Validation of the Omron M7 and Microlife 3BTO-A Blood Pressure Measuring Devices in Preeclampsia

被引:17
|
作者
Nouwen, Elisabeth [1 ]
Snijder, Marianne [1 ]
van Montfrans, Gert [2 ]
Wolf, Hans [1 ]
机构
[1] Univ Amsterdam, Dept Obstet & Gynaecol, Acad Med Ctr, NL-1100 DD Amsterdam, Netherlands
[2] Univ Amsterdam, Dept Vasc Med, Acad Med Ctr, NL-1100 DD Amsterdam, Netherlands
关键词
Blood pressure device; Preeclampsia; Validation; INTERNATIONAL-PROTOCOL; MONITORING DEVICE; PREGNANCY;
D O I
10.3109/10641955.2010.544799
中图分类号
R71 [妇产科学];
学科分类号
100211 ;
摘要
Objective. Validation of two automatic blood pressure devices (Microlife 3BTO-A, Microlife, Taipei, Taiwan and Omron M7, OMRON, Kyoto, Japan) in preeclampsia. Methods. Women (n = 34) admitted to the obstetric department of the Academic Medical Center and the Onze Lieve Vrouwe Gasthuis, Amsterdam, the Netherlands, because of preeclampsia were included. Intermittent auscultatory and oscillometric device measurements were performed to form three measurement pairs for each oscillometric device. Comparison was according to the International Protocol for validation of blood pressure measuring devices that was designed by the Working Group on Blood Pressure Monitoring of the European Society of Hypertension. Results. Both devices passed the first phase of the protocol. In the second phase, OMRON M7 passed for diastolic blood pressure, but failed for systolic blood pressure, whereas Microlife 3BTO-A failed for both. A difference between device and observers of more than 10 mmHg was observed in over 20% of systolic measurements in both devices and in 5% (OMRON M7) or 16% (Microlife 3BTO-A) of diastolic measurements. Conclusions. Although the use of Omron M7 and Microlife 3BTO-A is acceptable in pregnant women with normal or moderately elevated blood pressure, these devices cannot be recommended in women with severe preeclampsia.
引用
收藏
页码:131 / 139
页数:9
相关论文
共 50 条
  • [1] A validated algorithm for blood pressure measurement in pregnancy andpre-eclampsia: Microlife 3BTO-A vs microlife 3AC1.
    De Greeff, A.
    Reggiori, F.
    Anthony, J.
    Shennan, A. H.
    HYPERTENSION IN PREGNANCY, 2006, 25 : 163 - 163
  • [2] An accurate automated blood pressure device for use in pregnancy and pre-eclampsia: the Microlife 3BTO-A
    Reinders, A
    Cuckson, AC
    Lee, JTM
    Shennan, AH
    BJOG-AN INTERNATIONAL JOURNAL OF OBSTETRICS AND GYNAECOLOGY, 2005, 112 (07) : 915 - 920
  • [3] Validation of Blood Pressure Measuring Devices in Women with Preeclampsia
    Nouwen, Ilse
    Post, Joris
    Boer, Kees
    Ganzevoort, Wessel
    Wolf, Hans
    HYPERTENSION IN PREGNANCY, 2008, 27 (04) : 516 - 516
  • [4] Validation of the Omron M7 (HEM-780-E) oscillometric blood pressure monitoring device according to the British Hypertension Society protocol
    Coleman, Andrew
    Steel, Stephen
    Freeman, Paul
    de Greeff, Annemarie
    Shennan, Andrew
    BLOOD PRESSURE MONITORING, 2008, 13 (01) : 49 - 54
  • [5] Validation of the OMRON M7 (HEM-780-E) blood pressure measuring device in a population requiring large cuff use according to the International Protocol of the European Society of Hypertension
    El Feghali, Ramzi N.
    Topouchian, Jirar A.
    Pannier, Bruno M.
    El Assaad, Hiba A.
    Asmar, Roland G.
    BLOOD PRESSURE MONITORING, 2007, 12 (03) : 173 - 178
  • [6] Validation of the microlife BP3BU1-5 wrist blood pressure measuring devices according to the International Protocol in adults
    Altunkan, S.
    Iliman, N.
    Altunkan, E.
    JOURNAL OF HYPERTENSION, 2007, 25 : S29 - S29
  • [7] Validation of the Microlife BP3BTO-A oscillometric blood pressure monitoring device according to a modified British Hypertension Society protocol
    Cuckson, AC
    Reinders, A
    Shabeeh, H
    Shennan, AH
    BLOOD PRESSURE MONITORING, 2002, 7 (06) : 319 - 324
  • [8] Validation of blood pressure measuring devices in France
    Vaisse, B
    Bobrie, G
    Raveau-Landon, C
    Asmar, R
    JOURNAL OF HYPERTENSION, 2004, 22 : S306 - S306
  • [9] VALIDATION OF BLOOD-PRESSURE MEASURING DEVICES
    OBRIEN, E
    PADFIELD, P
    BLAND, M
    ATKINS, N
    COATS, A
    PETRIE, J
    ALTMAN, D
    LITTLER, W
    DESWIET, M
    JOURNAL OF CLINICAL MONITORING, 1995, 11 (04): : 257 - 257
  • [10] Accuracy of the OMRON M4 automatic blood pressure measuring device
    JE Naschitz
    L Loewenstein
    R Lewis
    D Keren
    L Gaitini
    A Tamir
    D Yeshurun
    Journal of Human Hypertension, 2000, 14 : 423 - 427