Experimental evaluation of two simple thermal models using transient temperature analysis

被引:53
作者
Kolios, MC
Worthington, AE
Sherar, MD
Hunt, JW
机构
[1] Univ Toronto, Ontario Canc Inst, Div Expt Therapeut, Toronto, ON M5G 29, Canada
[2] Univ Toronto, Dept Med Biophys, Toronto, ON M5G 29, Canada
[3] Univ Toronto, Ontario Canc Inst, Div Med Phys, Toronto, ON M5G 29, Canada
关键词
D O I
10.1088/0031-9155/43/11/011
中图分类号
R318 [生物医学工程];
学科分类号
0831 ;
摘要
Thermal models are used to predict temperature distributions of heated tissues during thermal therapies. Recent interest in short duration high temperature therapeutic procedures necessitates the accurate modelling of transient temperature profiles in heated tissues. Blood flow plays an important role in tissue heat transfer and the resultant temperature distribution. This work examines the transient predictions of two simple mathematical models of heat transfer by blood flow (the bioheat transfer equation model and the effective thermal conductivity equation model) and compares their predictions to measured transient temperature data. Large differences between the two models are predicted in the tissue temperature distribution as a function of blood flow for a short heat pulse. In the experiments a hot water needle, similar to 30 degrees C above ambient, delivered a 20 s heating pulse to an excised fixed porcine kidney that was used as a Bow model. Temperature profiles of a thermocouple that primarily traversed the kidney cortex were examined. Kidney locations with large vessels were avoided in the temperature profile analysis by examination of the vessel geometry using high resolution computed tomography angiography and the detection of the characteristic large vessel localized cooling or heating patterns in steady-state temperature profiles. It was found that for regions without large vessels, predictions of the Pennes bioheat transfer equation were in much better agreement with the experimental data when compared to predictions of the scalar effective thermal conductivity equation model. For example, at a location r similar to 2 mm away from the source, the measured delay time was 10.6 +/- 0.5 s compared to predictions of 9.4 s and 5.4 s of the BHTE and ETCE models, respectively. However, for the majority of measured locations, localized cooling and heating effects were detected close to large vessels when the kidney was perfused. Finally, it is shown that increasing Bow in regions without large vessels minimally perturbs temperature profiles for short exposure times; regions with large vessels still have a significant effect.
引用
收藏
页码:3325 / 3340
页数:16
相关论文
共 49 条
  • [1] RECENT DEVELOPMENTS IN MODELING HEAT-TRANSFER IN BLOOD-PERFUSED TISSUES
    ARKIN, H
    XU, LX
    HOLMES, KR
    [J]. IEEE TRANSACTIONS ON BIOMEDICAL ENGINEERING, 1994, 41 (02) : 97 - 107
  • [2] EFFECTS OF PHYSICAL PARAMETERS ON HIGH-TEMPERATURE ULTRASOUND HYPERTHERMIA
    BILLARD, BE
    HYNYNEN, K
    ROEMER, RB
    [J]. ULTRASOUND IN MEDICINE AND BIOLOGY, 1990, 16 (04) : 409 - 420
  • [3] ESTIMATION OF THE THERMAL EFFECT OF BLOOD-FLOW IN A BRANCHING COUNTERCURRENT NETWORK USING A 3-DIMENSIONAL VASCULAR MODEL
    BRINCK, H
    WERNER, J
    [J]. JOURNAL OF BIOMECHANICAL ENGINEERING-TRANSACTIONS OF THE ASME, 1994, 116 (03): : 324 - 330
  • [4] EFFICIENCY FUNCTION - IMPROVEMENT OF CLASSICAL BIOHEAT APPROACH
    BRINCK, H
    WERNER, J
    [J]. JOURNAL OF APPLIED PHYSIOLOGY, 1994, 77 (04) : 1617 - 1622
  • [5] HEAT-TRANSFER TO BLOOD-VESSELS
    CHATO, JC
    [J]. JOURNAL OF BIOMECHANICAL ENGINEERING-TRANSACTIONS OF THE ASME, 1980, 102 (02): : 110 - 118
  • [6] Chen M M, 1980, Ann N Y Acad Sci, V335, P137, DOI 10.1111/j.1749-6632.1980.tb50742.x
  • [7] MAGNETIC RESONANCE-GUIDED THERMAL SURGERY
    CLINE, HE
    SCHENCK, JF
    WATKINS, RD
    HYNYNEN, K
    JOLESZ, FA
    [J]. MAGNETIC RESONANCE IN MEDICINE, 1993, 30 (01) : 98 - 106
  • [8] Simultaneous magnetic resonance phase and magnitude temperature maps in muscle
    Cline, HE
    Hynynen, K
    Schneider, E
    Hardy, CJ
    Maier, SE
    Watkins, RD
    Jolesz, FA
    [J]. MAGNETIC RESONANCE IN MEDICINE, 1996, 35 (03) : 309 - 315
  • [9] EXPERIMENTAL-VERIFICATION OF BIOHEAT TRANSFER THEORIES - MEASUREMENT OF TEMPERATURE PROFILES AROUND LARGE ARTIFICIAL VESSELS IN PERFUSED TISSUE
    CREZEE, J
    LAGENDIJK, JJW
    [J]. PHYSICS IN MEDICINE AND BIOLOGY, 1990, 35 (07) : 905 - 923
  • [10] INTERSTITIAL HEATING - EXPERIMENTS IN ARTIFICIALLY PERFUSED BOVINE TONGUES
    CREZEE, J
    MOOIBROEK, J
    BOS, CK
    LAGENDIJK, JJW
    [J]. PHYSICS IN MEDICINE AND BIOLOGY, 1991, 36 (06) : 823 - 833