Validation of the Omron M7 and Microlife 3BTO-A Blood Pressure Measuring Devices in Preeclampsia

被引:17
|
作者
Nouwen, Elisabeth [1 ]
Snijder, Marianne [1 ]
van Montfrans, Gert [2 ]
Wolf, Hans [1 ]
机构
[1] Univ Amsterdam, Dept Obstet & Gynaecol, Acad Med Ctr, NL-1100 DD Amsterdam, Netherlands
[2] Univ Amsterdam, Dept Vasc Med, Acad Med Ctr, NL-1100 DD Amsterdam, Netherlands
关键词
Blood pressure device; Preeclampsia; Validation; INTERNATIONAL-PROTOCOL; MONITORING DEVICE; PREGNANCY;
D O I
10.3109/10641955.2010.544799
中图分类号
R71 [妇产科学];
学科分类号
100211 ;
摘要
Objective. Validation of two automatic blood pressure devices (Microlife 3BTO-A, Microlife, Taipei, Taiwan and Omron M7, OMRON, Kyoto, Japan) in preeclampsia. Methods. Women (n = 34) admitted to the obstetric department of the Academic Medical Center and the Onze Lieve Vrouwe Gasthuis, Amsterdam, the Netherlands, because of preeclampsia were included. Intermittent auscultatory and oscillometric device measurements were performed to form three measurement pairs for each oscillometric device. Comparison was according to the International Protocol for validation of blood pressure measuring devices that was designed by the Working Group on Blood Pressure Monitoring of the European Society of Hypertension. Results. Both devices passed the first phase of the protocol. In the second phase, OMRON M7 passed for diastolic blood pressure, but failed for systolic blood pressure, whereas Microlife 3BTO-A failed for both. A difference between device and observers of more than 10 mmHg was observed in over 20% of systolic measurements in both devices and in 5% (OMRON M7) or 16% (Microlife 3BTO-A) of diastolic measurements. Conclusions. Although the use of Omron M7 and Microlife 3BTO-A is acceptable in pregnant women with normal or moderately elevated blood pressure, these devices cannot be recommended in women with severe preeclampsia.
引用
收藏
页码:131 / 139
页数:9
相关论文
共 50 条
  • [21] VALIDATION OF THREE DIFFERENT PROFESSIONAL DEVICES MEASURING OFFICE BLOOD PRESSURE (BP). THE OMRON BP10, THE OMRON HBP T105 AND THE PIC INDOLOR PROFESSIONAL CHECK
    Germano, G.
    Valente, L.
    Psimenos, A.
    Sarullo, F.
    Pecchioli, V.
    Asmar, R.
    JOURNAL OF HYPERTENSION, 2010, 28 : E84 - E84
  • [22] Validation of the Omron® M5-I device for blood pressure measurement according to the International validation Protocol
    El Assaad, M
    Topouchian, J
    Labaki, G
    Asmar, R
    JOURNAL OF HYPERTENSION, 2003, 21 : S23 - S23
  • [23] Comparison of Microlife BP A200 Plus and Omron M6 Blood Pressure Monitors to Detect Atrial Fibrillation in Hypertensive Patients
    Marazzi, Giuseppe
    Iellamo, Ferdinando
    Volterrani, Maurizio
    Lombardo, Mauro
    Pelliccia, Francesco
    Righi, Daniela
    Grieco, Fabrizia
    Cacciotti, Luca
    Iaia, Luigi
    Caminiti, Giuseppe
    Rosano, Giuseppe
    ADVANCES IN THERAPY, 2012, 29 (01) : 64 - 70
  • [24] Comparison of Microlife BP A200 Plus and Omron M6 Blood Pressure Monitors to Detect Atrial Fibrillation in Hypertensive Patients
    Giuseppe Marazzi
    Ferdinando Iellamo
    Maurizio Volterrani
    Mauro Lombardo
    Francesco Pelliccia
    Daniela Righi
    Fabrizia Grieco
    Luca Cacciotti
    Luigi Iaia
    Giuseppe Caminiti
    Giuseppe Rosano
    Advances in Therapy, 2012, 29 : 64 - 70
  • [25] Validation of the Noninvasive Assessment of Central Blood Pressure by the SphygmoCor and Omron Devices Against the Invasive Catheter Measurement
    Ding, Feng-Hua
    Fan, Wang-Xiang
    Zhang, Rui-Yan
    Zhang, Qi
    Li, Yan
    Wang, Ji-Guang
    AMERICAN JOURNAL OF HYPERTENSION, 2011, 24 (12) : 1306 - 1311
  • [26] Erratum to: Comparison of Microlife BP A200 Plus and Omron M6 Blood Pressure Monitors to Detect Atrial Fibrillation in Hypertensive Patients
    Giuseppe Marazzi
    Ferdinando Iellamo
    Maurizio Volterrani
    Mauro Lombardo
    Francesco Pelliccia
    Daniela Righi
    Fabrizia Grieco
    Luca Cacciotti
    Luigi Iaia
    Giuseppe Caminiti
    Giuseppe Rosano
    Advances in Therapy, 2014, 31 : 1317 - 1317
  • [27] BLOOD PRESSURE VARIABILITY DURING VALIDATION OF BLOOD PRESSURE MEASURING DEVICES: A NEGLECTED, BUT IMPORTANT FACTOR
    Braam, R.
    Verrij, E.
    Thien, T.
    Montfrans, G.
    Bos, W.
    JOURNAL OF HYPERTENSION, 2009, 27 : S4 - S4
  • [28] Validation of three professional devices measuring office blood pressure according to three different methods: the Omron BP10, the Omron HBP T105 and the Pic Indolor Professional
    Asmar, Roland
    Khabouth, Jose
    Mattar, Jimmy
    Pecchioli, Valerio
    Germano, Giuseppe
    JOURNAL OF HYPERTENSION, 2010, 28 (03) : 452 - 458
  • [29] Accuracy of the Omron RX-M, an automated blood pressure measuring device, measuring blood pressure at the wrist, according to a modified British Hypertension Society protocol
    Braam, RL
    Aslan, B
    Thien, T
    BLOOD PRESSURE MONITORING, 2004, 9 (01) : 25 - 30
  • [30] VALIDATION OF THE OMRON(R) M3500 BLOOD PRESSURE MEASURING DEVICE IN THE ADULT AND SPECIFIC POPULATIONS (OBESE AND CHILDREN) ACCORDING TO THE AAMI PROTOCOL
    Chahine, M.
    Assemaani, N.
    Sayed-Hasan, G.
    Chami-Chebbo, M.
    Salameh, P.
    Asmar, R.
    JOURNAL OF HYPERTENSION, 2015, 33 : E412 - E412