Evaluation of the overall efficacy of the Omron office digital blood pressure HEM-907 monitor in adults

被引:175
作者
White, WB [1 ]
Anwar, YA [1 ]
机构
[1] Univ Connecticut, Ctr Hlth, Sect Hypertens & Clin Pharmacol, Farmington, CT 06030 USA
关键词
non-invasive blood pressure; Association for the Advancement of Medical Instrumentation (AAMI) validation;
D O I
10.1097/00126097-200104000-00007
中图分类号
R6 [外科学];
学科分类号
1002 ; 100210 ;
摘要
Background Non-invasive self blood pressure monitoring has become increasingly popular. To assure the accuracy of devices used for this purpose, all need to be validated independently prior to marketing. The objective of this study was to assess the accuracy of the HEM-907, a new semi-automatic, non-invasive, osciliometric blood pressure monitoring device specifically designed to be used in the clinic or physician's office setting. Methods Blood pressure measurements taken employing this device were compared with the results obtained by two experienced observers using a mercury sphygmomanometer on 100 subjects and patients (384 measurements). The limits of agreement were calculated for the device compared with the results of the two observers according to the standards of the Association for the Advancement of Medical instrumentation (AAMI). Results The agreement between the two observers was -0.36 +/- 2.32 mmHg for systolic blood pressure and 0.02 +/- 2.42 mmHg for diastolic blood pressure. The agreement between the device and the observers was 1.56 +/- 4.42 mmHg and 3.49 +/- 4.61 mmHg for systolic and diastolic blood pressure respectively. Conclusions The Omron HEM-907 satisfied the AAMI criteria for accuracy for a non-invasive blood pressure monitoring device. Blood Press Monit 6:107-110 (C) 2001 Lippincott Williams & Wilkins.
引用
收藏
页码:107 / 110
页数:4
相关论文
共 9 条
[1]  
ANISANWAR Y, 1998, BLOOD PRESS MONIT, V3, P261
[2]  
ANISANWAR Y, 1997, BLOOD PRESS MONIT, V2, P105
[3]  
*ASS ADV MED INSTR, 1992, AM NAT STAND EL AUT
[4]   STATISTICAL METHODS FOR ASSESSING AGREEMENT BETWEEN TWO METHODS OF CLINICAL MEASUREMENT [J].
BLAND, JM ;
ALTMAN, DG .
LANCET, 1986, 1 (8476) :307-310
[5]   CHARACTERIZATION OF THE OSCILLOMETRIC METHOD FOR MEASURING INDIRECT BLOOD-PRESSURE [J].
GEDDES, LA ;
VOELZ, M ;
COMBS, C ;
REINER, D ;
BABBS, CF .
ANNALS OF BIOMEDICAL ENGINEERING, 1982, 10 (06) :271-280
[6]   Recommendations for the use of home (self) and ambulatory blood pressure monitoring [J].
Pickering, T .
AMERICAN JOURNAL OF HYPERTENSION, 1996, 9 (01) :1-11
[7]  
WHITE W, 1998, HYPERTENSION PRIMER, P302
[8]   NATIONAL STANDARD FOR MEASUREMENT OF RESTING AND AMBULATORY BLOOD PRESSURES WITH AUTOMATED SPHYGMOMANOMETERS [J].
WHITE, WB ;
BERSON, AS ;
ROBBINS, C ;
JAMIESON, MJ ;
PRISANT, LM ;
ROCCELLA, E ;
SHEPS, SG .
HYPERTENSION, 1993, 21 (04) :504-509
[9]  
White WB, 1998, BLOOD PRESS MONIT S, V3, pS7