Microfabricated on-chip-type electrochemical flow immunoassay system for the detection of histamine released in whole blood samples

被引:62
作者
Lim, TK [1 ]
Ohta, H [1 ]
Matsunaga, T [1 ]
机构
[1] Tokyo Univ Agr & Technol, Dept Biotechnol & Life Sci, Koganei, Tokyo 1848588, Japan
关键词
D O I
10.1021/ac020749n
中图分类号
O65 [分析化学];
学科分类号
070302 ; 081704 ;
摘要
This paper describes an on-chip-type electrochemical flow immunoassay system with a multichanneled matrix column. The multichanneled matrix column was functionally coated with cation-exchange resin and used for separation of proteins. Antihistamine immunoglobulin G (IgG) antibody conjugated with ferrocenemonocarboxylic acid (Fc) was also prepared and used as a novel analytical reagent. Antibody-antigen complexes were separated from free Fc-conjugated IgG antibody (Fc-IgG) on the basis of differences in isoelectric point (pI) using the multichanneled matrix column coated with cation-exchange resin. The assay yields a good relationship between current and histamine concentration in the range of 200-2000 ng/ mL. This simple technique enables the assay of histamine released in whole blood within 2 min. Furthermore, a good correlation was found between the response of the electrochemical immunoassay described in this paper and the conventional RIA (radioimmunoassay). This on-chip-type electrochemical flow immunoassay requires only minute quantities of whole blood samples and generates highly reproducible results.
引用
收藏
页码:3316 / 3321
页数:6
相关论文
共 46 条
  • [1] Protein microchips: Use for immunoassay and enzymatic reactions
    Arenkov, P
    Kukhtin, A
    Gemmell, A
    Voloshchuk, S
    Chupeeva, V
    Mirzabekov, A
    [J]. ANALYTICAL BIOCHEMISTRY, 2000, 278 (02) : 123 - 131
  • [2] Passive sensitization of human airways induces mast cell degranulation and release of tryptase
    Berger, P
    N'guyen, C
    Buckley, M
    Scotto-Gomez, E
    Marthan, R
    Tunon-de-Lara, JM
    [J]. ALLERGY, 2002, 57 (07) : 592 - 599
  • [3] Analytic chemistry: Everyone's a (future) chemist
    Burns, MA
    [J]. SCIENCE, 2002, 296 (5574) : 1818 - 1819
  • [4] ASSAY OF PROTEIN-QUINONE COUPLING INVOLVING COMPOUNDS STRUCTURALLY RELATED TO ACTIVE PRINCIPLE OF POISON IVY
    BYCK, JS
    DAWSON, CR
    [J]. ANALYTICAL BIOCHEMISTRY, 1968, 25 (1-3) : 123 - &
  • [5] Microfabricated devices in biotechnology and biochemical processing
    Chován, T
    Guttman, A
    [J]. TRENDS IN BIOTECHNOLOGY, 2002, 20 (03) : 116 - 122
  • [6] Electron transfer between ferrocene-modified Au/octadecanethiol/lipid BLM electrode and redox couples in solution
    Cui, XL
    Jiang, DL
    Diao, P
    Li, JX
    Tong, RT
    Wang, XK
    [J]. BIOELECTROCHEMISTRY AND BIOENERGETICS, 1999, 48 (01): : 243 - 247
  • [7] ZERO-LENGTH CROSSLINKING PROCEDURE WITH THE USE OF ACTIVE ESTERS
    GRABAREK, Z
    GERGELY, J
    [J]. ANALYTICAL BIOCHEMISTRY, 1990, 185 (01) : 131 - 135
  • [8] Fabrication of nanocolumns for liquid chromatography
    He, B
    Tait, N
    Regnier, F
    [J]. ANALYTICAL CHEMISTRY, 1998, 70 (18) : 3790 - 3797
  • [9] Initial study of using a laminar fluid diffusion interface for sample preparation in high-performance liquid chromatography
    Jandik, P
    Weigl, BH
    Kessler, N
    Cheng, J
    Morris, CJ
    Schulte, T
    Avdalovic, N
    [J]. JOURNAL OF CHROMATOGRAPHY A, 2002, 954 (1-2) : 33 - 40
  • [10] Kakuta M, 2001, CHEM REC, V1, P395