Kinetic performance of a 50 mm long 1.8 μm chiral column in supercritical fluid chromatography

被引:17
作者
Berger, Terry A. [1 ]
机构
[1] SFC Solut Inc, 9435 Downing St, Englewood, FL 34224 USA
关键词
Chiral column; 1.8 mu m particles; h(r) <2d(p); Variable vanDeemter minima; Sub-1 min chromatograms; Extra-column dispersion; ULTRA-HIGH-PRESSURE; LIQUID-CHROMATOGRAPHY; M PARTICLES; STATIONARY PHASES; SIZE DISTRIBUTION; HIGH-EFFICIENCY; DRUG DISCOVERY; UHPLC; PURIFICATION; SEPARATION;
D O I
10.1016/j.chroma.2016.07.012
中图分类号
Q5 [生物化学];
学科分类号
071010 ; 081704 ;
摘要
Reduced plate heights (h(r)) of <2 were observed for the first time during the chiral separation of enantiomers, on sub-2 mu m particles with supercritical fluid chromatography (SFC). The enantiomers of trans-stilbene oxide, were separated on a 4.6 x 50 mm, 1.8 mu m R,R-Whelk-O1 column, with hr as low as 1.93. The plumbing of a commercial SFC instrument was modified to create a low dispersion version. Without the modification performance was considerably worse. vanDeemter like plots of reduced plate height vs. flow rate, for trans-stilbene oxide, indicate that the optimum flow varied with% modifier. On a 4.6 x 250 mm, 5 mu m R,R- Whelk-O1 column, the optimum flow was >4mL/min for 5% methanol in CO2, decreasing to <2 mL/min for 40% methanol (more than a factor of 2). For a 4.6 x 50 mm column packed with 1.8 mu m particles the optimum appeared to be near, or >5 mL/min with 2.5%, 5%, and 10% methanol, decreasing to between 3 and 3.5 mL/min at 40% methanol. This is the first time such shifts have been characterized. Since the solutes were the same in all cases, the differences are likely due to changes in solute diffusion coefficients caused by changes in modifier concentration, and pressure. Pump pressure requirements sometimes exceeded 500 bar. It is shown that a 5 mL/min flow rate is inadequate for use with 1.8 mu m particles in a 4.6 mm ID column format. Instead, it is suggested to decrease the ID of the column to 3 mm, where the optimum flow rates are on the order of 2 mi./min with decreased tubing variance. Nevertheless, a number of sub-1 min chromatograms are presented. (C) 2016 Elsevier B.V. All rights reserved.
引用
收藏
页码:136 / 144
页数:9
相关论文
共 18 条
[1]   High efficiency, narrow particle size distribution, sub-2 μm based macrocyclic glycopeptide chiral stationary phases in HPLC and SFC [J].
Barhate, Chandan L. ;
Wahab, M. Farooq ;
Breitbach, Zachary S. ;
Bell, David S. ;
Armstrong, Daniel W. .
ANALYTICA CHIMICA ACTA, 2015, 898 :128-137
[2]   Instrument modifications that produced reduced plate heights &lt;2 with sub-2 μm particles and 95% of theoretical efficiency at k=2 in supercritical fluid chromatography [J].
Berger, Terry A. .
JOURNAL OF CHROMATOGRAPHY A, 2016, 1444 :129-144
[3]   Demonstration of High Speeds with Low Pressure Drops Using 1.8 μm Particles in SFC [J].
Berger, Terry A. .
CHROMATOGRAPHIA, 2010, 72 (7-8) :597-602
[4]  
Broeckhoven K., 2014, TRAC, V63, P21
[5]   Transition from enantioselective high performance to ultra-high performance liquid chromatography: A case study of a brush-type chiral stationary phase based on sub-5-micron to sub-2-micron silica particles [J].
Cancelliere, Giovanna ;
Ciogli, Alessia ;
D'Acquarica, Ilaria ;
Gasparrini, Francesco ;
Kocergin, Jelena ;
Misiti, Domenico ;
Pierini, Marco ;
Ritchie, Harald ;
Simone, Patrizia ;
Villani, Claudio .
JOURNAL OF CHROMATOGRAPHY A, 2010, 1217 (07) :990-999
[6]   Ultra-high-pressure liquid chromatography (UHPLC) in method development [J].
Dong, Michael W. ;
Zhang, Kelly .
TRAC-TRENDS IN ANALYTICAL CHEMISTRY, 2014, 63 :21-30
[7]   Current and future trends in UHPLC [J].
Fekete, Szabolcs ;
Schappler, Julie ;
Veuthey, Jean-Luc ;
Guillarme, Davy .
TRAC-TRENDS IN ANALYTICAL CHEMISTRY, 2014, 63 :2-13
[8]   Fast Chiral Separation of Drugs Using Columns Packed with Sub-2 μm Particles and Ultra-High Pressure [J].
Guillarme, Davy ;
Bonvin, Gregoire ;
Badoud, Flavia ;
Schappler, Julie ;
Rudaz, Serge ;
Veuthey, Jean-Luc .
CHIRALITY, 2010, 22 (03) :320-330
[9]   The evaluation of 25 chiral stationary phases and the utilization of sub-2.0 μm coated polysaccharide chiral stationary phases via supercritical fluid chromatography [J].
Hamman, Chris ;
Wong, Mengling ;
Aliagas, Ignacio ;
Ortwine, Daniel F. ;
Pease, Joseph ;
Schmidt, Donald E., Jr. ;
Victorino, Joseph .
JOURNAL OF CHROMATOGRAPHY A, 2013, 1305 :310-319
[10]   Ultra-fast high-efficiency enantioseparations by means of a teicoplanin-based chiral stationary phase made on sub-2 μm totally porous silica particles of narrow size distribution [J].
Ismail, Omar H. ;
Ciogli, Alessia ;
Villani, Claudio ;
De Martino, Michela ;
Pierini, Marco ;
Cavazzini, Alberto ;
Bell, David S. ;
Gasparrini, Francesco .
JOURNAL OF CHROMATOGRAPHY A, 2016, 1427 :55-68