Intravenous Administration Errors Intercepted by Smart Infusion Technology in an Adult Intensive Care Unit

被引:7
作者
Ibarra-Perez, Rebecca [1 ]
Puertolas-Balint, Fabiola [1 ]
Lozano-Cruz, Elizabeth [2 ]
Zamora-Gomez, Sergio E. [2 ]
Castro-Pastrana, Lucila, I [1 ]
机构
[1] Univ Americas Puebla, Dept Ciencias Quim Biol, Cholula, Mexico
[2] Hosp Juarez Mexico, Unidad Cuidados Intensivos Adultos, Ciudad De Mexico, Mexico
关键词
dose error reduction system; smart pump; intravenous medication safety; medication errors; intensive care; infusion therapy; ASHP NATIONAL-SURVEY; PUMPS; SAFETY; IMPACT; IMPLEMENTATION;
D O I
10.1097/PTS.0000000000000374
中图分类号
R19 [保健组织与事业(卫生事业管理)];
学科分类号
摘要
Objectives The aim of the study was to investigate the efficacy of intravenous (IV) smart pumps with drug libraries and dose error reduction system (DERS) to intercept programming errors entailing high risk for patients in an adult intensive care unit (ICU). Methods A 2-year retrospective study was conducted in the adult ICU of the Hospital Juarez de Mexico in Mexico City to evaluate the impact of IV smart pump/DERS (Hospira MedNet) technology implementation. We conducted a descriptive analysis of the reports generated by the system's software from April 2014 through May 2016. Our study focused on the upper hard limit alerts and used the systems' variance reports and IV Medication Harm Index methodology to determine the severity of the averted overdoses for medications with the highest number of edits. Results The system monitored 124,229 infusion programs and averted on 36,942 deviations of the preset safe limits. Upper hard limit alerts accounted for 26.4% of pump reprogramming events. One hundred sixty-six significant administration errors were intercepted and prevented, and IV Medication Harm Index analysis identified 83 of them as highest-risk averted overdoses with insulin accounting for 51.8% of those. The rate of compliance with the safety software during the study period was 69.8%. Conclusions Our study contributes additional evidence of the impact of IV smart pump/DERS technology. These pumps effectively intercepted severe infusion errors and significantly prevented adverse drug events related to dosing. Our results support the implementation of this technology in ICUs as a minimum safety standard and could help drive an IV infusion safety initiative in Mexico.
引用
收藏
页码:430 / 436
页数:7
相关论文
共 51 条
[21]  
Maddox RR, 2008, ADV PATIENT SAFETY N
[22]   Impact of implementing smart infusion pumps in a pediatric intensive care unit [J].
Manrique-Rodriguez, Silvia ;
Sanchez-Galindo, Amelia C. ;
Lopez-Herce, Jesus ;
Angel Calleja-Hernandez, Miguel ;
Martinez-Martinez, Fernando ;
Iglesias-Peinado, Irene ;
Carrillo-Alvarez, Angel ;
Sanjurjo Saez, Maria ;
Fernandez-Llamazares, Cecilia M. .
AMERICAN JOURNAL OF HEALTH-SYSTEM PHARMACY, 2013, 70 (21) :1897-1906
[23]  
Mansfield Jennifer, 2013, Hosp Pharm, V48, P942, DOI 10.1310/hpj4811-942
[24]   Clinical review:: Medication errors in critical care [J].
Moyen, Eric ;
Camire, Eric ;
Stelfox, Henry Thomas .
CRITICAL CARE, 2008, 12 (02)
[25]  
Murdoch Linda J, 2008, Br J Nurs, V17, P630
[26]   Benefits and Risks of Using Smart Pumps to Reduce Medication Error Rates: A Systematic Review [J].
Ohashi, Kumiko ;
Dalleur, Olivia ;
Dykes, Patricia C. ;
Bates, David W. .
DRUG SAFETY, 2014, 37 (12) :1011-1020
[27]  
Pa Patient Safety Authority, 2010, PA PATIENT SAF ADVIS, V7, P9
[28]   ASHP national survey of pharmacy practice in hospital settings: Dispensing and administration-2014 [J].
Pedersen, Craig A. ;
Schneider, Philip J. ;
Scheckelhoff, Douglas J. .
AMERICAN JOURNAL OF HEALTH-SYSTEM PHARMACY, 2015, 72 (13) :1119-1137
[29]   ASHP national survey of pharmacy practice in hospital settings: Dispensing and administration-2011 [J].
Pedersen, Craig A. ;
Schneider, Philip J. ;
Scheckelhoff, Douglas J. .
AMERICAN JOURNAL OF HEALTH-SYSTEM PHARMACY, 2012, 69 (09) :768-785
[30]  
Pennsylvania Patient Safety Authority, 2007, PA PSRS PATIENT SAF, V4, P139