Finite element stress analysis of Ti-6Al-4V and partially stabilized zirconia dental implant during clenching

被引:12
|
作者
Choi, Andy H. [1 ,2 ]
Matinlinna, Jukka Pekka [1 ]
Ben-Nissan, Besim [2 ]
机构
[1] Univ Hong Kong, Fac Dent, Hong Kong, Hong Kong, Peoples R China
[2] Univ Technol Sydney, Fac Sci, Sydney, NSW 2007, Australia
关键词
finite element analysis; FEA; mandible; dental implant materials; clenching; metallic dental implant; ceramic dental implant; partially stabilized zirconia; titanium; Ti-6Al-4V; PS-ZrO2; GUIDED TISSUE REGENERATION; ELASTIC PROPERTIES; BIOMECHANICS; IMMEDIATE; PRINCIPLE; STRAIGHT; BONE; BITE;
D O I
10.3109/00016357.2011.600723
中图分类号
R78 [口腔科学];
学科分类号
1003 ;
摘要
Objective. The purpose of this paper is to compare the differences in stress between Ti-6Al-4V and PS-ZrO2 dental implant during clenching and whether these changes are clinically significant to limit the use of zirconia in oral implantology. Materials and methods. The model geometry was derived from position measurements taken from 28 diamond blade cut cross-sections of an average size human adult edentulous mandible and generated using a special sequencing method. Data on anatomical, structural, functional aspects and material properties were obtained from measurements and published data. Ti6-Al-4V and PS-ZrO2 dental implants were modelled as cylindrical structure with a diameter of 3.26 mm and length of 12.00 mm was placed in the first molar region on the right hemimandible. Results. The analysis revealed an increase of 2-3% in the averaged tensile and compressive stress and an increase of 8% in the averaged Von Mises stress were recorded in the bone-implant interface when PS-ZrO2 dental implant was used instead of Ti-6Al-4V dental implant. The results also revealed only relatively low levels of stresses were transferred from the implant to the surrounding cortical and cancellous bone, with the majority of the stresses transferred to the cortical bone. Conclusion. Even though high magnitudes of tensile, compressive and Von Mises stresses were recorded on the Ti-6Al-4V and PS-ZrO2 dental implants and in the surrounding osseous structures, the stresses may not be clinically critical since the mechanical properties of the implant material and the cortical and cancellous bone could withstand stress magnitudes far greater than those recorded in this analysis.
引用
收藏
页码:353 / 361
页数:9
相关论文
共 50 条
  • [1] Effects of micromovement on the changes in stress distribution of partially stabilized zirconia (PS-ZrO2) dental implants and bridge during clenching: A three-dimensional finite element analysis
    Choi, Andy H.
    Matinlinna, Jukka
    Ben-Nissan, Besim
    ACTA ODONTOLOGICA SCANDINAVICA, 2013, 71 (01) : 72 - 81
  • [2] Influence of Ti-6Al-4V and Ti-15Zr Dental Implants on the Stress in Mandibular Bone A finite element analysis study
    Dima, Cosmin
    Agop-Forna, Doriana
    Moldoveanu, Sanziana Butnaru
    Forna, Consuela Norina
    MATERIALE PLASTICE, 2019, 56 (02) : 469 - 473
  • [3] Finite element simulation of residual stress in machining of Ti-6Al-4V with a microstructural consideration
    Pan, Zhipeng
    Liang, Steven Y.
    Garmestani, Hamid
    PROCEEDINGS OF THE INSTITUTION OF MECHANICAL ENGINEERS PART B-JOURNAL OF ENGINEERING MANUFACTURE, 2019, 233 (04) : 1103 - 1111
  • [4] Finite element analysis of spring back in Ti-6Al-4V alloy
    Wahed, M. A.
    Gadi, V. S. R.
    Gupta, A. K.
    Supradeepan, K.
    Singh, S. K.
    Kotkunde, N. R.
    MATERIALS TODAY-PROCEEDINGS, 2019, 18 : 2693 - 2699
  • [5] The Role of Nicotine in the Corrosive Behavior of a Ti-6Al-4V Dental Implant
    Royhman, Dmitry
    Dominguez-Benetton, Xochitl
    Yuan, Judy Chia-Chun
    Shokuhfar, Tolou
    Takoudis, Christos
    Mathew, Mathew T.
    Sukotjo, Cortino
    CLINICAL IMPLANT DENTISTRY AND RELATED RESEARCH, 2015, 17 : E352 - E363
  • [6] Finite Element Analysis of Ultrasonic Vibration Assisted Milling of Ti-6Al-4V
    Huang, Hongfu
    Lu, Dong
    Wu, Yongbo
    Yang, Mingming
    MATERIALS PROCESSING TECHNOLOGY II, PTS 1-4, 2012, 538-541 : 823 - 827
  • [7] Finite element analysis of stress distribution and burst failure of SiCf/Ti-6Al-4V composite ring
    Zhang, Hong-yuan
    Yang, Yan-qing
    Luo, Xian
    TRANSACTIONS OF NONFERROUS METALS SOCIETY OF CHINA, 2015, 25 (01) : 261 - 270
  • [8] Finite Element Analysis of Zirconia Dental Implant
    Fiorillo, Luca
    Milone, Dario
    D'Andrea, Danilo
    Santonocito, Dario
    Risitano, Giacomo
    Cervino, Gabriele
    Cicciu, Marco
    PROSTHESIS, 2022, 4 (03): : 490 - 499
  • [9] Diffusion Bonding of Ti-6Al-4V Sheet with Ti-6Al-4V Foam for Biomedical Implant Applications
    Hamilton, Brittany
    Oppenheimer, Scott
    Dunand, David C.
    Lewis, Daniel
    METALLURGICAL AND MATERIALS TRANSACTIONS B-PROCESS METALLURGY AND MATERIALS PROCESSING SCIENCE, 2013, 44 (06): : 1554 - 1559
  • [10] A Finite Element Analysis of Errors in Axisymmetricisothermal and Gleeble Compression Testing of Ti-6Al-4V
    C. J. Bennett
    S. B. Leen
    P. H. Shipway
    International Journal of Material Forming, 2010, 3 : 1155 - 1158