Electrode migration after cochlear implant surgery: more common than expected?

被引:51
作者
Dietz, Aarno [1 ]
Wennstrom, Minna [1 ]
Lehtimaki, Antti [2 ]
Lopponen, Heikki [1 ,3 ]
Valtonen, Hannu [4 ]
机构
[1] Kuopio Univ Hosp, Dept Otorhinolaryngol, POB 100, Kuopio 70029, Finland
[2] Kuopio Univ Hosp, Dept Clin Radiol, Kuopio 70029, Finland
[3] Univ Eastern Finland, Inst Clin Med, Dept Otorhinolaryngol, Kuopio, Finland
[4] Univ Helsinki, Cent Hosp, Dept Otorhinolaryngol, Helsinki, Finland
关键词
Cochlear implants; Electrode migration; Cone-beam computed tomography; Impedances; REVISION; COMPLICATIONS; ADULTS;
D O I
10.1007/s00405-015-3716-4
中图分类号
R76 [耳鼻咽喉科学];
学科分类号
100213 ;
摘要
The overall complication rate of cochlear implant surgery is low and so-called electrode failures (electrode migration, misplacement, etc.,) account for only a minority of all complications. The aim of this study was to explore the prevalence of electrode migration as the cause for increased impedance values and non-auditory stimulation in the basal channels. Within the scope of a quality control process, the cochlear implant database of the Kuopio University Hospital (Finland) was reviewed. Patients with gradual elevation of impedance values and/or non-auditory stimulation of the basal electrode channels were re-examined and cone-beam computed tomography was administered. There were 162 cochlear implant recipients and 201 implanted devices registered in the database. A total of 18 patients (18 devices) were identified having significantly increased impedance values or non-auditory stimulation of the basal electrodes. Cone-beam computed tomography revealed extra-cochlear electrodes in 12 of these patients due to the migration of the electrode array. All extruded electrodes were lateral wall electrodes, i.e., straight electrode arrays (Cochlear CI422 and Med-El devices). The most common feature of electrode migration was the gradual increase of the impedance values in the basal electrodes, even though telemetry could also be unsuspicious. Electrode migration after cochlear implant surgery may be more common than previously reported. At surgery, special attention should be paid to the reliable fixation of the electrode array. This study underlines the importance of postoperative imaging after cochlear implant surgery.
引用
收藏
页码:1411 / 1418
页数:8
相关论文
共 25 条
[1]   Quality control after cochlear implant surgery by means of rotational tomography [J].
Aschendorff, A ;
Kubalek, T ;
Turowski, B ;
Zanella, T ;
Hochmuth, T ;
Schumacher, T ;
Klenzner, T ;
Laszig, R .
OTOLOGY & NEUROTOLOGY, 2005, 26 (01) :34-37
[2]  
Aschendorff A, 2011, Laryngorhinootologie, V90 Suppl 1, pS16, DOI 10.1055/s-0030-1270448
[3]   FIXATION OF THE ELECTRODE CABLE DURING COCHLEAR IMPLANTATION - THE SPLIT BRIDGE TECHNIQUE [J].
BALKANY, T ;
TELISCHI, FF .
LARYNGOSCOPE, 1995, 105 (02) :217-218
[4]   Incidence and Indications for Revision Cochlear Implant Surgery in Adults and Children [J].
Brown, Kevin D. ;
Connell, Sarah S. ;
Balkany, Thomas J. ;
Eshraghi, Adrien E. ;
Telischi, Fred F. ;
Angeli, Simon A. .
LARYNGOSCOPE, 2009, 119 (01) :152-157
[5]   Revision Cochlear Implant Surgery in Patients With Suspected Soft Failures [J].
Chung, Doug ;
Kim, Ana H. ;
Parisier, Simon ;
Linstrom, Christopher ;
Alexiades, George ;
Hoffman, Ronald ;
Kohan, Darius .
OTOLOGY & NEUROTOLOGY, 2010, 31 (08) :1194-1198
[6]  
Cohen N. L., 1995, Annals of Otology Rhinology and Laryngology, V104, P402
[7]   Electrode migration after cochlear implantation [J].
Connell, Sarah S. ;
Balkany, Thomas J. ;
Hodges, Annelle V. ;
Telischi, Fred F. ;
Angeli, Simon I. ;
Eshraghi, Adrien A. .
OTOLOGY & NEUROTOLOGY, 2008, 29 (02) :156-159
[8]   Revision cochlear implant surgery in children [J].
Cullen, Robert D. ;
Fayad, Jose N. ;
Luxford, William M. ;
Buchman, Craig A. .
OTOLOGY & NEUROTOLOGY, 2008, 29 (02) :214-220
[9]   Long-term evaluation of the effect of intracochlear steroid deposition on electrode impedance in cochlear implant patients [J].
De Ceulaer, G ;
Johnson, S ;
Yperman, M ;
Daemers, K ;
Offeciers, FE ;
O'Donoghue, GM ;
Govaerts, PJ .
OTOLOGY & NEUROTOLOGY, 2003, 24 (05) :769-774
[10]  
Dietz A, 2013, AUDIOL NEURO-OTOL, V18, P17