Brain MR Perfusion-Weighted Imaging With Alternate Ascending/Descending Directional Navigation

被引:22
作者
Park, Sung-Hong [1 ,2 ]
Duong, Timothy Q. [1 ,2 ]
机构
[1] Univ Texas Hlth Sci Ctr San Antonio, Res Imaging Inst, San Antonio, TX 78229 USA
[2] Univ Texas Hlth Sci Ctr San Antonio, Dept Radiol, San Antonio, TX 78229 USA
关键词
arterial spin labeling; perfusion-weighted imaging; perfusion directionality; white matter perfusion; balanced steady state free precession; alternate ascending/descending directional navigation; CEREBRAL-BLOOD-FLOW; RADIO-FREQUENCY; SPIN; MULTISLICE; RESONANCE; INVERSION; SSFP; WATER;
D O I
10.1002/mrm.22580
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
In this study, a new arterial spin labeling technique that requires no separate spin preparation pulse was developed. Sequential two-dimensional slices were acquired in ascending and descending orders by turns using balanced steady state free precession for pair-wise subtraction. Simulation studies showed this new technique, alternate ascending/descending directional navigation (ALADDIN), has high sensitivity to both slow- (1-10 cm/sec) and fast-moving (>10 cm/sec) blood because of the presence of multiple labeling planes proximal to imaging planes and sensitivity of balanced steady state free precession to initial magnetization differences. ALADDIN provided high-resolution multislice perfusion-weighted images in similar to 3 min. About 80-90% of signals in a slice were ascribed to spins saturated in the four prior slices. Three to five edge slices on each side of imaging group were affected by transient magnetization transfer effects and incomplete T(1) recovery between successive acquisitions. ALADDIN signals were dependent on many imaging parameters, implying room for improvement. Sagittal and coronal ALADDIN images demonstrated perfusion direction in gray matter regions was mostly from center to lateral, anterior, or posterior, whereas that in some white matter regions was reversed. ALADDIN is likely useful for many studies requiring perfusion-weighted imaging with short scan time, insensitiveness to arterial transit time, directional information, high resolution, and/or wide coverage. Magn Reson Med 65:1578-1591, 2011. (C)2010 Wiley-Liss, Inc.
引用
收藏
页码:1578 / 1591
页数:14
相关论文
共 35 条
[1]   Multisection cerebral blood flow MR imaging with continuous arterial spin labeling [J].
Alsop, DC ;
Detre, JA .
RADIOLOGY, 1998, 208 (02) :410-416
[2]   Reduced transit-time sensitivity in noninvasive magnetic resonance imaging of human cerebral blood flow [J].
Alsop, DC ;
Detre, JA .
JOURNAL OF CEREBRAL BLOOD FLOW AND METABOLISM, 1996, 16 (06) :1236-1249
[3]   Analysis of multiple-acquisition SSFP [J].
Bangerter, NK ;
Hargreaves, BA ;
Vasanawala, SS ;
Pauly, JM ;
Gold, GE ;
Nishimura, DG .
MAGNETIC RESONANCE IN MEDICINE, 2004, 51 (05) :1038-1047
[4]   FAIR-TrueFISP imaging of cerebral perfusion in areas of high magnetic susceptibility differences at 1.5 and 3 Tesla [J].
Boss, Andreas ;
Martirosian, Petros ;
Klose, Uwe ;
Naegele, Thomas ;
Claussen, Claus D. ;
Schick, Fritz .
JOURNAL OF MAGNETIC RESONANCE IMAGING, 2007, 25 (05) :924-931
[5]   Continuous Flow-Driven Inversion for Arterial Spin Labeling Using Pulsed Radio Frequency and Gradient Fields [J].
Dai, Weiying ;
Garcia, Dairon ;
de Bazelaire, Cedric ;
Alsop, David C. .
MAGNETIC RESONANCE IN MEDICINE, 2008, 60 (06) :1488-1497
[6]   PERFUSION IMAGING [J].
DETRE, JA ;
LEIGH, JS ;
WILLIAMS, DS ;
KORETSKY, AP .
MAGNETIC RESONANCE IN MEDICINE, 1992, 23 (01) :37-45
[7]   INCIDENTAL MAGNETIZATION TRANSFER CONTRAST IN STANDARD MULTISLICE IMAGING [J].
DIXON, WT ;
ENGELS, H ;
CASTILLO, M ;
SARDASHTI, M .
MAGNETIC RESONANCE IMAGING, 1990, 8 (04) :417-422
[8]   QUALITATIVE MAPPING OF CEREBRAL BLOOD-FLOW AND FUNCTIONAL LOCALIZATION WITH ECHO-PLANAR MR-IMAGING AND SIGNAL TARGETING WITH ALTERNATING RADIO-FREQUENCY [J].
EDELMAN, RR ;
SIEWERT, B ;
DARBY, DG ;
THANGARAJ, V ;
NOBRE, AC ;
MESULAM, MM ;
WARACH, S .
RADIOLOGY, 1994, 192 (02) :513-520
[9]   Nonlinear averaging reconstruction method for phase-cycle SSFP [J].
Elliott, Andrew M. ;
Bernstein, Matt A. ;
Ward, Heidi A. ;
Lane, John ;
Witte, Robert J. .
MAGNETIC RESONANCE IMAGING, 2007, 25 (03) :359-364
[10]   Perfusion Tensor Imaging [J].
Frank, Lawrence R. ;
Lu, Kun ;
Wong, Eric C. .
MAGNETIC RESONANCE IN MEDICINE, 2008, 60 (06) :1284-1291