摘要
To identify range deviations by using Compton cameras (CCs), tomographic image reconstruction of CC data is needed. Within this context, image reconstruction is usually performed using maximum likelihood expectation maximization (MLEM), and more recently, the origin ensemble (OE) algorithm. In this article, we investigate how MLEM and OE affect the precision and accuracy of estimated range deviations. In particular, we focus on the effects of data selection, statistical fluctuations, and artifact reduction. The use of external information of the beam path through a hodoscope was also explored. Additionally, two methods to calculate range deviations were tested. To this aim, realistic proton beams were simulated using GATE and data from single spots as well as from seven contiguous spots of an energy layer were reconstructed. MLEM and OE reacted differently to the poor data statistics. In general, both algorithms were able to detect range shifts for single spots, particularly when multiple coincidences were also considered. Selection of events corresponding to the most relevant energy peaks decreased the identification performance due to the lower statistics. When data from several contiguous spots were jointly reconstructed, the accuracy of the results degraded significantly, and nonzero shifts were assigned when no shifts had occurred. The limited size of the cameras and the subsequent restriction in the orientation and aperture of detected cones, as well as in the number of detected events are major challenges. Future efforts should be devoted to noise regularization and compensation for data truncation.
引用
收藏
页码:233 / 242
页数:10
相关论文
共 43 条
  • [1] Resolution recovery for Compton camera using origin ensemble algorithm
    Andreyev, A.
    Celler, A.
    Ozsahin, I.
    Sitek, A.
    [J]. MEDICAL PHYSICS, 2016, 43 (08) : 4866 - 4876
  • [2] Fast image reconstruction for Compton camera using stochastic origin ensemble approach
    Andreyev, Andriy
    Sitek, Arkadiusz
    Celler, Anna
    [J]. MEDICAL PHYSICS, 2011, 38 (01) : 429 - 438
  • [3] [Anonymous], GESAMMELTE WERKE
  • [4] Berger M.J., 1999, ESTAR, PSTAR, and ASTAR: computer programs for calculating stopping-power and range tables for electrons, protons, and helium ions
  • [5] 3D prompt gamma imaging for proton beam range verification
    Draeger, E.
    Mackin, D.
    Peterson, S.
    Chen, H.
    Avery, S.
    Beddar, S.
    Polf, J. C.
    [J]. PHYSICS IN MEDICINE AND BIOLOGY, 2018, 63 (03)
  • [6] Feasibility Studies of a New Event Selection Method to Improve Spatial Resolution of Compton Imaging for Medical Applications
    Draeger E.
    Peterson S.
    Mackin D.
    Chen H.
    Beddar S.
    Polf J.C.
    [J]. IEEE Trans. Radiat. Plasma Med. Sci., 4 (358-367): : 358 - 367
  • [7] Double-Field Hadrontherapy Treatment Monitoring With the INSIDE In-Beam PET Scanner: Proof of Concept
    Ferrero, Veronica
    Bisogni, Maria Giuseppina
    Camarlinghi, Niccolo'
    Fiorina, Elisa
    Giraudo, Giuseppe
    Morrocchi, Matteo
    Pennazio, Francesco
    Sportelli, Giancarlo
    Wheadon, Richard
    Cerello, Piergiorgio
    [J]. IEEE TRANSACTIONS ON RADIATION AND PLASMA MEDICAL SCIENCES, 2018, 2 (06) : 588 - 593
  • [8] Large surface gamma cameras for medical imaging: characterization of the bismuth germanate blocks
    Fontana, M.
    Dauvergne, D.
    Della Negra, R.
    Letang, J. M.
    Mounier, F.
    Testa, E.
    Zanetti, L.
    Zoccarato, Y.
    Collaboration, CLaRyS
    [J]. JOURNAL OF INSTRUMENTATION, 2018, 13
  • [9] A Tracking Compton-Scattering Imaging System for Hadron Therapy Monitoring
    Frandes, Mirela
    Zoglauer, Andreas
    Maxim, Voichita
    Prost, Remy
    [J]. IEEE TRANSACTIONS ON NUCLEAR SCIENCE, 2010, 57 (01) : 144 - 150
  • [10] Gillam JE, 2012, IEEE NUCL SCI CONF R, P3298