MR-guided focused-ultrasound ablation system: Determination of precision and accuracy in preparation for clinical trials

被引:4
作者
Kendell, KR [1 ]
Wu, T [1 ]
Felmlee, JP [1 ]
Lewis, BD [1 ]
Ehman, RL [1 ]
机构
[1] Mayo Clin & Mayo Fdn, Dept Radiol, MR Res Lab, Rochester, MN 55905 USA
关键词
magnetic resonance (MR); guidance; ultrasound; (US); technology; therapeutic;
D O I
10.1148/radiology.209.3.9844687
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
The authors analyzed the accuracy and precision of focal-spot placement with a magnetic resonance-guided, focused-ultrasound system. Average absolute accuracy errors ranged from 0.2 to 1.0 mm, and average absolute individual precision errors ranged from 0.2 to 0.3 mm. To prevent damage to vital structures, single sonications and sonication grids should be placed beyond approximately 2 and 1 mm, respectively.
引用
收藏
页码:856 / 861
页数:6
相关论文
共 24 条
[1]  
BUROV AK, 1956, DOKL AKAD NAUK SSSR+, V106, P445
[2]  
BUROV AK, 1956, DOKL AKAD NAUK SSSR+, V106, P239
[3]   FOCUSED US SYSTEM FOR MR IMAGING-GUIDED TUMOR ABLATION [J].
CLINE, HE ;
HYNYNEN, K ;
WATKINS, RD ;
ADAMS, WJ ;
SCHENCK, JF ;
ETTINGER, RH ;
FREUND, WR ;
VETRO, JP ;
JOLESZ, FA .
RADIOLOGY, 1995, 194 (03) :731-737
[4]   MR-GUIDED FOCUSED ULTRASOUND SURGERY [J].
CLINE, HE ;
SCHENCK, JF ;
HYNYNEN, K ;
WATKINS, RD ;
SOUZA, SP ;
JOLESZ, FA .
JOURNAL OF COMPUTER ASSISTED TOMOGRAPHY, 1992, 16 (06) :956-965
[5]  
COLEMAN DJ, 1985, OPHTHALMOLOGY, V92, P339
[6]   THE EFFECT OF VARIOUS PHYSICAL PARAMETERS ON THE SIZE AND SHAPE OF NECROSED TISSUE VOLUME DURING ULTRASOUND SURGERY [J].
DAMIANOU, C ;
HYNYNEN, K .
JOURNAL OF THE ACOUSTICAL SOCIETY OF AMERICA, 1994, 95 (03) :1641-1649
[7]   FOCAL SPACING AND NEAR-FIELD HEATING DURING PULSED HIGH-TEMPERATURE ULTRASOUND THERAPY [J].
DAMIANOU, C ;
HYNYNEN, K .
ULTRASOUND IN MEDICINE AND BIOLOGY, 1993, 19 (09) :777-787
[8]   THE EFFECT OF WAVE REFLECTION AND REFRACTION AT SOFT-TISSUE INTERFACES DURING ULTRASOUND HYPERTHERMIA TREATMENTS [J].
FAN, XB ;
HYNYNEN, K .
JOURNAL OF THE ACOUSTICAL SOCIETY OF AMERICA, 1992, 91 (03) :1727-1736
[9]  
FOSTER RS, 1993, EUR UROL, V23, P29
[10]  
FRY F J, 1978, Ultrasound in Medicine and Biology, V4, P337, DOI 10.1016/0301-5629(78)90022-4