Bringing Light into Darkness-Comparison of Different Personal Dosimeters for Assessment of Solar Ultraviolet Exposure

被引:6
作者
Strehl, Claudine [1 ]
Heepenstrick, Timo [1 ]
Knuschke, Peter [2 ]
Wittlich, Marc [3 ]
机构
[1] Inst Occupat Safety & Hlth German Social Accid In, Dept Ergon Phys Environm Factors, D-53757 St Augustin, Germany
[2] Tech Univ Dresden, Fac Med, Dept Dermatol, D-01307 Dresden, Germany
[3] German Social Accid Insurance, Dept Accid Prevent Digitalisat Technol, Inst Occupat Safety & Hlth, D-53757 St Augustin, Germany
关键词
personal dosimetry; UV radiation; intercalibration; UV-RADIATION; POLYSULFONE FILM; SPORE DOSIMETRY; VALENCIA; ERYTHEMA; MODEL; BODY;
D O I
10.3390/ijerph18179071
中图分类号
X [环境科学、安全科学];
学科分类号
08 ; 0830 ;
摘要
(1) Measuring personal exposure to solar ultraviolet radiation (UVR) poses a major challenges for researchers. Often, the study design determines the measuring devices that can be used, be it the duration of measurements or size restrictions on different body parts. It is therefore of great importance that measuring devices produce comparable results despite technical differences and modes of operation. Particularly when measurement results from different studies dealing with personal UV exposure are to be compared with each other, the need for intercomparability and intercalibration factors between different measurement systems becomes significant. (2) Three commonly used dosimeter types-(polysulphone film (PSF), biological, and electronic dosimeters)-were selected to perform intercalibration measurements. They differ in measurement principle and sensitivity, measurement accuracy, and susceptibility to inaccuracies. The aim was to derive intercalibration factors for these dosimeter types. (3) While a calibration factor between PSF and electronic dosimeters of about 1.3 could be derived for direct irradiation of the dosimeters, this was not the case for larger angles of incidence of solar radiation with increasing fractions of diffuse irradiation. Electronic dosimeters show small standard deviation across all measurements. For biological dosimeters, no intercalibration factor could be found with respect to PSF and electronic dosimeters. In a use case, the relation between steady-state measurements and personal measurements was studied. On average, persons acquired only a small fraction of the ambient radiation.
引用
收藏
页数:13
相关论文
共 56 条
[1]  
[Anonymous], 2011, 503111201104 DIN
[2]   Erythemal Ultraviolet Exposure in Two Groups of Outdoor Workers in Valencia, Spain [J].
Antonia Serrano, Maria ;
Canada, Javier ;
Carlos Moreno, Juan .
PHOTOCHEMISTRY AND PHOTOBIOLOGY, 2009, 85 (06) :1468-1473
[3]  
Arrandale VH, 2016, CAN J PUBLIC HEALTH, V107, pE119, DOI [10.17269/CJPH.107.5039, 10.17269/cjph.107.5039]
[4]   Behavioral and structural prevention of skin cancer Implementation and effectiveness [J].
Baldermann, C. ;
Weiskopf, D. .
HAUTARZT, 2020, 71 (08) :572-579
[5]  
BioSense, VIOSPOR BLUE LIN TEC
[6]  
BLUM HF, 1959, JNCI-J NATL CANCER I, V23, P319
[7]   PERSONNEL MONITORING OF EXPOSURE TO UV-RADIATION [J].
CHALLONER, AVJ ;
CORLESS, D ;
DAVIS, A ;
DEANE, GHW ;
DIFFEY, BL ;
GUPTA, SP ;
MAGNUS, IA .
CLINICAL AND EXPERIMENTAL DERMATOLOGY, 1976, 1 (02) :175-179
[8]   Estimation of Individual Exposure to Erythemal Weighted UVR by Multi-Sensor Measurements and Integral Calculation [J].
Cheng, Wenwen ;
Brown, Robert ;
Vernez, David ;
Goldberg, Daniel .
SENSORS, 2020, 20 (15) :1-13
[9]  
Cockell CS, 2001, PHOTOCHEM PHOTOBIOL, V74, P570, DOI 10.1562/0031-8655(2001)074<0570:EOAFST>2.0.CO
[10]  
2