Moving sum of number of positive patient results as a quality control tool

被引:37
作者
Liu, Jiakai [2 ]
Tan, Chin Hon [2 ]
Badrick, Tony [3 ]
Loh, Tze Ping [1 ]
机构
[1] Natl Univ Singapore Hosp, Dept Lab Med, 5 Lower Kent Ridge Rd, Singapore 119074, Singapore
[2] Natl Univ Singapore, Dept Ind & Syst Engn, Singapore, Singapore
[3] Bond Univ, Fac Hlth Sci & Med, Robina, Qld, Australia
关键词
analytical error; average of normal; bias; erroneous; laboratory error; moving average; moving sum; quality control; shift; spurious; AVERAGE;
D O I
10.1515/cclm-2016-0950
中图分类号
R446 [实验室诊断]; R-33 [实验医学、医学实验];
学科分类号
1001 ;
摘要
Background: Recently, the total prostate-specific antigen (PSA) assay used in a laboratory had a positive bias of 0.03 mu g/L, which went undetected. Consequently, a number of post-prostatectomy patients with previously undetectable PSA concentrations (defined as <0.03 mu g/L in that laboratory) were being reported as having detectable PSA, which suggested poorer prognosis according to clinical guidelines. Methods: Through numerical simulations, we explored (1) how a small bias may evade the detection of routine quality control (QC) procedures with specific reference to the concentration of the QC material, (2) whether the use of 'average of normals' approach may detect such a small bias, and (3) describe the use of moving sum of number of patient results with detectable PSA as an adjunct QC procedure. Results: The lowest QC level (0.86 mu g/L) available from a commercial kit had poor probability (< 10%) of a bias of 0.03 mu g/L regardless of QC rule (i.e. 1 : 2S, 2 : 2S, 1 : 3S, 4 : 1S) used. The average number of patient results affected before error detection (ANPed) was high when using the average of normals approach due to the relatively wide control limits. By contrast, the ANPed was significantly lower for the moving sum of number of patient results with a detectable PSA approach. Conclusions: Laboratory practitioners should ensure their QC strategy can detect small but critical bias, and may require supplementation of ultra-low QC levels that are not covered by commercial kits with in-house preparations. The use of moving sum of number of patient results with a detectable result is a helpful adjunct QC tool.
引用
收藏
页码:1709 / 1714
页数:6
相关论文
共 14 条
  • [1] Diagnostic and prognostic utility of early measurement with high-sensitivity troponin T assay in patients presenting with chest pain
    Aldous, Sally J.
    Richards, Mark
    Cullen, Louise
    Troughton, Richard
    Than, Martin
    [J]. CANADIAN MEDICAL ASSOCIATION JOURNAL, 2012, 184 (05) : E260 - E268
  • [2] Failure of Current Laboratory Protocols to Detect Lot-to-Lot Reagent Differences: Findings and Possible Solutions
    Algeciras-Schimnich, Alicia
    Bruns, David E.
    Boyd, James C.
    Bryant, Sandra C.
    La Fortune, Kristin A.
    Grebe, Stefan K. G.
    [J]. CLINICAL CHEMISTRY, 2013, 59 (08) : 1187 - 1194
  • [3] Changing the paradigm of laboratory quality control through implementation of real-time test results monitoring: For patients by patients
    Fleming, James K.
    Katayev, Alexander
    [J]. CLINICAL BIOCHEMISTRY, 2015, 48 (7-8) : 508 - 513
  • [4] EAU Guidelines on Prostate Cancer. Part II: Treatment of Advanced, Relapsing, and Castration-Resistant Prostate Cancer
    Heidenreich, Axel
    Bastian, Patrick J.
    Bellmunt, Joaquim
    Bolla, Michel
    Joniau, Steven
    van der Kwast, Theodor
    Mason, Malcolm
    Matveev, Vsevolod
    Wiegel, Thomas
    Zattoni, Filiberto
    Mottet, Nicolas
    [J]. EUROPEAN UROLOGY, 2014, 65 (02) : 467 - 479
  • [5] HOFFMANN RG, 1965, AM J CLIN PATHOL, V43, P134
  • [6] Average of delta: a new quality control tool for clinical laboratories
    Jones, Graham R. D.
    [J]. ANNALS OF CLINICAL BIOCHEMISTRY, 2016, 53 (01) : 133 - 140
  • [7] Patient results and laboratory test accuracy
    Katayev, Alexander
    Fleming, James K.
    [J]. INTERNATIONAL JOURNAL OF HEALTH CARE QUALITY ASSURANCE, 2014, 27 (01) : 65 - +
  • [8] Prognostic Role of High-Sensitivity Cardiac Troponin T in Patients With Nonischemic Dilated Cardiomyopathy
    Kawahara, Chiho
    Tsutamoto, Takayoshi
    Nishiyama, Keizo
    Yamaji, Masayuki
    Sakai, Hiroshi
    Fujii, Masanori
    Yamamoto, Takashi
    Horie, Minoru
    [J]. CIRCULATION JOURNAL, 2011, 75 (03) : 656 - 661
  • [9] Laboratory quality control: Using patient data to assess analytical performance
    Kazmierczak, SC
    [J]. CLINICAL CHEMISTRY AND LABORATORY MEDICINE, 2003, 41 (05) : 617 - 627
  • [10] Verification of out-of-control situations detected by "average of normal" approach
    Liu, Jiakai
    Tan, Chin Hon
    Loh, Tze Ping
    Badrick, Tony
    [J]. CLINICAL BIOCHEMISTRY, 2016, 49 (16-17) : 1248 - 1253