Pre-planning of intramedullary nailing procedures: A methodology for predicting the position of the distal hole

被引:4
|
作者
Mortazavi, Javad [1 ]
Farahmand, Farzam [1 ,2 ]
Behzadipour, Saeed [1 ]
Yeganeh, Ali [3 ]
机构
[1] Sharif Univ Technol, Mech Engn Dept, Azadi Ave, Tehran, Iran
[2] Univ Tehran Med Sci, RCBTR, Tehran, Iran
[3] Iran Univ Med Sci, Dept Orthoped Surg, Tehran, Iran
基金
美国国家科学基金会;
关键词
Long bone fractures; Distal locking; Nail deformation; Analytical modeling; Computer assisted surgery; TARGETING DEVICE; AIMING DEVICE; FEMORAL NAILS; HUMAN FEMUR; LOCKING; SCREWS; NAVIGATION; INSERTION; EFFICACY; GUIDANCE;
D O I
10.1016/j.medengphy.2019.09.012
中图分类号
R318 [生物医学工程];
学科分类号
0831 ;
摘要
Inserting the distal locking screws is a challenging step of the intramedullary nailing procedures due to the nail deformation that makes the proximally mounted targeting systems ineffective. A pre-planning methodology is proposed, based on an analytical model of the nail-bone construct, to predict the nail deformation during surgery using orthogonal preoperative radiographs. Each of the femoral shaft and the nail was modeled as a curved tubular Euler-Bernoulli beam. The unknown positions and forces of the nail-bone interaction were found using a systematic trial and error approach, which minimized the total strain energy of the system while satisfying the force and geometrical constraints. The predictions of the model for the nail deformation were compared with the experimental results of five cadaver specimens in 15 test conditions. Relatively large displacements (up to 13 mm) were found for the distal hole in sagittal plane only. The model predictions were in close agreement with the experimental results, with a root mean square error of 1.2 mm. It was concluded that the proposed pre-planning methodology is promising for practical clinical use in intramedullary nailing operations, in order to provide the compensatory information that is required for tuning of proximally mounted targeting systems. (C) 2019 IPEM. Published by Elsevier Ltd. All rights reserved.
引用
收藏
页码:172 / 179
页数:8
相关论文
共 12 条
  • [1] Recovery of distal hole axis in intramedullary nail trajectory planning
    Zhu, Y
    Phillips, R
    Griffiths, JG
    Viant, W
    Mohsen, A
    Bielby, M
    PROCEEDINGS OF THE INSTITUTION OF MECHANICAL ENGINEERS PART H-JOURNAL OF ENGINEERING IN MEDICINE, 2002, 216 (H5) : 323 - 332
  • [2] Recovery of distal hole axis in IntraMedullary nail trajectory planning
    Zhu, Y
    Phillips, R
    Griffiths, JG
    Viant, W
    Mohsen, A
    Bielby, M
    2001 IEEE INTERNATIONAL CONFERENCE ON ROBOTICS AND AUTOMATION, VOLS I-IV, PROCEEDINGS, 2001, : 1561 - 1566
  • [3] Distal Tibial Intramedullary Nailing Using an Extraarticular, Lateral Parapatellar Approach in the Semiextended Position
    McAndrew, Christopher M.
    Ricci, William M.
    Miller, Anna N.
    Avery, Matthew C.
    JOURNAL OF ORTHOPAEDIC TRAUMA, 2018, 32 : S34 - S35
  • [4] A patient specific finite element simulation of intramedullary nailing to predict the displacement of the distal locking hole
    Mortazavi, Javad
    Farahmand, Farzam
    Behzadipour, Saeed
    Yeganeh, Ali
    Aghighi, Mohammad
    MEDICAL ENGINEERING & PHYSICS, 2018, 55 : 34 - 42
  • [5] Effect of Distal Interlocking Screw Number and Position After Intramedullary Nailing of Distal Tibial Fractures: A Biomechanical Study Simulating Immediate Weight-Bearing
    Chan, Daniel S.
    Nayak, Aniruddh N.
    Blaisdell, Greg
    James, Chris R.
    Denard, Antony
    Miles, Jeremy
    Santoni, Brandon G.
    JOURNAL OF ORTHOPAEDIC TRAUMA, 2015, 29 (02) : 98 - 104
  • [6] A comparison of free-hand method and electromagnetic navigation technique for the distal locking during intramedullary nailing procedures: a meta-analysis
    Zhao, Xiaoxu
    Fan, Yonggang
    Chen, Juwu
    ARCHIVES OF ORTHOPAEDIC AND TRAUMA SURGERY, 2021, 141 (01) : 45 - 53
  • [7] A comparison of free-hand method and electromagnetic navigation technique for the distal locking during intramedullary nailing procedures: a meta-analysis
    Xiaoxu Zhao
    Yonggang Fan
    Juwu Chen
    Archives of Orthopaedic and Trauma Surgery, 2021, 141 : 45 - 53
  • [8] Meta-analysis suggests that the electromagnetic technique is better than the free-hand method for the distal locking during intramedullary nailing procedures
    Yanbin Zhu
    Hengrui Chang
    Yiyang Yu
    Wei Chen
    Song Liu
    Yingze Zhang
    International Orthopaedics, 2017, 41 : 1041 - 1048
  • [9] Meta-analysis suggests that the electromagnetic technique is better than the free-hand method for the distal locking during intramedullary nailing procedures
    Zhu, Yanbin
    Chang, Hengrui
    Yu, Yiyang
    Chen, Wei
    Liu, Song
    Zhang, Yingze
    INTERNATIONAL ORTHOPAEDICS, 2017, 41 (05) : 1041 - 1048
  • [10] 3D visualization of myocardial substrate using Delayed Enhancement MRI for pre-planning and guidance of ablation procedures of ventricular tachycardia
    Rubio-Guivernau J.L.
    Perez-David E.
    Arenal Á.
    Bermejo J.
    Santos A.
    Ledesma-Carbayo M.J.