A Time-Frequency Approach for Cerebral Embolic Load Monitoring

被引:8
作者
Imaduddin, Syed M. [1 ,2 ]
LaRovere, Kerri L. [3 ]
Kussman, Barry D. [4 ]
Heldt, Thomas [1 ,2 ]
机构
[1] MIT, Dept Elect Engn & Comp Sci, Inst Med Engn & Sci, Cambridge, MA 02139 USA
[2] MIT, Res Lab Elect, 77 Massachusetts Ave, Cambridge, MA 02139 USA
[3] Boston Childrens Hosp, Dept Neurol, Boston, MA USA
[4] Boston Childrens Hosp, Dept Anesthesiol Crit Care & Pain Med, Boston, MA USA
关键词
Doppler effect; Ultrasonic imaging; Microsoft Windows; Catheterization; Monitoring; Training; Testing; Emboli; patient monitoring; stroke; time-frequency analysis; transcranial ultrasound; VENTRICULAR ASSIST DEVICE; TRANSCRANIAL DOPPLER; MICROEMBOLI; SUPPORT;
D O I
10.1109/TBME.2019.2927709
中图分类号
R318 [生物医学工程];
学科分类号
0831 ;
摘要
Objective: To enable reliable cerebral embolic load monitoring from high-intensity transient signals (HITS) recorded with single-channel transcranial Doppler (TCD) ultrasound. Methods: We propose a HITS detection and characterization method using a weighted-frequency Fourier linear combiner that estimates baseline Doppler signal power. An adaptive threshold is determined by examining the Doppler signal power variance about the baseline estimate, and HITS are extracted if their Doppler power exceeds this threshold. As signatures from multiple emboli may be superimposed, we analyze the detected HITS in the time-frequency (TF) domain to segment the signals into individual emboli. A logistic regression classification approach is employed to classify HITS into emboli or artifacts. Data were collected using a commercial TCD device with emboli-detection capabilities from 12 children undergoing mechanical circulatory support or cardiac catheterization. A subset of 696 HITS were reviewed, annotated, and split into training and testing sets for developing and evaluating the HITS classification algorithm. Results: The classifier yielded 98% and 96% sensitivity for 100% specificity on training and testing data, respectively. The TF approach decomposed 38% of candidate embolic signals into two or more embolic events that ultimately account for 69% of the overall embolic counts. Our processing pipeline resulted in highly accurate emboli identification and produced emboli counts that were lower (by a median of 64%) compared to the commercial ultrasound system's estimates. Significance: Using only single-channel, single-frequency Doppler ultrasound, the proposed method enables sensitive detection and segmentation of embolic signatures. Our approach paves the way toward accurate real-time cerebral emboli monitoring.
引用
收藏
页码:1007 / 1018
页数:12
相关论文
共 41 条
  • [21] Lounissen T., 2017, ULTRASCHALL MED, V39, P198
  • [22] Lubbers J, 1977, Ultrasound Med Biol, V2, P301, DOI 10.1016/0301-5629(77)90031-X
  • [23] Cerebral microemboli detected by transcranial doppler in patients treated with extracorporeal membrane oxygenation
    Marinoni, M.
    Migliaccio, M. L.
    Trapani, S.
    Bonizzoli, M.
    Gucci, L.
    Cianchi, G.
    Gallerini, A.
    Buoninsegni, L. Tadini
    Cramaro, A.
    Valente, S.
    Chiostri, M.
    Peris, A.
    [J]. ACTA ANAESTHESIOLOGICA SCANDINAVICA, 2016, 60 (07) : 934 - 944
  • [24] Interrater reliability: the kappa statistic
    McHugh, Mary L.
    [J]. BIOCHEMIA MEDICA, 2012, 22 (03) : 276 - 282
  • [25] PULSE DOPPLER ULTRASOUND DETECTION, CHARACTERIZATION AND SIZE ESTIMATION OF EMBOLI IN FLOWING BLOOD
    MOEHRING, MA
    KLEPPER, JR
    [J]. IEEE TRANSACTIONS ON BIOMEDICAL ENGINEERING, 1994, 41 (01) : 35 - 44
  • [26] Power M-mode Doppler (PMD) for observing cerebral blood flow and tracking emboli
    Moehring, MA
    Spencer, MP
    [J]. ULTRASOUND IN MEDICINE AND BIOLOGY, 2002, 28 (01) : 49 - 57
  • [27] Neurologic Complications of Extracorporeal Membrane Oxygenation
    Nasr, Deena M.
    Rabinstein, Alejandro A.
    [J]. JOURNAL OF CLINICAL NEUROLOGY, 2015, 11 (04): : 383 - 389
  • [28] Oppenheim A.V., 2002, Discrete-time signal processing, V2nd
  • [29] THRESHOLD SELECTION METHOD FROM GRAY-LEVEL HISTOGRAMS
    OTSU, N
    [J]. IEEE TRANSACTIONS ON SYSTEMS MAN AND CYBERNETICS, 1979, 9 (01): : 62 - 66
  • [30] THE IMPACT OF MICROEMBOLI DURING CARDIOPULMONARY BYPASS ON NEUROPSYCHOLOGICAL FUNCTIONING
    PUGSLEY, W
    KLINGER, L
    PASCHALIS, C
    TREASURE, T
    HARRISON, M
    NEWMAN, S
    [J]. STROKE, 1994, 25 (07) : 1393 - 1399