Stabilised Biosensing Using Needle-Based Recess Electrodes

被引:8
作者
Anastasova, Salzitsa [1 ]
Spehar-Deleze, Anna-Maria [1 ]
Bickham, Dale [1 ]
Uebel, Patrick [2 ]
Schmidt, Markus [2 ]
Russell, Philip [2 ]
Vadgama, Pankaj [1 ]
机构
[1] Univ London, London, England
[2] Max Planck Inst Sci Light, D-91058 Erlangen, Germany
基金
英国工程与自然科学研究理事会;
关键词
Recessed tip electrode; Oxygen; Glucose; Biofouling; Selectivity; in vivo; IN-VIVO; GLUCOSE BIOSENSORS; EXTENDED LINEARITY; ENZYME ELECTRODE; OXYGEN DELIVERY; SENSOR; DIFFUSION; MEMBRANE; PERFORMANCE; TRANSPORT;
D O I
10.1002/elan.201100511
中图分类号
O65 [分析化学];
学科分类号
070302 ; 081704 ;
摘要
A recess disk electrode for amperometric monitoring of oxygen and glucose is reported. The basic design of sensor is a needle structure for easy implantation, embodying a gold filled silica capillary with a 80 mu m inner diameter working electrode. Recess and inlaid disc electrodes, placed inside stainless steel tubes which served as a counter/reference electrode, were compared. The working electrode surface was modified with different treatments and barrier membranes to achieve selectivity for the analytes of interest. The basic needle format is ideal for in vivo use, and the format is easily extendable to other analyte targets. The sensors show linear working range for oxygen up to 160 mmHg partial pressure of oxygen, and for glucose from 1 to 10 mM. Bio-fouling, as assessed by exposure to bovine serum albumin, was significantly reduced. Response times for the recess construct was increased but remained within the acceptable range for physiological monitoring. The operational stability of the sensors is demonstrated as well as the interference-free detection of peroxide in the presence of physiologically relevant levels of ascorbic acid, uric acid, acetaminophen, and catechol. Preliminary in vivo tests showed excellent response towards glucose.
引用
收藏
页码:529 / 538
页数:10
相关论文
共 41 条
[1]   The determination of diffusion coefficients of counter ion in an ion exchange membrane using electrical conductivity measurement [J].
Amang, DN ;
Alexandrova, S ;
Schaetzel, P .
ELECTROCHIMICA ACTA, 2003, 48 (18) :2563-2569
[2]  
[Anonymous], 1993, NEW ENGL J MED, V329, P977, DOI [DOI 10.1056/NEJM199309303291401, 10.1056/NEJM199309303291401]
[3]  
[Anonymous], BIOELECTROCHEMISTRY
[4]  
[Anonymous], SENS ACTUATORS B
[5]  
[Anonymous], MEMBRANES IMPROVE AM
[6]   A glucose-sensing contact lens: from bench top to patient [J].
Badugu, R ;
Lakowicz, JR ;
Geddes, CD .
CURRENT OPINION IN BIOTECHNOLOGY, 2005, 16 (01) :100-107
[7]   Dynamic delay and maximal dynamic error in continuous biosensors [J].
Baker, DA ;
Gough, DA .
ANALYTICAL CHEMISTRY, 1996, 68 (08) :1292-1297
[8]   GLUCOSE BIOSENSORS BASED ON THICK-FILM TECHNOLOGY [J].
BILITEWSKI, U ;
RUGER, P ;
SCHMID, RD .
BIOSENSORS & BIOELECTRONICS, 1991, 6 (04) :369-373
[9]   Continuous Glucose Sensing: Future Technology Developments [J].
Brauker, Jim .
DIABETES TECHNOLOGY & THERAPEUTICS, 2009, 11 :S25-S36
[10]   OXYGEN DELIVERY AND UPTAKE IN DOGS DURING ANEMIC AND HYPOXIC HYPOXIA [J].
CAIN, SM .
JOURNAL OF APPLIED PHYSIOLOGY, 1977, 42 (02) :228-234