Unskilled unawareness and the learning curve in robotic spine surgery

被引:96
作者
Schatlo, Bawarjan [1 ]
Martinez, Ramon [1 ]
Alaid, Awad [1 ]
von Eckardstein, Kajetan [1 ]
Akhavan-Sigari, Reza [1 ]
Hahn, Anina [1 ]
Stockhammer, Florian [1 ]
Rohde, Veit [1 ]
机构
[1] Univ Med Gottingen, Dept Neurosurg, Gottingen, Germany
关键词
Spine instrumentation; Pedicle screw; Robotic surgery; Spinal fusion; Computer-assisted surgery; Learning curve; PEDICLE SCREW PLACEMENT; LUMBAR SPINE; ACCURACY; NAVIGATION; FIXATION; GUIDANCE; SYSTEM;
D O I
10.1007/s00701-015-2535-0
中图分类号
R74 [神经病学与精神病学];
学科分类号
摘要
Robotic assistance for the placement of pedicle screws has been established as a safe technique. Nonetheless rare instances of screw misplacement have been reported.The aim of the present retrospective study is to assess whether experience and time affect the accuracy of screws placed with the help of the SpineAssist (TM) robot system. Postoperative computed tomography (CT) scans of 258 patients requiring thoracolumbar pedicle screw instrumentation from 2008 to 2013 were reviewed. Overall, 13 surgeons performed the surgeries. A pedicle breach of > 3 mm was graded as a misplacement. Surgeons were dichotomised into an early and experienced period in increments of five surgeries. In 258 surgeries, 1,265 pedicle screws were placed with the aid of the robot system. Overall, 1,217 screws (96.2 %) were graded as acceptable. When displayed by surgeon, the development of percent misplacement rates peaked between 5 and 25 surgeries in 12 of 13 surgeons. The overall misplacement rate in the first five surgeries was 2.4 % (6/245). The misplacement rate rose to 6.3 % between 11 and 15 surgeries (10/158; p = 0.20), and reached a significant peak between 16 and 20 surgeries with a rate of 7.1 % (8/112; p = 0.03). Afterwards, misplacement rates declined. A major peak in screw inaccuracies occurred between cases 10 and 20, and a second, smaller one at about 40 surgeries. One potential explanation could be a transition from decreased supervision (unskilled but aware) to increased confidence of a surgeon (unskilled but unaware) who adopts this new technique prior to mastering it (skilled). We therefore advocate ensuring competent supervision for new surgeons at least during the first 25 procedures of robotic spine surgery to optimise the accuracy of robot-assisted pedicle screws.
引用
收藏
页码:1819 / 1823
页数:5
相关论文
共 19 条
[1]   Learning curve of computer-assisted navigation system in spine surgery [J].
Bai Yu-shu ;
Zhang Ye ;
Chen Zi-qiang ;
Wang Chuan-feng ;
Zhao Ying-chuan ;
Shi Zhi-cai ;
Li Ming ;
Liu, Ka Po Gabriel .
CHINESE MEDICAL JOURNAL, 2010, 123 (21) :2989-2994
[2]   Clinical Acceptance and Accuracy Assessment of Spinal Implants Guided With SpineAssist Surgical Robot Retrospective Study [J].
Devito, Dennis P. ;
Kaplan, Leon ;
Dietl, Rupert ;
Pfeiffer, Michael ;
Horne, Dale ;
Silberstein, Boris ;
Hardenbrook, Mitchell ;
Kiriyanthan, George ;
Barzilay, Yair ;
Bruskin, Alexander ;
Sackerer, Dieter ;
Alexandrovsky, Vitali ;
Stueer, Carsten ;
Burger, Ralf ;
Maeurer, Johannes ;
Gordon, Donald G. ;
Schoenmayr, Robert ;
Friedlander, Alon ;
Knoller, Nachshon ;
Schmieder, Kirsten ;
Pechlivanis, Ioannis ;
Kim, In-Se ;
Meyer, Bernhard ;
Shoham, Moshe .
SPINE, 2010, 35 (24) :2109-2115
[3]   The use of pedicle-screw internal fixation for the operative treatment of spinal disorders [J].
Gaines, RW .
JOURNAL OF BONE AND JOINT SURGERY-AMERICAN VOLUME, 2000, 82A (10) :1458-1476
[4]   Clinically relevant complications related to pedicle screw placement in thoracolumbar surgery and their management: a literature review of 35,630 pedicle screws [J].
Gautschi, Oliver P. ;
Schatlo, Bawarjan ;
Schaller, Karl ;
Tessitore, Enrico .
NEUROSURGICAL FOCUS, 2011, 31 (04)
[5]   ACCURACY OF PEDICULAR SCREW PLACEMENT INVIVO [J].
GERTZBEIN, SD ;
ROBBINS, SE .
SPINE, 1990, 15 (01) :11-14
[6]   Robotic-assisted laparoscopic prostatectomy: What is the learning curve? [J].
Herrell, SD ;
Smith, JA .
UROLOGY, 2005, 66 (5A) :105-107
[7]   Intraoperative spinal navigation [J].
Holly, LT ;
Foley, KT .
SPINE, 2003, 28 (15) :S54-S61
[8]   What Is the Learning Curve for Robotic-assisted Pedicle Screw Placement in Spine Surgery? [J].
Hu, Xiaobang ;
Lieberman, Isador H. .
CLINICAL ORTHOPAEDICS AND RELATED RESEARCH, 2014, 472 (06) :1839-1844
[9]   Perioperative course and accuracy of screw positioning in conventional, open robotic-guided and percutaneous robotic-guided, pedicle screw placement [J].
Kantelhardt, Sven Rainer ;
Martinez, Ramon ;
Baerwinkel, Stefan ;
Burger, Ralf ;
Giese, Alf ;
Rohde, Veit .
EUROPEAN SPINE JOURNAL, 2011, 20 (06) :860-868
[10]   Monitoring the Quality of Robot-Assisted Pedicle Screw Fixation in the Lumbar Spine by Using a Cumulative Summation Test [J].
Kim, Ho-Joong ;
Lee, Seung Hoo ;
Chang, Bong-Soon ;
Lee, Choon-Ki ;
Lim, Teck Onn ;
Hoo, Ling Ping ;
Yi, Je-Min ;
Yeom, Jin S. .
SPINE, 2015, 40 (02) :87-94