Whole body personnel monitoring via ionization detection

被引:1
作者
Koster, JE [1 ]
Bounds, JA [1 ]
Kerr, PL [1 ]
Steadman, PA [1 ]
Whitley, CR [1 ]
机构
[1] Univ Calif Los Alamos Natl Lab, Los Alamos, NM 87545 USA
关键词
D O I
10.1109/23.682691
中图分类号
TM [电工技术]; TN [电子技术、通信技术];
学科分类号
0808 ; 0809 ;
摘要
A project between Fernald EMP and LANL is to field a monitor for the detection of alpha-emitting contamination on a human body. Traditional personnel monitoring for alpha emitters involves either frisking with a probe or pressing against large detectors in order to overcome the short range of alpha particles. These methods have a low alpha collection efficiency, and can miss contamination on; less accessible surfaces. We have investigated the sensitivity and practicality of measuring the entire subject simultaneously using the technique of ionization monitoring. The goal is to create a booth that personnel step into quickly during egress from radiological facilities. The detection technique relies on a breeze of air passing over the subject. Alpha emission produces copious ions In the ambient air which are transported by the air current to an ion collector, resulting in a small electrical current proportional to the amount of contamination. Results indicate a conservative sensitivity of 3000 disintegrations per minute localized to one of five areas of the body in a measurement lasting less than 2 minutes.