Generation of wedge-shaped dose distributions through dynamic multileaf collimator dose delivery

被引:7
作者
Zhu, Jingeng [1 ]
机构
[1] Provena St Joseph Med Ctr, Canc Care Dept, Joliet, IL 60435 USA
关键词
dynamic wedge; dynamic multileaf collimator; intensity modulation;
D O I
10.1120/jacmp.2025.25351
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
A new method of generating wedge- shaped dose distributions through dynamic multileaf collimator dose delivery rather than computer- controlled jaw motion is presented. The method starts with the calculation of a wedge- shaped beam profile for the desired wedge angle. The resultant wedge beam profile is then passed to the intensity- modulated radiation therapy ( IMRT) leaf sequence generation algorithm to create multileaf collimator ( MLC) segments for dose delivery. Wedge- shaped dose distributions are achieved through dynamic MLC dose delivery guided with the generated MLC segments. The method has been tested by generating wedge-shaped doses for a set of conventional wedge angles ( i. e., 15, 30, 45, and 60). Film dosimetry is used for dose distribution verification. For each dose delivery guided with MLC segments created for the indicated wedge angle, the desired wedge- shaped dose distributions are observed. It is concluded that the dynamic MLC can be used to implement dynamic wedges in the clinic. This technique is different from the virtual wedge or the dynamic wedge developed for a particular type of LINAC. The same method can be applied to any machine equipped with a MLC. Other advantages are that it can generate a wedge field at an arbitrary orientation as the omni wedge does, and it creates wedged and shaped fields using a MLC only.
引用
收藏
页码:37 / 45
页数:9
相关论文
共 12 条
[1]   DESIGN OF A UNIVERSAL WEDGE FILTER SYSTEM FOR A COBALT 60 UNIT [J].
ARON, BS ;
SCAPICCH.M .
AMERICAN JOURNAL OF ROENTGENOLOGY RADIUM THERAPY AND NUCLEAR MEDICINE, 1966, 96 (01) :70-&
[2]   X-RAY FIELD COMPENSATION WITH MULTILEAF COLLIMATORS [J].
BORTFELD, TR ;
KAHLER, DL ;
WALDRON, TJ ;
BOYER, AL .
INTERNATIONAL JOURNAL OF RADIATION ONCOLOGY BIOLOGY PHYSICS, 1994, 28 (03) :723-730
[3]  
BRAHME A, 1988, STRAHLENTHER ONKOL, V164, P343
[4]   THE GENERATION OF INTENSITY-MODULATED FIELDS FOR CONFORMAL RADIOTHERAPY BY DYNAMIC COLLIMATION [J].
CONVERY, DJ ;
ROSENBLOOM, ME .
PHYSICS IN MEDICINE AND BIOLOGY, 1992, 37 (06) :1359-1374
[5]  
*INT COMM RAD UN M, 1976, 24 NAT BUR STAND INT
[6]  
KIJEWSKI P K, 1978, Medical Physics (Woodbury), V5, P426, DOI 10.1118/1.594440
[7]   Multiple machine implementation of enhanced dynamic wedge [J].
Klein, EE ;
Gerber, R ;
Zhu, XR ;
Oehmke, F ;
Purdy, JA .
INTERNATIONAL JOURNAL OF RADIATION ONCOLOGY BIOLOGY PHYSICS, 1998, 40 (04) :977-985
[8]   A semi-automatic technique for measurement of arterial wall from black blood MRI [J].
Ladak, HM ;
Thomas, JB ;
Mitchell, JR ;
Rutt, BK ;
Steinman, DA .
MEDICAL PHYSICS, 2001, 28 (06) :1098-1107
[9]   DYNAMIC WEDGE FIELD TECHNIQUES THROUGH COMPUTER-CONTROLLED COLLIMATOR MOTION AND DOSE DELIVERY [J].
LEAVITT, DD ;
MARTIN, M ;
MOELLER, JH ;
LEE, WL .
MEDICAL PHYSICS, 1990, 17 (01) :87-91
[10]   The omni wedge: A method to produce wedged fields at arbitrary orientations [J].
Milliken, BD ;
Hamilton, RJ ;
Rubin, SJ .
MEDICAL PHYSICS, 1996, 23 (03) :337-342