Modular neck total hip arthroplasty - a perfect storm

被引:1
作者
Humadi, Ali [1 ]
Jamieson, Richard [1 ]
Chao, Tat [1 ]
Tacey, Mark [2 ,3 ]
Khoury, Elie [1 ]
机构
[1] Univ New S Wales, Albury Base Hosp, Sydney, NSW 2052, Australia
[2] Univ Melbourne, Northern Hosp, Northern Clin Res Ctr, Melbourne, Vic, Australia
[3] Univ Melbourne, Dept Med, Melbourne, Vic, Australia
关键词
Arthroplasty; Failure; Hip; Modular; Replacement; Revision; REPLACEMENT; STEM; CORROSION; REVISION; FAILURE;
D O I
10.5301/hipint.5000327
中图分类号
R826.8 [整形外科学]; R782.2 [口腔颌面部整形外科学]; R726.2 [小儿整形外科学]; R62 [整形外科学(修复外科学)];
学科分类号
摘要
Purpose: This study was aimed to highlight neck notching as a potential cause of failure in modular neck total hip arthroplasty. It aimed to identify both the combination of modular components that place patients at greatest risk of failure and the potential mechanism for failure. Methods: This study involved a retrospective review of 301 total hip arthroplasties (THA) using Kinectiv modular neck Technology (Zimmer, Warsaw, Indiana). The primary outcome was the presence of neck notching requiring revision. Patient records, operative notes and postoperative radiographs were reviewed. A statistical analysis was performed. Results: A total of 301 THA were performed on 290 patients. There were 7 failures (2.3%). All 7 failures involved a modular combination of an anteverted neck, extended offset and length code -8 (failure rate of 36.8%, p<0.001). Conclusions: We identified femoral neck notching as a potential cause of failure in modular neck THA. The combination of an anteverted neck, extended offset and length code -8 was associated with a high rate of neck notching using the Kinectiv modular neck Technology (Zimmer, Warsaw, Indiana). Based on our experience with this prosthesis we advise caution when using this particular combination of implants.
引用
收藏
页码:128 / 131
页数:4
相关论文
共 13 条
  • [1] Charnley total hip arthroplasty with cement - Minimum twenty-five-year follow-up
    Callaghan, JJ
    Albright, JC
    Goetz, DD
    Olejniczak, JP
    Johnston, RC
    [J]. JOURNAL OF BONE AND JOINT SURGERY-AMERICAN VOLUME, 2000, 82A (04) : 487 - 497
  • [2] Risk factors for revision for early dislocation in total hip arthroplasty
    Conroy, Jonathan L.
    Whitehouse, Sarah L.
    Graves, Stephen E.
    Pratt, Nicole L.
    Ryan, Philip
    Crawford, Ross W.
    [J]. JOURNAL OF ARTHROPLASTY, 2008, 23 (06) : 867 - 872
  • [3] Adverse Local Tissue Reaction Arising from Corrosion at the Femoral Neck-Body Junction in a Dual-Taper Stem with a Cobalt-Chromium Modular Neck
    Cooper, H. John
    Urban, Robert M.
    Wixson, Richard L.
    Meneghini, R. Michael
    Jacobs, Joshua J.
    [J]. JOURNAL OF BONE AND JOINT SURGERY-AMERICAN VOLUME, 2013, 95A (10) : 865 - 872
  • [4] Dunbar Michael J, 2010, Orthopedics, V33, P640, DOI 10.3928/01477447-20100722-30
  • [5] Modular titanium alloy neck adapter failures in hip replacement - failure mode analysis and influence of implant material
    Grupp, Thomas M.
    Weik, Thomas
    Bloemer, Wilhelm
    Knaebel, Hanns-Peter
    [J]. BMC MUSCULOSKELETAL DISORDERS, 2010, 11
  • [6] Impingement with total hip replacement
    Malik, Aamer
    Maheshwari, Aditya
    Dorr, Lawrence D.
    [J]. JOURNAL OF BONE AND JOINT SURGERY-AMERICAN VOLUME, 2007, 89A (08) : 1832 - 1842
  • [7] Meftah M, 2014, J BONE JOINT SURG AM, V96A, P481, DOI [10.2106/JBJS.M.00979, 10.2106/JBJS.L.01205]
  • [8] How Have Alternative Bearings and Modularity Affected Revision Rates in Total Hip Arthroplasty?
    Mihalko, William M.
    Wimmer, Markus A.
    Pacione, Carol A.
    Laurent, Michel P.
    Murphy, Robert F.
    Rider, Carson
    [J]. CLINICAL ORTHOPAEDICS AND RELATED RESEARCH, 2014, 472 (12) : 3747 - 3758
  • [9] Modular Hip Implant Fracture at the Stem-Sleeve Interface
    Parisi, Thomas
    Burroughs, Brian
    Kwon, Young-Min
    [J]. ORTHOPEDICS, 2015, 38 (03) : E234 - E239
  • [10] Midterm Results of a Femoral Stem With a Modular Neck Design: Clinical Outcomes and Metal Ion Analysis
    Silverton, Craig D.
    Jacobs, Joshua J.
    Devitt, Jeffrey W.
    Cooper, H. John
    [J]. JOURNAL OF ARTHROPLASTY, 2014, 29 (09) : 1768 - 1773