bone; Muscle; Validation; Phantom; Dynamic; 3-DIMENSIONAL PATELLAR TRACKING; IN-VIVO; KINEMATICS; JOINT; FEASIBILITY; FLUOROSCOPY;
D O I
10.1016/j.jbiomech.2010.08.029
中图分类号
Q6 [生物物理学];
学科分类号
071011 ;
摘要
The rising cost of musculoskeletal pathology, disease, and injury creates a pressing need for accurate and reliable methods to quantify 3D musculoskeletal motion, fostering a renewed interest in this area over the past few years. To date, cine-phase contrast (PC) MRI remains the only technique capable of non-invasively tracking in vivo 3D musculoskeletal motion during volitional activity, but current scan times are long on the 1.5T MR platform (similar to 2.5 min or 75 movement cycles). With the clinical availability of higher field strength magnets (3.0T) that have increased signal-to-noise ratios, it is likely that scan times can be reduced while improving accuracy. Therefore, the purpose of this study is to validate cine-PC MRI on a 3.0T platform, in terms of accuracy, precision, and subject-repeatability, and to determine if scan time could be minimized. On the 3.0T platform it is possible to limit scan time to 2 min, with sub-millimeter accuracy (<0.33 mm/0.97 degrees), excellent technique precision (<0.18 degrees), and strong subject-repeatability (<0.73 mm/1.10 degrees). This represents reduction in imaging time by 25% (42 s), a 50% improvement in accuracy, and a 72% improvement in technique precision over the original 1.5T platform. Scan time can be reduced to 1 min (30 movement cycles), but the improvements in accuracy are not as large. Published by Elsevier Ltd.
引用
收藏
页码:193 / 197
页数:5
相关论文
共 24 条
[1]  
*ASTM, 2010, STAND PRACT US TERMS, pE177
[2]   Accuracy of biplane x-ray imaging combined with model-based tracking for measuring in-vivo patellofemoral joint motion [J].
Bey M.J. ;
Kline S.K. ;
Tashman S. ;
Zauel R. .
Journal of Orthopaedic Surgery and Research, 3 (1)
[3]  
Boling M., 2009, SCAND J MED SCI SPOR
[4]   TRACKING THE MOTION OF SKELETAL-MUSCLE WITH VELOCITY-ENCODED MR-IMAGING [J].
DRACE, JE ;
PELC, NJ .
JOURNAL OF MAGNETIC RESONANCE IMAGING, 1994, 4 (06) :773-778
[5]   Feasibility of using real-time MRI to measure joint kinematics in 1.5T and open-bore 0.5T systems [J].
Draper, Christine E. ;
Santos, Juan M. ;
Kourtis, Lampros C. ;
Besier, Thor F. ;
Fredericson, Michael ;
Beaupre, Gary S. ;
Gold, Garry E. ;
Delp, Scott L. .
JOURNAL OF MAGNETIC RESONANCE IMAGING, 2008, 28 (01) :158-166
[6]   Magnetic resonance imaging for in vivo assessment of three-dimensional patellar tracking [J].
Fellows, RA ;
Hill, NA ;
Gill, HS ;
MacIntyre, NJ ;
Harrison, MM ;
Ellis, RE ;
Wilson, DR .
JOURNAL OF BIOMECHANICS, 2005, 38 (08) :1643-1652
[7]   Repeatability of a novel technique for in vivo measurement of three-dimensional patellar tracking using magnetic resonance Imaging [J].
Fellows, RA ;
Hill, NA ;
MacIntyre, NJ ;
Harrison, MM ;
Ellis, RE ;
Wilson, DR .
JOURNAL OF MAGNETIC RESONANCE IMAGING, 2005, 22 (01) :145-153
[8]   Integrating modelling, motion capture and x-ray fluoroscopy to investigate patellofemoral function during dynamic activity [J].
Fernandez, J. W. ;
Akbarshahi, M. ;
Kim, H. J. ;
Pandy, M. G. .
COMPUTER METHODS IN BIOMECHANICS AND BIOMEDICAL ENGINEERING, 2008, 11 (01) :41-53
[9]   Theoretical accuracy of model-based shape matching for measuring natural knee kinematics with single-plane fluoroscopy [J].
Fregly, BJ ;
Rahman, HA ;
Banks, SA .
JOURNAL OF BIOMECHANICAL ENGINEERING-TRANSACTIONS OF THE ASME, 2005, 127 (04) :692-699
[10]   Retrospective Case Evaluation of Gender Differences in Sports Injuries in a Japanese Sports Medicine Clinic [J].
Iwamoto, Jun ;
Takeda, Tsuyoshi ;
Sato, Yoshihiro ;
Matsumoto, Hideo .
GENDER MEDICINE, 2008, 5 (04) :405-414