Electrochemical impedance spectroscopy characterization of passive film formed on implant Ti-6Al-7Nb alloy in Hank's solution

被引:190
作者
Lavos-Valereto, IC
Wolynec, S
Ramires, I
Guastaldi, AC
Costa, I
机构
[1] Univ Sao Paulo, Polytech Sch, Dept Met & Mat Engn, BR-05508900 Sao Paulo, Brazil
[2] Sao Paulo State Univ, Inst Chem, Dept Phys Chem, Araraquara, SP, Brazil
[3] IPEN CNEN SP, Ctr Mat Sci & Technol, Sao Paulo, Brazil
基金
巴西圣保罗研究基金会;
关键词
D O I
10.1023/B:JMSM.0000010097.86245.74
中图分类号
R318 [生物医学工程];
学科分类号
0831 ;
摘要
The influence of potential on electrochemical behavior of Ti-6Al-7Nb alloy under simulate physiological conditions was investigated by electrochemical impedance spectroscopy (EIS). The experimental results were compared with those obtained by potentiodynamic polarization curves. All measurements were carried out in Hank's aerated solution at 25degreesC, at pH 7.8 and at different potentials (corrosion potential, 0 mV(SCE), 1000 mV(SCE), and 2000 mV(SCE)). The EIS spectra exhibited a two-step or a two-time constant system, suggesting the formation of a two-layer oxide film on the metal surface. The high corrosion resistance, displayed by this alloy in electrochemical polarization tests, is due to the dense inner layer, while its osseointegration ability can be ascribed to the presence of the outer porous layer. (C) 2004 Kluwer Academic Publishers.
引用
收藏
页码:55 / 59
页数:5
相关论文
共 20 条
[1]   Electrochemical characterization of cast titanium alloys [J].
Cai, Z ;
Shafer, T ;
Watanabe, I ;
Nunn, ME ;
Okabe, T .
BIOMATERIALS, 2003, 24 (02) :213-218
[2]   ELECTROCHEMICAL-BEHAVIOR OF PASSIVE LAYERS ON TITANIUM [J].
ELBASIOUNY, MS ;
MAZHAR, AA .
CORROSION, 1982, 38 (05) :237-240
[3]  
Gluszek J, 1997, BIOMATERIALS, V18, P789
[4]   Study of the corrosion behavior of titanium and some of its alloys for biomedical and dental implant applications [J].
González, JEG ;
Mirza-Rosca, JC .
JOURNAL OF ELECTROANALYTICAL CHEMISTRY, 1999, 471 (02) :109-115
[5]  
HEFNY MM, 1982, BRIT CORROS J, V17, P38, DOI 10.1179/000705982798274633
[6]   AC-IMPEDANCE MEASUREMENTS ON CORRODED POROUS ALUMINUM-OXIDE FILMS [J].
HITZIG, J ;
JUTTNER, K ;
LORENZ, WJ .
JOURNAL OF THE ELECTROCHEMICAL SOCIETY, 1986, 133 (05) :887-892
[7]   Electrochemical corrosion of titanium and titanium-based alloys [J].
Kuphasuk, C ;
Oshida, Y ;
Andres, CJ ;
Hovijitra, ST ;
Barco, MT ;
Brown, DT .
JOURNAL OF PROSTHETIC DENTISTRY, 2001, 85 (02) :195-202
[8]   The electrochemical behavior of Ti-6Al-7Nb alloy with and without plasma-sprayed hydroxyapatite coating in Hank's solution [J].
Lavos-Valereto, C ;
Costa, I ;
Wolynec, S .
JOURNAL OF BIOMEDICAL MATERIALS RESEARCH, 2002, 63 (05) :664-670
[9]   A study of histological responses from Ti-6Al-7Nb alloy dental implants with and without plasma-sprayed hydroxyapatite coating in dogs [J].
Lavos-Valereto, IC ;
Konig, B ;
Rossa, C ;
Marcantonio, E ;
Zavaglia, AC .
JOURNAL OF MATERIALS SCIENCE-MATERIALS IN MEDICINE, 2001, 12 (03) :273-276
[10]  
LAVOSVALERETO IC, UNPUB LONG TERM PERF