Accuracy and Precision of Two Computer-Assisted Methods of Radiographic Wear Measurement in Total Hip Arthroplasty

被引:13
作者
Crockarell, John R., Jr. [1 ]
Snearly, Christine M. [2 ]
机构
[1] Univ Tennessee, Campbell Clin, Dept Orthopaed Surg, Collierville, TN 38017 USA
[2] InMot Orthopaed Res Ctr, Memphis, TN USA
关键词
Hip Analysis Suite; Roentgen Monographic Analysis Tool; total hip arthroplasty; MEASURING POLYETHYLENE WEAR;
D O I
10.1016/j.arth.2011.04.008
中图分类号
R826.8 [整形外科学]; R782.2 [口腔颌面部整形外科学]; R726.2 [小儿整形外科学]; R62 [整形外科学(修复外科学)];
学科分类号
摘要
Roentgen Monographic Analysis Tool (ROMAN) and Hip Analysis Suite (HAS) were used to analyze radiographs of a phantom hip model. Displacements of known magnitude and direction were produced using dial micrometers. Differences between the known displacement and the programs reported displacement were compared. Hip Analysis Suite was superior with a median error of 0.075 mm (range, 0.019-0.205 mm) compared with 0.137 mm (range, 0.008-0.389 mm) for ROMAN (P = .002). Hip Analysis Suite was also more precise when evaluating intraobserver variability, with a standard deviation between radiographs of 0.007 mm (range, 0.002-0.009 mm), whereas ROMAN's standard deviation was 0.117 mm (range, 0.007-0.153 mm). Repeatability for HAS was 0.019 mm and 0.325 mm for ROMAN. Hip Analysis Suite was more accurate and precise than ROMAN under experimental conditions with digital radiographs.
引用
收藏
页码:37 / 40
页数:4
相关论文
共 13 条
[1]  
*ASTM INT, 2008, E17708 ASTM INT
[2]  
Bland JM, 1996, BRIT MED J, V313, P744
[3]   STATISTICAL METHODS FOR ASSESSING AGREEMENT BETWEEN TWO METHODS OF CLINICAL MEASUREMENT [J].
BLAND, JM ;
ALTMAN, DG .
LANCET, 1986, 1 (8476) :307-310
[4]   Comparison of femoral head penetration using RSA and the Martell method [J].
Bragdon, Charles R. ;
Martell, John M. ;
Greene, Meridith E. ;
Estok, Daniel M., II ;
Thanner, Jonas ;
Karrholm, Johan ;
Harris, William H. ;
Malchau, Henrik .
CLINICAL ORTHOPAEDICS AND RELATED RESEARCH, 2006, (448) :52-57
[5]   A new approach for the Martell 3-D method of measuring polyethylene wear without requiring the cross-table lateral films [J].
Bragdon, CR ;
Martell, JM ;
Estok, DM ;
Greene, ME ;
Malchau, H ;
Harris, WH .
JOURNAL OF ORTHOPAEDIC RESEARCH, 2005, 23 (04) :720-725
[6]   Accuracy of measurement of polyethylene wear with use of radiographs of total hip replacements [J].
Ebramzadeh, E ;
Sangiorgio, SN ;
Lattuada, F ;
Kang, JS ;
Chiesa, R ;
McKellop, HA ;
Dorr, LD .
JOURNAL OF BONE AND JOINT SURGERY-AMERICAN VOLUME, 2003, 85A (12) :2378-2384
[7]   The determination of linear and angular penetration of the femoral head into the acetabular component as an assessment of wear in total hip replacement - A comparison of four computer-assisted methods [J].
Geerdink, C. H. ;
Grimm, B. ;
Vencken, W. ;
Heyligers, I. C. ;
Tonino, A. J. .
JOURNAL OF BONE AND JOINT SURGERY-BRITISH VOLUME, 2008, 90B (07) :839-846
[8]   Assessing the pattern of femoral head penetration after total hip arthroplasty [J].
Hopper, RH ;
Young, AM ;
Engh, CA ;
McAuley, JP .
JOURNAL OF ARTHROPLASTY, 2004, 19 (07) :22-29
[9]   Validation of two and three-dimensional radiographic techniques for measuring polyethylene wear after total hip arthroplasty [J].
Hui, AJ ;
McCalden, FW ;
Martell, JM ;
MacDonald, SJ ;
Bourne, RB ;
Rorabeck, CH .
JOURNAL OF BONE AND JOINT SURGERY-AMERICAN VOLUME, 2003, 85A (03) :505-511
[10]   ON A TEST OF WHETHER ONE OF 2 RANDOM VARIABLES IS STOCHASTICALLY LARGER THAN THE OTHER [J].
MANN, HB ;
WHITNEY, DR .
ANNALS OF MATHEMATICAL STATISTICS, 1947, 18 (01) :50-60