Continuous non-invasive finger arterial pressure monitoring reflects intra-arterial pressure changes in children undergoing cardiac surgery

被引:37
作者
Hofhuizen, C. M. [2 ]
Lemson, J. [1 ]
Hemelaar, A. E. A. [1 ]
Settels, J. J. [4 ]
Schraa, O. [4 ]
Singh, S. K. [3 ]
van der Hoeven, J. G. [1 ]
Scheffer, G. J. [2 ]
机构
[1] Radboud Univ Nijmegen, Med Ctr, Dept Intens Care Med, NL-6500 HB Nijmegen, Netherlands
[2] Radboud Univ Nijmegen, Med Ctr, Dept Anaesthesiol, NL-6500 HB Nijmegen, Netherlands
[3] Radboud Univ Nijmegen, Med Ctr, Dept Cardiothorac Surg, NL-6500 HB Nijmegen, Netherlands
[4] BMEYE BV Ctr Point 1, NL-1101 BE Amsterdam, Netherlands
关键词
arterial pressure; haemodynamic monitoring; paediatric anaesthesia; paediatrics; BLOOD-PRESSURE; ACCURACY; FINAPRES; FEASIBILITY; ANESTHESIA; AGREEMENT;
D O I
10.1093/bja/aeq201
中图分类号
R614 [麻醉学];
学科分类号
100217 ;
摘要
Continuous non-invasive measurement of finger arterial pressure (FAP) is a reliable technology in adults. FAP is measured with an inflatable cuff around the finger and simultaneously converted to a reconstructed brachial artery pressure waveform (reBAP) by the Nexfin (TM) device. We assessed the adequacy of a prototype device (Nexfin-paediatric), designed for a paediatric population, for detecting rapid arterial pressure changes in children during cardiac surgery. Thirteen anaesthetized children with a median age of 11 months (2 months-7 yr) undergoing congenital cardiac surgery were included in the study. reBAP and intra-arterial pressure (IAP) were recorded simultaneously during the surgical procedure. To assess the accuracy of reBAP in tracking arterial pressure changes, the four largest IAP variations within a 5 min time interval were identified from each procedure. These variations were compared offline with reBAP during a 10 s control period before and a 10 s period after an arterial pressure change had occurred. In 10 out of 13 children, a non-invasive arterial pressure recording could be obtained. Therefore, recordings from these 10 children were eligible for further analysis, resulting in 40 data points. The correlation coefficient between reBAP and IAP in tracking mean arterial pressure (MAP) changes was 0.98. reBAP followed changes in IAP with a mean bias for systolic, diastolic arterial pressure, and MAP of 0.0 mm Hg (sd 5.8), 0.1 (sd 2.8), and 0.19 (sd 2.7), respectively. The prototype device closely follows arterial pressure changes in children. However, in a considerable number of attempts, obtaining a signal was time-consuming or unsuccessful. This technique seems promising but requires further technical development.
引用
收藏
页码:493 / 500
页数:8
相关论文
共 32 条
[1]  
*AAMI, 2003, MAN EL AUT SPHYGM
[2]   Feasibility of Noninvasive continuous finger arterial blood pressure measurements in very young children, aged 0-4 years [J].
Andriessen, Peter ;
Schraa, Olaf ;
Van Den Bosch-Ruis, Wendy ;
Ten Harkel, Derk Jan ;
Settels, Jos J. ;
Oetomo, Sidarto Bambang ;
Blanco, Carlos E. .
PEDIATRIC RESEARCH, 2008, 63 (06) :691-696
[3]   Agreement between methods of measurement with multiple observations per individual [J].
Bland, J. Martin ;
Altman, Douglas G. .
JOURNAL OF BIOPHARMACEUTICAL STATISTICS, 2007, 17 (04) :571-582
[4]   COMPARING METHODS OF MEASUREMENT - WHY PLOTTING DIFFERENCE AGAINST STANDARD METHOD IS MISLEADING [J].
BLAND, JM ;
ALTMAN, DG .
LANCET, 1995, 346 (8982) :1085-1087
[5]   STATISTICAL METHODS FOR ASSESSING AGREEMENT BETWEEN TWO METHODS OF CLINICAL MEASUREMENT [J].
BLAND, JM ;
ALTMAN, DG .
LANCET, 1986, 1 (8476) :307-310
[6]   Reconstruction of brachial artery pressure from noninvasive finger pressure measurements [J].
Bos, WJW ;
vanGoudoever, J ;
vanMontfrans, GA ;
vandenMeiracker, AH ;
Wesseling, KH .
CIRCULATION, 1996, 94 (08) :1870-1875
[7]   Flow-related techniques for preoperative goal-directed fluid optimization [J].
Bundgaard-Nielsen, M. ;
Ruhnau, B. ;
Secher, N. H. ;
Kehlet, H. .
BRITISH JOURNAL OF ANAESTHESIA, 2007, 98 (01) :38-44
[8]   Factors influencing the accuracy of oscillometric blood pressure measurement in critically ill patients [J].
Bur, A ;
Herkner, H ;
Vlcek, M ;
Woisetschläger, C ;
Derhaschnig, U ;
Delle Karth, G ;
Laggner, AN ;
Hirschl, MM .
CRITICAL CARE MEDICINE, 2003, 31 (03) :793-799
[9]   Assessment of short-term blood pressure variability in anesthetized children: A comparative study between intraarterial and finger blood pressure [J].
Constant, I ;
Laude, D ;
Elghozi, JL ;
Murat, I .
JOURNAL OF CLINICAL MONITORING AND COMPUTING, 1999, 15 (3-4) :205-214
[10]   Feasibility of using finger arterial pressure in neonates [J].
Drouin, E ;
Gournay, V ;
Calamel, J ;
Mouzard, A ;
Roze, JC .
ARCHIVES OF DISEASE IN CHILDHOOD-FETAL AND NEONATAL EDITION, 1997, 77 (02) :F139-F140