Measurement of volumetric flow by real-time 3-dimensional Doppler echocardiography in children

被引:16
作者
Lu, Xiuzhang
Nadvoretskiy, Vyacheslav
Klas, Berthold
Bu, Liping
Stolpen, Alan
Ayres, Nancy A.
Sahn, David J.
Ge, Shuping
机构
[1] Baylor Coll Med, Dept Pediat, Lillie Frank Abercrombie Sect Cardiol, Houston, TX 77030 USA
[2] Texas Childrens Hosp, Houston, TX 77030 USA
[3] Texas Heart Inst, Houston, TX 77025 USA
[4] TomTec Imaging Syst, Munich, Germany
[5] Univ Iowa, Iowa City, IA 52242 USA
[6] Oregon Hlth & Sci Univ, Portland, OR 97201 USA
关键词
D O I
10.1016/j.echo.2007.01.025
中图分类号
R5 [内科学];
学科分类号
1002 ; 100201 ;
摘要
Background: We sought to assess the accuracy and reproducibility of an automated real-time (RT) 3-dimensional OD) Doppler echocardiography (RT3DDE) technique for measuring volumetric flow (VF) in children. Methods: A total of 19 healthy children (age 11.5 +/- 3.5 years) were studied to measure VF through mitral valve (MV), aortic valve (AV), pulmonary valve (PV), and tricuspid valve (W) by RT3DDE. RT 3D echocardiography was also performed to measure left ventricular (LV) end-systolic volume, LV end-diastolic volume, and stroke volume (stroke volume = LV end-diastolic volume - LV end-systolic volume), which served as a reference standard for comparison with VF by RT3DDE. Results: Compared with stroke volume by RT 3D echocardiography, the correlation with VF was excellent for MV (r = 0.91), good for AV (r = 0.89) and PV (r = 0.89), but poor for TV (r = 0.20) by RT3DDE. There were good agreements for AV (bias = 0.9 +/- 5.0 mL), PV (bias = 0.4 +/- 5.7 mL), and MV (bias = 4.1 +/- 4.7 mL), and marked underestimation for TV (bias = -24.4 +/- 14.6 mL). Conclusions: Our data demonstrated that VF measurement by RT3DDE is feasible and reasonably accurate for MV, AV, and PV but problematic for TV.
引用
收藏
页码:915 / 920
页数:6
相关论文
共 30 条
[1]   Volumetric blood flow measurement with the use of dynamic 3-dimensional ultrasound color flow imaging [J].
Berg, S ;
Torp, H ;
Haugen, BO ;
Samstad, S .
JOURNAL OF THE AMERICAN SOCIETY OF ECHOCARDIOGRAPHY, 2000, 13 (05) :393-402
[2]   STATISTICAL METHODS FOR ASSESSING AGREEMENT BETWEEN TWO METHODS OF CLINICAL MEASUREMENT [J].
BLAND, JM ;
ALTMAN, DG .
LANCET, 1986, 1 (8476) :307-310
[3]   Rapid full volume data acquisition by real-time 3-dimensional echocardiography for assessment of left ventricular indexes in children: A validation study compared with magnetic resonance imaging [J].
Bu, LP ;
Munns, S ;
Zhang, HH ;
Disterhoft, M ;
Dixon, M ;
Stolpen, A ;
Sonka, M ;
Scholz, TD ;
Mahoney, LT ;
Ge, SP .
JOURNAL OF THE AMERICAN SOCIETY OF ECHOCARDIOGRAPHY, 2005, 18 (04) :299-305
[4]   THE MITRAL-VALVE ORIFICE METHOD FOR NON-INVASIVE TWO-DIMENSIONAL ECHO DOPPLER DETERMINATIONS OF CARDIAC-OUTPUT [J].
FISHER, DC ;
SAHN, DJ ;
FRIEDMAN, MJ ;
LARSON, D ;
VALDESCRUZ, LM ;
HOROWITZ, S ;
GOLDBERG, SJ ;
ALLEN, HD .
CIRCULATION, 1983, 67 (04) :872-877
[5]  
Ge S, 1996, J Am Soc Echocardiogr, V9, P700, DOI 10.1016/S0894-7317(96)90067-X
[6]   A real-time 3-dimensional digital Doppler method for measurement of flow rate and volume through mitral valve in children: A validation study compared with magnetic resonance imaging [J].
Ge, SP ;
Bu, LP ;
Zhang, HH ;
Schelbert, E ;
Disterhoft, M ;
Li, XN ;
Li, XK ;
Sahn, D ;
Stolpen, A ;
Sonka, M .
JOURNAL OF THE AMERICAN SOCIETY OF ECHOCARDIOGRAPHY, 2005, 18 (01) :1-7
[7]   NORMAL INTRACARDIAC AND GREAT VESSEL DOPPLER FLOW VELOCITIES IN INFANTS AND CHILDREN [J].
GRENADIER, E ;
LIMA, CO ;
ALLEN, HD ;
SAHN, DJ ;
BARRON, JV ;
VALDESCRUZ, LM ;
GOLDBERG, SJ .
JOURNAL OF THE AMERICAN COLLEGE OF CARDIOLOGY, 1984, 4 (02) :343-350
[8]   Comparison of left ventricular volumes and ejection fractions measured by three-dimensional echocardiography versus by two-dimensional echocardiography and cardiac magnetic resonance in patients with various cardiomyopathies [J].
Gutiérrez-Chico, JL ;
Zamorano, JL ;
de Isla, LP ;
Orejas, M ;
Almería, C ;
Rodrigo, JL ;
Ferreirós, J ;
Serra, V ;
Macaya, C .
AMERICAN JOURNAL OF CARDIOLOGY, 2005, 95 (06) :809-813
[9]  
Haugen B O, 2000, Eur J Echocardiogr, V1, P252, DOI 10.1053/euje.2000.0041
[10]  
Haugen B O, 2000, Eur J Echocardiogr, V1, P204, DOI 10.1053/euje.2000.0044