A study of MDL improvement for the in vivo measurement of lead in bone

被引:26
作者
Nie, HL [1 ]
Chettle, D
Stronach, I
Arnold, M
Huang, SB
McNeill, F
O'Meara, J
机构
[1] McMaster Univ, Med Phys & Appl Radiat Sci Unit, Hamilton, ON L8S 4K1, Canada
[2] Queen Elizabeth Hosp, Dept Med Phys, Birmingham, W Midlands, England
[3] Weber State Univ, Dept Phys, Ogden, UT 84408 USA
[4] Fudan Univ, Sch Med, Shanghai, Peoples R China
[5] Univ Guelph, Dept Phys, Guelph, ON N1G 2W1, Canada
关键词
MDL; lead measurement; cloverleaf system;
D O I
10.1016/S0168-583X(03)01675-6
中图分类号
TH7 [仪器、仪表];
学科分类号
0804 ; 080401 ; 081102 ;
摘要
Cd-109 induced K XRF has been used for in vivo lead measurement for about two decades. Recently, the need for improvement of this system has been emphasized due to an increased understanding of low-level lead exposure. The conventional XRF bone-lead measurement system includes a 25 mm radius HPGe detector and one set of electronics. In this work, a cloverleaf detector system is investigated. This system consists of four 8 mm radius detectors and four sets of electronics. We measured bare plaster of Paris phantoms and phantoms in a soft tissue equivalent leg phantom to represent in vivo measurement. A Monte Carlo simulation for XRF measurement was also used to simulate this system. We compared both the experimental results and the simulation and found that the minimum detectable limit (MDL) is greatly improved by using the cloverleaf system and a stronger source. The effect of geometry is also discussed. An overall MDL ratio of about 0.3 (experimental value 0.278 +/- 0.016 and simulation value 0.273) is obtained by using the cloverleaf system compared to the conventional system for in vivo measurement, which means a decrease of MDL from about 6-10 to about 2-3 mug/(g bone mineral). (C) 2003 Elsevier B.V. All rights reserved.
引用
收藏
页码:579 / 583
页数:5
相关论文
共 6 条
  • [1] AHLGREN L, 1976, J WORK ENV HLTH, V2, P82
  • [2] Bevington R., 1969, DATA REDUCTION ERROR
  • [3] Monte Carlo simulation of source-excited in vivo x-ray fluorescence measurements of heavy metals
    O'Meara, JM
    Chettle, DR
    McNeill, FE
    Prestwich, WV
    Svensson, CE
    [J]. PHYSICS IN MEDICINE AND BIOLOGY, 1998, 43 (06) : 1413 - 1428
  • [4] Improving the in vivo X-ray fluorescence (XRF) measurement of renal mercury
    O'Meara, JM
    Börjesson, J
    Chettle, DR
    [J]. APPLIED RADIATION AND ISOTOPES, 2000, 53 (4-5) : 639 - 646
  • [5] STRONACH IM, 2000, PHYS CANADA MAY, P62
  • [6] IN-VIVO X-RAY-FLUORESCENCE OF LEAD IN BONE - REVIEW AND CURRENT ISSUES
    TODD, AC
    CHETTLE, DR
    [J]. ENVIRONMENTAL HEALTH PERSPECTIVES, 1994, 102 (02) : 172 - 177