Effects of residual target motion for image-tracked spine radiosurgery

被引:40
作者
Chuang, Cynthia [1 ]
Sahgal, Arjun [1 ]
Lee, Letitia [1 ]
Larson, David [1 ]
Huang, Kim [1 ]
Petti, Paula [1 ]
Verhey, Lynn [1 ]
Ma, Lijun [1 ]
机构
[1] Univ Calif San Francisco, Dept Radiat Oncol, San Francisco, CA 94143 USA
关键词
hypofractionated frameless radiotherapy; image guidance tracking; intrafraction target motion; robotic linear accelerator; stereotactic radiosurgery;
D O I
10.1118/1.2790587
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
A quality assurance method was developed to investigate the effects of residual target motion for hypofractionated spine radiosurgery. The residual target motion (target movement between successive image-guided corrections) was measured on-line via dual x-ray imagers for patients treated with CyberKnife (Accuray, Inc., Sunnyvale, CA), a robotic linear accelerator with intrafractional image-tracking capability. The six degree-of-freedom characteristics of the residual target motion were analyzed, the effects of such motion on patient treatment delivery were investigated by incorporating the probability distribution of the residual motion into the treatment planning dose calculations, and deviations of the doses from those originally planned were calculated. Measurements using a programmable motion phantom were also carried out and compared with the static treatment plan calculations. It was found that the residual target motions were patient specific and typically on the order of 2 mm. The measured dose distributions incorporating the residual target motion also exhibited 2.0 mm discrepancy at the prescription isodose level when compared with the static treatment plan calculations. For certain patients, residual errors introduced significant uncertainties (similar to 1 Gy) for the dose delivered to the spinal cord, especially at the high dose levels covering a small volume of the spinal cord (e.g., 0.1 cc). In such cases, stringent cord constraints and frequent monitoring of the target position should be implemented. (c) 2007 American Association of Physicists in Medicine.
引用
收藏
页码:4484 / 4490
页数:7
相关论文
共 11 条
  • [1] Image-guided robotic radiosurgery
    Adler, JR
    Murphy, MJ
    Chang, SD
    Hancock, SL
    [J]. NEUROSURGERY, 1999, 44 (06) : 1299 - 1306
  • [3] An analysis of the accuracy of the Cyberknife: A robotic nameless stereotactic radiosurgical system
    Chang, SD
    Main, W
    Martin, DP
    Gibbs, IC
    Heilbrun, MP
    [J]. NEUROSURGERY, 2003, 52 (01) : 140 - 146
  • [4] CyberKnife stereotactic radiosurgical treatment of spinal tumors for pain control and quality of life
    Degen, JW
    Gagnon, GJ
    Voyadzis, JM
    McRae, DA
    Lunsden, M
    Dieterich, S
    Molzahn, I
    Henderson, FC
    [J]. JOURNAL OF NEUROSURGERY-SPINE, 2005, 2 (05) : 540 - 549
  • [5] Cyberknife radiosurgery for benign intradural extramedullary spinal tumors
    Dodd, RL
    Ryu, MR
    Kamnerdsupaphon, P
    Gibbs, IC
    Chang, SD
    Adler, JR
    [J]. NEUROSURGERY, 2006, 58 (04) : 674 - 684
  • [6] Radiosurgery for spinal metastases - Clinical experience in 500 cases from a single institution
    Gerszten, Peter C.
    Burton, Steven A.
    Ozhasoglu, Cihat
    Welch, William C.
    [J]. SPINE, 2007, 32 (02) : 193 - 199
  • [7] Gamma knife radiosurgery for benign cavernous sinus tumors: Quantitative analysis of treatment outcomes
    Kuo, JS
    Chen, JCT
    Yu, C
    Zelman, V
    Giannotta, SL
    Petrovich, Z
    MacPherson, D
    Apuzzo, MLJ
    [J]. NEUROSURGERY, 2004, 54 (06) : 1385 - 1393
  • [8] An automatic six-degree-of-freedom image registration algorithm for image-guided frameless stereotaxic radiosurgery
    Murphy, MJ
    [J]. MEDICAL PHYSICS, 1997, 24 (06) : 857 - 866
  • [9] Patterns of patient movement during frameless image-guided radiosurgery
    Murphy, MJ
    Chang, SD
    Gibbs, IC
    Le, QT
    Hai, J
    Kim, D
    Martin, DP
    Adler, JR
    [J]. INTERNATIONAL JOURNAL OF RADIATION ONCOLOGY BIOLOGY PHYSICS, 2003, 55 (05): : 1400 - 1408
  • [10] Respiration tracking in radiosurgery
    Schweikard, A
    Shiomi, H
    Adler, J