A new laboratory EUV reflectometer for large optics using a laser plasma source

被引:21
作者
van Loyen, L [1 ]
Böttger, T [1 ]
Braun, S [1 ]
Mai, H [1 ]
Leson, A [1 ]
Scholze, F [1 ]
Tümmler, J [1 ]
Ulm, G [1 ]
Legall, H [1 ]
Nickles, PV [1 ]
Sandner, W [1 ]
Stiel, H [1 ]
Rempel, C [1 ]
Schulze, M [1 ]
Brutscher, J [1 ]
Macco, F [1 ]
Müllender, S [1 ]
机构
[1] IWS Dresden, Fraunhofer Inst Mat & Beam Technol, D-01277 Dresden, Germany
来源
METROLOGY, INSPECTION, AND PROCESS CONTROL FOR MICROLITHOGRAPHY XVII, PTS 1 AND 2 | 2003年 / 5038卷
关键词
extreme ultraviolet; metrology; lithography; at-wavelength characterization; reflectometry;
D O I
10.1117/12.485042
中图分类号
TH7 [仪器、仪表];
学科分类号
0804 ; 080401 ; 081102 ;
摘要
The quality assurance for the production of optical components for Extreme UltraViolet Lithography (EUVL) strongly requires at-wavelength metrology. Presently, at-wavelength characterizations of mirrors and masks are done using the synchrotron radiation of electron storage rings, e.g. BESSY II. For the production process of EUV optics, however, the immediate access to metrology tools is necessary and the availability of laboratory devices is mandatory. Within the last few years a stand alone laboratory EUV reflectometer for large samples has been developed. It consists of a laser produced plasma (LPP) radiation source, a monochromator and a large goniometer system. The manipulation system of the reflectometer can handle samples with diameters of up to 500 mm, thicknesses of up to 200 mm and weights of up to 30 kg. The wavelength can be varied from 10 run to 16 nm. The spot size on the sample surface is about empty set 2 mm. The angle of incidence can be varied from 3degrees to 60degrees. In this paper, we describe the laboratory reflectometer in detail and discuss the achieved performance. First measurements of 4" mirrors are presented and discussed in comparison to the results obtained at the PTB soft X-ray radiometry beamline at BESSY II.
引用
收藏
页码:12 / 21
页数:10
相关论文
共 8 条
[1]  
BRAUN S, 2003, P SOC PHOTO-OPT INS, P5037
[2]  
Gullikson E M, 1992, J Xray Sci Technol, V3, P283, DOI 10.3233/XST-1992-3402
[3]  
HEINEMANN F, 1991, THESIS U BIELEFELD G
[4]   A laser produced plasma based reflectometer for EUV metrology [J].
Mrowka, S ;
Underwood, JH ;
Gullikson, E ;
Batson, P .
EMERGING LITHOGRAPHIC TECHNOLOGIES IV, 2000, 3997 :819-822
[5]   High-accuracy radiometry in the EUV range at the PTB soft x-ray beamline [J].
Scholze, F ;
Tümmler, J ;
Ulm, G .
METROLOGIA, 2003, 40 (01) :S224-S228
[6]  
SCHOLZE F, 2003, P SOC PHOTO-OPT INS, P5037
[7]  
Tümmler J, 2002, P SOC PHOTO-OPT INS, V4688, P338, DOI 10.1117/12.472307
[8]  
Windt D. L., 1991, P SOC PHOTO-OPT INS, V1547, P144