Closed-Loop Fluid Resuscitation Control Via Blood Volume Estimation

被引:15
作者
Bighamian, Ramin [1 ]
Kim, Chang-Sei [1 ]
Reisner, Andrew T. [2 ]
Hahn, Jin-Oh [1 ]
机构
[1] Univ Maryland, Dept Mech Engn, 2181 Glenn L Martin Hall, College Pk, MD 20742 USA
[2] Massachusetts Gen Hosp, Dept Emergency Med, 55 Fruit St, Boston, MA 02114 USA
来源
JOURNAL OF DYNAMIC SYSTEMS MEASUREMENT AND CONTROL-TRANSACTIONS OF THE ASME | 2016年 / 138卷 / 11期
关键词
SYSTOLIC PRESSURE VARIATION; ARTERIAL PULSE PRESSURE; STROKE VOLUME; CARDIAC-OUTPUT; HEART-RATE; CARDIOVASCULAR-RESPONSE; VENTRICULAR INTERACTION; CONTRACTILE STATE; EJECTION TIME; MANAGEMENT;
D O I
10.1115/1.4033833
中图分类号
TP [自动化技术、计算机技术];
学科分类号
0812 ;
摘要
This paper presents a closed-loop control of fluid resuscitation to overcome hypovolemia based on model-based estimation of relative changes in blood volume (BV). In this approach, the control system consists of a model-based relative BV (RBV) estimator and a feedback controller. The former predicts relative changes in the BV response to augmented fluid by analyzing an arterial blood pressure (BP) waveform and the electrocardiogram (ECG). Then, the latter determines the amount of fluid to be augmented by comparing target versus predicted relative changes in BV. In this way, unlike many previous methods for fluid resuscitation based on controlled variable(s) nonlinearly correlated with the changes in BV, fluid resuscitation can be guided by a controlled variable linearly correlated with the changes in BV. This paper reports initial design of the closed-loop fluid resuscitation system and its in silico evaluation in a wide range of hypovolemic scenarios. The results suggest that closed-loop fluid resuscitation guided by a controlled variable linearly correlated with the changes in BV can be effective in overcoming hypovolemia: across 100 randomly produced hypovolemia cases, it resulted in the BV regulation error of 7.98 +/- 171.6 ml, amounting to 0.18 +/- 3.04% of the underlying BV. When guided by pulse pressure (PP), a classical controlled variable nonlinearly correlated with the changes in BV; the same closed-loop fluid resuscitation system resulted in persistent under-resuscitation with the BV regulation error of -779.1 +/- 147.4 ml, amounting to -13.9 +/- 2.65% of the underlying BV.
引用
收藏
页数:8
相关论文
共 75 条
[11]   Noninvasive Hemodynamic Monitoring: Can We Have It All? [J].
Cannesson, Maxime ;
Pinsky, Michael .
ANESTHESIA AND ANALGESIA, 2015, 120 (01) :10-11
[12]  
Cannesson M, 2011, EXPERT REV MED DEVIC, V8, P635, DOI [10.1586/erd.11.30, 10.1586/ERD.11.30]
[13]   PULSE PRESSURE VARIATION: WHERE ARE WE TODAY? [J].
Cannesson, Maxime ;
Aboy, Mateo ;
Hofer, Christoph K. ;
Rehman, Mohamed .
JOURNAL OF CLINICAL MONITORING AND COMPUTING, 2011, 25 (01) :45-56
[14]   Arterial Pressure Variation and Goal-Directed Fluid Therapy [J].
Cannesson, Maxime .
JOURNAL OF CARDIOTHORACIC AND VASCULAR ANESTHESIA, 2010, 24 (03) :487-497
[15]  
CHIEN S, 1971, P SOC EXP BIOL MED, V136, P271
[16]  
Consortium Spinal Cord Med, 2008, J SPINAL CORD MED, V31, P403
[17]   Arterial pulse pressure and its association with reduced stroke volume during progressive central hypovolemia [J].
Convertino, Victor A. ;
Cooke, William H. ;
Holcomb, John B. .
JOURNAL OF TRAUMA-INJURY INFECTION AND CRITICAL CARE, 2006, 61 (03) :629-634
[18]   Can one predict fluid responsiveness in spontaneously breathing patients? [J].
De Backer, Daniel ;
Pinsky, Michael R. .
INTENSIVE CARE MEDICINE, 2007, 33 (07) :1111-1113
[19]  
DeBey R K, 1987, Biomed Sci Instrum, V23, P195
[20]  
DELLINGER RP, 2013, INTENS CARE MED, V39, P165, DOI DOI 10.1007/s00134-012-2769-8