Comparison of International Normalized Ratios provided by two point-of-care devices and laboratory-based venipuncture in a pharmacist-managed anticoagulation clinic

被引:27
作者
Donaldson, Mark [1 ,2 ]
Sullivan, John [1 ]
Norbeck, Amber [1 ]
机构
[1] Kalispell Reg Med Ctr, Kalispell, MT 59901 USA
[2] Univ Montana, Sch Pharm, Missoula, MT 59812 USA
关键词
Ambulatory care; Anticoagulants; Equipment; International normalized ratio; Methodology; Pharmacists; Tests; laboratory; Warfarin; ACCURACY; TIME; AGREEMENT; MONITOR; INRS;
D O I
10.2146/ajhp100096
中图分类号
R9 [药学];
学科分类号
1007 ;
摘要
Purpose. The International Normalized Ratios (INRs) measured by two point-of-care (POC) devices were compared with those obtained via laboratory-based venipuncture. Methods. In an outpatient, pharmacist-managed anticoagulation clinic, 52 patients receiving oral warfarin therapy had their INR values measured using two POC devices, i-STAT PT/INR (Abbott Laboratories, Abbott Park, IL) and CoaguChek XS Plus (Roche Diagnostics, Indianapolis, IN). At the same visit, one venous blood sample was collected from each patient for INR measurement by the reference laboratory's instrumentation (STAGO, Diagnostica-Stago, Parsippany, NJ). Accuracy was evaluated by calculating the absolute difference for each set of INR values. Clinical correlation was defined as an INR measurement obtained by the POC devices that would have resulted in the same therapeutic decision as the INR value measured by the reference laboratory's instrumentation. Results. Accuracy was superior with the CoaguChek XS Plus device. The absolute difference (mean +/- S.D.) in the INR measurements obtained using STAGO versus CoaguChek XS Plus was 0.28 +/- 0.31 (p < 0.0001). The absolute difference in INR values measured using STAGO and the i-STAT PT/INR device was 0.51 +/- 0.44 (p < 0.0001). For clinical correlation, 17 (33%) of 52 INR measurements with the CoaguChek XS Plus were sufficiently different from the STAGO-measured INR values to have resulted in a different therapeutic decision (p < 0.001), compared with 28 (54%) of 52 with the i-STAT PT/INR (p < 0.001). Conclusion. INR measurements generated by POC devices exhibited positive bias, compared with laboratory-based venipuncture, for INR values at the high end of the INR range.
引用
收藏
页码:1616 / 1622
页数:7
相关论文
共 27 条
[1]   Evolving models of warfarin management: Anticoagulation clinics, patient self-monitoring, and patient self-management [J].
Ansell, JE ;
Hughes, R .
AMERICAN HEART JOURNAL, 1996, 132 (05) :1095-1100
[2]  
Bland JM, 1999, STAT METHODS MED RES, V8, P135, DOI 10.1177/096228029900800204
[3]   STATISTICAL METHODS FOR ASSESSING AGREEMENT BETWEEN TWO METHODS OF CLINICAL MEASUREMENT [J].
BLAND, JM ;
ALTMAN, DG .
LANCET, 1986, 1 (8476) :307-310
[4]  
Bungard Tammy J, 2009, Open Med, V3, pe16
[5]   Accuracy, clinical correlation, and patient acceptance of two handheld prothrombin time monitoring devices in the ambulatory setting [J].
Chapman, DC ;
Stephens, MA ;
Hamann, GL ;
Bailey, LE ;
Dorko, CS .
ANNALS OF PHARMACOTHERAPY, 1999, 33 (7-8) :775-780
[6]   Comparison of an anticoagulation clinic with usual medical care -: Anticoagulation control, patient outcomes, and health care costs [J].
Chiquette, E ;
Amato, MG ;
Bussey, HI .
ARCHIVES OF INTERNAL MEDICINE, 1998, 158 (15) :1641-1647
[7]  
Deaton C, 1999, CIRCULATION, V100, P100
[8]   Point-of-care (POC) versus central laboratory instrumentation for monitoring oral anticoagulation [J].
Dorfman, DM ;
Goonan, EM ;
Boutilier, MK ;
Jarolim, P ;
Tanasijevic, M ;
Goldhaber, SZ .
VASCULAR MEDICINE, 2005, 10 (01) :23-27
[9]   Clinical impact of interlaboratory variation in International Normalized Ratio determinations [J].
Finck, KM ;
Doetkott, C ;
Miller, DR .
AMERICAN JOURNAL OF HEALTH-SYSTEM PHARMACY, 2001, 58 (08) :684-688
[10]   A comparison of point-of-gate instruments designed for monitoring oral anticoagulation with standard laboratory methods [J].
Gosselin, R ;
Owings, JT ;
White, RH ;
Hutchinson, R ;
Branch, J ;
Mahackian, K ;
Johnston, M ;
Larkin, EC .
THROMBOSIS AND HAEMOSTASIS, 2000, 83 (05) :698-703