Mechanical reliability of the AMS 700CXM inflatable penile prosthesis for the treatment of male erectile dysfunction

被引:23
作者
Choi, YD [1 ]
Choi, YJ
Kim, JH
Choi, HK
机构
[1] Yonsei Univ, Coll Med, Dept Urol, Seoul, South Korea
[2] Yonsei Univ, Coll Med, Inst Androl, Seoul, South Korea
关键词
penis; impotence; penile prosthesis;
D O I
10.1016/S0022-5347(05)66536-6
中图分类号
R5 [内科学]; R69 [泌尿科学(泌尿生殖系疾病)];
学科分类号
1002 ; 100201 ;
摘要
Purpose: The AMS 700CXM* inflatable penile prosthesis with medium controlled expansion cylinders was introduced in 1990 to provide a device with controlled expansion in girth and fitness for Asian men. We reviewed our experience with the AMS 700CXM prosthesis with particular regard to postoperative complications and reliability. Materials and Methods: From January 1991 to April 1999, 213 men with erectile dysfunction underwent primary penile prosthesis implantation with the AMS 700CXM. Functional status of the prosthesis was assessed retrospectively by reviewing the medical records and a telephone interview. Results: Patient age was 24 to 78 years and mean followup was 49 months. Of the 28 complications (10.3%) 8 (3%) were nonmechanical and 20 (7.3%) were mechanical. Nonmechanical complications included infection in 2 cases, erosion in 3, surgical complications in 2 and tube kinking in 1. Mechanical complications involved a cylinder tear in 11 cases (4%), pump failure in 4 (1.5%) and a reservoir tear in 2 (0.7%). Mechanical reliability of the device was 98.2% after 2, 95.7% after 3 and 90.4% after 5 years. Conclusions: The AMS 700CXM is suitable for Asian men and its mechanical reliability is excellent. However, more device modifications are needed to guarantee good reliability during long-term followup.
引用
收藏
页码:822 / 824
页数:3
相关论文
共 17 条
[1]   Fifteen years experience of penile prosthesis insertion [J].
Burns-Cox, N ;
Burston, A ;
Gingell, JC .
INTERNATIONAL JOURNAL OF IMPOTENCE RESEARCH, 1997, 9 (04) :211-216
[2]   Management of severe corporeal fibrosis with implantation of prosthesis via a transverse scrotal approach [J].
Carbone, DJ ;
Daitch, JA ;
Angermeier, KW ;
Lakin, MM ;
Montague, DK .
JOURNAL OF UROLOGY, 1998, 159 (01) :125-127
[3]  
Carson C. C., 1999, Journal of Urology, V161, P259
[4]   Long-term mechanical reliability of AMS 700 series inflatable penile prostheses: Comparison of CX/CXM and Ultrex cylinders [J].
Daitch, JA ;
Angermeier, KW ;
Lakin, MM ;
Ingleright, BJ ;
Montague, DK .
JOURNAL OF UROLOGY, 1997, 158 (04) :1400-1402
[5]   THE INFLATABLE PENILE PROSTHESIS - CLINICAL-EXPERIENCE WITH A NEW CONTROLLED EXPANSION CYLINDER [J].
FURLOW, WL ;
MOTLEY, RC .
JOURNAL OF UROLOGY, 1988, 139 (05) :945-946
[6]   5-YEAR FOLLOW-UP OF THE SCOTT INFLATABLE PENILE PROSTHESIS AND COMPARISON WITH SEMIRIGID PENILE PROSTHESIS [J].
KABALIN, JN ;
KESSLER, R .
JOURNAL OF UROLOGY, 1988, 140 (06) :1428-1430
[7]  
KNOLL LD, 1990, UROLOGY, V36, P502
[8]  
LEWIS RW, 1995, UROL CLIN N AM, V22, P847
[9]  
LEWIS RW, 1993, PROBL UROL, V7, P381
[10]   RELIABILITY OF AMS M700 INFLATABLE PENILE PROSTHESIS [J].
MALLOY, TR ;
WEIN, AJ ;
CARPINIELLO, VL .
UROLOGY, 1986, 28 (05) :385-387