Cortical auditory evoked potentials in cochlear implant patients

被引:0
作者
Hoth, S [1 ]
机构
[1] Univ Heidelberg, HNO Klin, D-69120 Heidelberg, Germany
关键词
cochlear implant; loudness coding; auditory evoked potentials; objective audiometry; hearing threshold;
D O I
10.1007/s001060050304
中图分类号
R76 [耳鼻咽喉科学];
学科分类号
100213 ;
摘要
The current adjustment of coch lear implant (CI) speech processors is based on a knowledge of the lower and upper limits (T- and C-levels) for electrical stimulus currents. These data are usually acquired from subjective classifications of individual patients. In cases with non-reliable patient responses, objective methods are necessary. Especially for the estimation of correct T-levels, auditory evoked potentials (AEP) can be applied, since they allow the determination of response thresholds in a frequency-specific manner. By determining the AEP of different latencies, late cortical responses can be registered almost without artifact contamination. These patients have been examined in 20 patients provided with 22- or 8-channel CI-systems (Nucleus(R) or Med-El(R) implants). In all cases, clear responses and clearly discernible threshold transitions could be detected. By making use of acoustical stimulation in a free sound field,subjective hearing threshold and the T-levels of electrical stimulation could be verified. Since late responses are generated in the primary auditory cortex, their assessment permits a nearly integral functional control of the aided hearing system. To date, no problems have occured from maturation or attentional effects in either pediatric or adult patients. The applicability in very young children remains to be explored.
引用
收藏
页码:739 / 747
页数:9
相关论文
共 31 条
[1]   ELECTRICALLY EVOKED AUDITORY BRAIN-STEM RESPONSE - GROWTH OF RESPONSE WITH CURRENT LEVEL [J].
ABBAS, PJ ;
BROWN, CJ .
HEARING RESEARCH, 1991, 51 (01) :123-138
[2]   CHARACTERISTICS OF ELECTRICALLY EVOKED AUDITORY BRAIN-STEM RESPONSES ELICITED WITH THE NUCLEUS 22-ELECTRODE INTRACOCHLEAR IMPLANT [J].
ALLUM, JHJ ;
SHALLOP, JK ;
HOTZ, M ;
PFALTZ, CR .
SCANDINAVIAN AUDIOLOGY, 1990, 19 (04) :263-267
[3]   RELIABILITY AND PREDICTIVE VALUE OF THE ELECTRICALLY-EVOKED AUDITORY BRAIN-STEM RESPONSE [J].
AUBERT, LR ;
CLARKE, GP .
BRITISH JOURNAL OF AUDIOLOGY, 1994, 28 (03) :121-124
[4]   ELECTRICALLY EVOKED WHOLE-NERVE ACTION-POTENTIALS - DATA FROM HUMAN COCHLEAR IMPLANT USERS [J].
BROWN, CJ ;
ABBAS, PJ ;
GANTZ, B .
JOURNAL OF THE ACOUSTICAL SOCIETY OF AMERICA, 1990, 88 (03) :1385-1391
[5]   INTRAOPERATIVE AND POSTOPERATIVE ELECTRICALLY-EVOKED AUDITORY BRAIN-STEM RESPONSES IN NUCLEUS COCHLEAR IMPLANT USERS - IMPLICATIONS FOR THE FITTING PROCESS [J].
BROWN, CJ ;
ABBAS, PJ ;
FRYAUFBERTSCHY, H ;
KELSAY, D ;
GANTZ, BJ .
EAR AND HEARING, 1994, 15 (02) :168-176
[6]  
GAME C J A, 1990, British Journal of Audiology, V24, P145, DOI 10.3109/03005369009076548
[7]   MULTIVARIATE PREDICTORS OF AUDIOLOGICAL SUCCESS WITH MULTICHANNEL COCHLEAR IMPLANTS [J].
GANTZ, BJ ;
WOODWORTH, GG ;
KNUTSON, JF ;
ABBAS, PJ ;
TYLER, RS .
ANNALS OF OTOLOGY RHINOLOGY AND LARYNGOLOGY, 1993, 102 (12) :909-916
[8]  
GERHARDT HJ, 1986, HNO PRAXIS, V11, P125
[9]  
Groenen P, 1997, AUDIOLOGY, V36, P83
[10]  
HOKE M, 1989, AUDIOLOGY, V28, P152