Injectable hydrogel microbeads for fluorescence-based in vivo continuous glucose monitoring

被引:247
作者
Shibata, Hideaki [1 ,2 ,3 ]
Heo, Yun Jung [1 ,2 ]
Okitsu, Teru [1 ,4 ]
Matsunaga, Yukiko [1 ,2 ]
Kawanishi, Tetsuro [1 ,3 ]
Takeuchi, Shoji [1 ,2 ]
机构
[1] Univ Tokyo, Life Bio Electromech Autonomous Nano Syst BEANS C, BEANS Project, Meguro Ku, Tokyo 1538505, Japan
[2] Univ Tokyo, Inst Ind Sci, Meguro Ku, Tokyo 1538505, Japan
[3] TERUMO Co R&D Headquarters, Nakai, Kanagawa 2590151, Japan
[4] Kyoto Univ Hosp, Transplant Unit, Sakyo Ku, Kyoto 6068507, Japan
关键词
fluorescent hydrogel; microfluidics; diboronic acid; continuous glucose monitoring; diabetes mellitus; BORONIC ACID; CONCANAVALIN-A; BIOSENSORS; SENSOR;
D O I
10.1073/pnas.1006911107
中图分类号
O [数理科学和化学]; P [天文学、地球科学]; Q [生物科学]; N [自然科学总论];
学科分类号
07 ; 0710 ; 09 ;
摘要
Fluorescent microbeads hold great promise for in vivo continuous glucose monitoring with wireless transdermal transmission and long-lasting activity. The full potential of fluorescent microbeads has yet to be realized due to insufficient intensity for transdermal transmission and material toxicity. This paper illustrates the highly-sensitive, biostable, long-lasting, and injectable fluorescent microbeads for in vivo continuous glucose monitoring. We synthesized a fluorescent monomer composed of glucose-recognition sites, a fluorogenic site, spacers, and polymerization sites. The spacers are designed to be long and hydrophilic for increasing opportunities to bind glucose molecules; consequently, the fluorescent monomers enable high-intensive responsiveness to glucose. We then fabricated injectable-sized fluorescent polyacrylamide hydrogel beads with high uniformity and high throughput. We found that our fluorescent beads provide sufficient intensity to transdermally monitor glucose concentrations in vivo. The fluorescence intensity successfully traced the blood glucose concentration fluctuation, indicating our method has potential uses in highly-sensitive and minimally invasive continuous blood glucose monitoring.
引用
收藏
页码:17894 / 17898
页数:5
相关论文
共 33 条
[1]   Biosensors for in vivo glucose measurement: can we cross the experimental stage [J].
Abel, PU ;
von Woedtke, T .
BIOSENSORS & BIOELECTRONICS, 2002, 17 (11-12) :1059-1070
[2]   Formation of dispersions using "flow focusing" in microchannels [J].
Anna, SL ;
Bontoux, N ;
Stone, HA .
APPLIED PHYSICS LETTERS, 2003, 82 (03) :364-366
[3]   Interstitial glucose concentration and glycemia: implications for continuous subcutaneous glucose monitoring [J].
Aussedat, B ;
Dupire-Angel, M ;
Gifford, R ;
Klein, JC ;
Wilson, GS ;
Reach, G .
AMERICAN JOURNAL OF PHYSIOLOGY-ENDOCRINOLOGY AND METABOLISM, 2000, 278 (04) :E716-E728
[4]   Concanavalin A for in vivo glucose sensing: A biotoxicity review [J].
Ballerstadt, Ralph ;
Evans, Colton ;
McNichols, Roger ;
Gowda, Ashok .
BIOSENSORS & BIOELECTRONICS, 2006, 22 (02) :275-284
[5]  
BANDERAS LM, 2005, SMALL, V1, P2
[6]   Glucose measurement in patients with diabetes mellitus with dermal interstitial fluid [J].
Bantle, JP ;
Thomas, W .
JOURNAL OF LABORATORY AND CLINICAL MEDICINE, 1997, 130 (04) :436-441
[7]   Biochemistry and molecular cell biology of diabetic complications [J].
Brownlee, M .
NATURE, 2001, 414 (6865) :813-820
[8]   Functional and molecular defects of pancreatic islets in human type 2 diabetes [J].
Del Guerra, S ;
Lupi, R ;
Marselli, L ;
Masini, M ;
Bugliani, M ;
Sbrana, S ;
Torri, S ;
Pollera, M ;
Boggi, U ;
Mosca, F ;
Del Prato, S ;
Marchetti, P .
DIABETES, 2005, 54 (03) :727-735
[9]   Polymer vesicles [J].
Discher, DE ;
Eisenberg, A .
SCIENCE, 2002, 297 (5583) :967-973
[10]   Operation of a miniature redox hydrogel-based pyruvate sensor in undiluted deoxygenated calf serum [J].
Gajovic, N ;
Binyamin, G ;
Warsinke, A ;
Scheller, FW ;
Heller, A .
ANALYTICAL CHEMISTRY, 2000, 72 (13) :2963-2968