A preliminary biomechanical evaluation in a simulated spinal fusion model

被引:12
作者
Ferrara, Lisa A.
Gordon, Illya
Coquillette, Madeline
Milks, Ryan
Fleischman, Aaron J.
Roy, Shuvo
Goel, Vijay K.
Benzel, Edward C.
机构
[1] Cleveland Clin Spine Inst, Spine Res Lab, Cleveland, OH USA
[2] Cleveland Clin Fdn, Dept Biomed Engn, BioMEMS Lab, Cleveland, OH USA
[3] Univ Toledo, Coll Engn & Med, Dept Bioengn, Toledo, OH USA
[4] Univ Toledo, Coll Engn & Med, Dept Orthopaed Surg, Toledo, OH USA
关键词
biomechanical testing; cervical spine; goat; polymethyhnethacrylate; spinal fusion;
D O I
10.3171/SPI-07/11/542
中图分类号
R74 [神经病学与精神病学];
学科分类号
摘要
Object. A preliminary in vitro biomechanical study was conducted to determine if the pressure at a bone graft-mortise interface and the load transmitted along a ventral cervical plate could be used as parameters to assess fusion status. Methods. An interbody bone graft and a ventral plate were placed at the C3-4 motion segment in six fresh cadaveric goat spines. Polymethylmethacrylate (PMMA) was used to simulate early bone fusion at the bone graft site. The loads along the plate and the simultaneous pressures induced at the graft-endplate interfaces were monitored during simulated stages of bone healing. Each specimen was nondestructively tested in compression loading while the pressures and loads at the graft site were recorded continuously. Each specimen was tested under five conditions (Disc, Graft, Plate, PMMA, and Removal). Results. The pressure at the interface of the bone graft and vertebral endplate did not change significantly with the addition of the ventral plate. The interface pressure and segmental stiffness did increase following PMMA augmentation of the bone graft (simulating an intermediate phase of bone fusion). The load transmitted along the ventral plate in compression increased after the addition of the bone graft, but decreased after PMMA augmentation. Thus, there was an increase in pressure at the graft-endplate interface and a decrease in load transferred along the ventral plate after the simulation of bone fusion. Upon removal of the ventral plate, the simulated fusion bore most of the axial load, thus explaining a further increase in graft site pressure. Conclusions. These observations support the notions of load sharing and the redistribution of loads occurring during and after bone graft incorporation. In the clinical setting, these parameters may be useful in the assessment of fusion after spine surgery. Although feasibility has been demonstrated in this preliminary study, further research is needed.
引用
收藏
页码:542 / 548
页数:7
相关论文
共 33 条
[1]   Adjacent segment motion after a simulated lumbar fusion in different sagittal alignments - A biomechanical analysis [J].
Akamaru, T ;
Kawahara, N ;
Yoon, ST ;
Minamide, A ;
Kim, KS ;
Tomita, K ;
Hutton, WC .
SPINE, 2003, 28 (14) :1560-1566
[2]   Micromachines in spine surgery [J].
Benzel, E ;
Ferrara, L ;
Roy, S ;
Fleischman, A .
SPINE, 2004, 29 (06) :601-606
[3]  
Benzel Edward C, 2002, Clin Neurosurg, V49, P209
[4]   Anterior cervical interbody fusion using autogeneic and allogeneic bone graft substrate: A prospective comparative analysis [J].
Bishop, RC ;
Moore, KA ;
Hadley, MN .
JOURNAL OF NEUROSURGERY, 1996, 85 (02) :206-210
[5]   CAN LUMBAR SPINE RADIOGRAPHS ACCURATELY DETERMINE FUSION IN POSTOPERATIVE-PATIENTS - CORRELATION OF ROUTINE RADIOGRAPHS WITH A 2ND SURGICAL LOOK AT LUMBAR FUSIONS [J].
BLUMENTHAL, SL ;
GILL, K .
SPINE, 1993, 18 (09) :1186-1189
[6]   CORRELATION OF RADIOLOGIC ASSESSMENT OF LUMBAR SPINE FUSIONS WITH SURGICAL EXPLORATION [J].
BRODSKY, AE ;
KOVALSKY, ES ;
KHALIL, MA .
SPINE, 1991, 16 (06) :S261-S265
[7]   Effects of short anterior lumbar interbody fusion on biomechanics of neighboring unfused segments [J].
Chow, DHK ;
Luk, KDK ;
Evans, JH ;
Leong, JCY .
SPINE, 1996, 21 (05) :549-555
[8]   COMPARISON OF IN-VIVO AND IN-VITRO ADJACENT SEGMENT MOTION AFTER LUMBAR FUSION [J].
DEKUTOSKI, MB ;
SCHENDEL, MJ ;
OGILVIE, JW ;
OLSEWSKI, JM ;
WALLACE, LJ ;
LEWIS, JL .
SPINE, 1994, 19 (15) :1745-1751
[9]   Anterior cervical plating reverses load transfer through multilevel strut-grafts [J].
DiAngelo, DJ ;
Foley, KT ;
Vossel, KA ;
Rampersaud, YR ;
Jansen, TH .
SPINE, 2000, 25 (07) :783-795
[10]   An in vivo biocompatibility assessment of MEMS materials for spinal fusion monitoring [J].
Ferrara, LA ;
Fleischman, AJ ;
Togawa, D ;
Bauer, TW ;
Benzel, EC ;
Roy, S .
BIOMEDICAL MICRODEVICES, 2003, 5 (04) :297-302