Systematic set-up errors for IMRT in the head and neck region: effect on dose distribution

被引:45
作者
Samuelsson, A [1 ]
Mercke, C
Johansson, KA
机构
[1] Univ Gothenburg, Sahlgrens Univ Hosp, Dept Radiat Phys, SE-41345 Gothenburg, Sweden
[2] Univ Gothenburg, Sahlgrens Univ Hosp, Dept Oncol, SE-41345 Gothenburg, Sweden
关键词
set-up error; intensity modulated radiation therapy; head and neck; geometric uncertainty;
D O I
10.1016/S0167-8140(02)00389-4
中图分类号
R73 [肿瘤学];
学科分类号
100214 ;
摘要
Background and purpose: There is a general concern about intensity modulated radiation therapy (IMRT) treatments being more sensitive to patient positioning than conventional treatments. The aim of this study was to evaluate the International Commission on radiation units and measurements (ICRU) method for taking systematic set-up errors into account for IMRT treatments and to compare the effects on the dose distribution with the effects of conventional treatments. Material and methods: A planning margin to account for set-up errors was added to the clinical target volumes and to the spinal cords, for three head and neck patients, according to the ICRU. No margin was added to organs at risk with mainly parallel structure if they were situated adjacent to the target volume, for example, the parotid glands. The effects of set-up errors in six IMRT plans and three conventional plans were simulated in the planning system and analysed with physical dose parameters. Results and, conclusions: In general, the ICRU method of taking set-up errors into account works satisfactorily for IMRT treatments as well as for conventional treatments with no difference between the treatment techniques. The sensitivity to set-up errors regarding the target volume is dependent on the quality of the treatment plan, i.e. the part of the target covered with a dose > 95 and < 105% and the effect in the critical organs is dependent on the sharpness of the dose gradients outside the critical organ. However, the method makes it difficult to include organs at risk with mainly parallel structure if they are situated adjacent to the target volume. (C) 2002 Elsevier Science Ireland Ltd. All rights reserved.
引用
收藏
页码:303 / 311
页数:9
相关论文
共 28 条
  • [1] METHODS OF IMAGE-RECONSTRUCTION FROM PROJECTIONS APPLIED TO CONFORMATION RADIOTHERAPY
    BORTFELD, T
    BURKELBACH, J
    BOESECKE, R
    SCHLEGEL, W
    [J]. PHYSICS IN MEDICINE AND BIOLOGY, 1990, 35 (10) : 1423 - 1434
  • [2] OPTIMIZATION OF STATIONARY AND MOVING BEAM RADIATION-THERAPY TECHNIQUES
    BRAHME, A
    [J]. RADIOTHERAPY AND ONCOLOGY, 1988, 12 (02) : 129 - 140
  • [3] SOLUTION OF AN INTEGRAL-EQUATION ENCOUNTERED IN ROTATION THERAPY
    BRAHME, A
    ROOS, JE
    LAX, I
    [J]. PHYSICS IN MEDICINE AND BIOLOGY, 1982, 27 (10) : 1221 - 1229
  • [4] CHAO C, 2000, INT J CANC RAD ONCOL, V90, P92
  • [5] TREATMENT DELIVERY ACCURACY IN INTENSITY-MODULATED CONFORMAL RADIOTHERAPY
    CONVERY, DJ
    ROSENBLOOM, ME
    [J]. PHYSICS IN MEDICINE AND BIOLOGY, 1995, 40 (06) : 979 - 999
  • [6] THE GENERATION OF INTENSITY-MODULATED FIELDS FOR CONFORMAL RADIOTHERAPY BY DYNAMIC COLLIMATION
    CONVERY, DJ
    ROSENBLOOM, ME
    [J]. PHYSICS IN MEDICINE AND BIOLOGY, 1992, 37 (06) : 1359 - 1374
  • [7] Parotid gland sparing in patients undergoing bilateral head and neck irradiation: Techniques and early results
    Eisbruch, A
    Ship, JA
    Martel, MK
    TenHaken, RK
    Marsh, LH
    Wolf, GT
    Esclamado, RM
    Bradford, CR
    Terrell, JE
    Gebarski, SS
    Lichter, AS
    [J]. INTERNATIONAL JOURNAL OF RADIATION ONCOLOGY BIOLOGY PHYSICS, 1996, 36 (02): : 469 - 480
  • [8] Dose, volume, and function relationships in parotid salivary glands following conformal and intensity-modulated irradiation of head and neck cancer
    Eisbruch, A
    Ten Haken, RK
    Kim, HM
    Marsh, LH
    Ship, JA
    [J]. INTERNATIONAL JOURNAL OF RADIATION ONCOLOGY BIOLOGY PHYSICS, 1999, 45 (03): : 577 - 587
  • [9] Set-up verification using portal imaging; review of current clinical practice
    Hurkmans, CW
    Remeijer, P
    Lebesque, JV
    Mijnheer, BJ
    [J]. RADIOTHERAPY AND ONCOLOGY, 2001, 58 (02) : 105 - 120
  • [10] *ICRU, 2002, INT COMM RAD UN MEAS