Design of optimal digital filter and digital signal processing for a CdZnTe high resolution gamma-ray system

被引:4
作者
Kim, B. J. [1 ,2 ]
Lee, K. B. [1 ,2 ]
Lee, J. M. [1 ,2 ]
Hwang, S. H. [1 ]
Heo, D. H. [1 ]
Han, K. H. [3 ]
机构
[1] Korea Res Inst Stand & Sci KRISS, Ionizing Radiat Ctr, Daejeon 34113, South Korea
[2] Univ Sci & Technol UST, Daejeon 34113, South Korea
[3] ULS Co Ltd, Daejeon 34186, South Korea
关键词
Digital signal processing; FPGA; Moving window deconvolution; CdZnTe detector;
D O I
10.1016/j.apradiso.2020.109171
中图分类号
O61 [无机化学];
学科分类号
070301 ; 081704 ;
摘要
We have developed an online digital signal processing system based on an FPGA. The system consists of pile-up rejection, baseline restorer, peak detection and pole-zero cancellation for evaluation of deposited energy in the detector. The shaping algorithm employed is a Moving Window Deconvolution (MWD) to shape digitized data into a trapezoidal form. For the purpose of verification, the developed digital system was tested for Cs-137 gamma rays. The entire system is programmed using the LabVIEW environment.
引用
收藏
页数:3
相关论文
共 11 条
[1]  
Bogovac M., 2007, SIGNAL PROCESS ELECT, V20, P17
[2]   DIGITAL PULSE PROCESSING IN HIGH-RESOLUTION, HIGH-THROUGHPUT GAMMA-RAY SPECTROSCOPY [J].
GEORGIEV, A ;
GAST, W .
IEEE TRANSACTIONS ON NUCLEAR SCIENCE, 1993, 40 (04) :770-779
[3]  
Guzik Z, 2013, NUKLEONIKA, V58, P333
[4]   Characteristics of depth-sensing coplanar grid CdZnTe detectors [J].
He, Z ;
Sturm, BW .
NUCLEAR INSTRUMENTS & METHODS IN PHYSICS RESEARCH SECTION A-ACCELERATORS SPECTROMETERS DETECTORS AND ASSOCIATED EQUIPMENT, 2005, 554 (1-3) :291-299
[5]   On nuclear spectrometry pulses digital shaping and processing [J].
Imperiale, C ;
Imperiale, A .
MEASUREMENT, 2001, 30 (01) :49-73
[6]   DIGITAL SYNTHESIS OF PULSE SHAPES IN REAL-TIME FOR HIGH-RESOLUTION RADIATION SPECTROSCOPY [J].
JORDANOV, VT ;
KNOLL, GF .
NUCLEAR INSTRUMENTS & METHODS IN PHYSICS RESEARCH SECTION A-ACCELERATORS SPECTROMETERS DETECTORS AND ASSOCIATED EQUIPMENT, 1994, 345 (02) :337-345
[7]   Development of a position-sensitive CZT detector with coplanar grid electrode [J].
Kim, B. J. ;
Lee, K. B. ;
Lee, J. M. ;
Park, T. S. ;
Hwang, S. H. .
JOURNAL OF RADIOANALYTICAL AND NUCLEAR CHEMISTRY, 2018, 316 (03) :1221-1225
[8]   Development of FPGA-based digital signal processing system for radiation spectroscopy [J].
Lee, Pil Soo ;
Lee, Chun Sik ;
Lee, Ju Hahn .
RADIATION MEASUREMENTS, 2013, 48 :12-17
[9]   UNIPOLAR CHARGE SENSING WITH COPLANAR ELECTRODES - APPLICATION TO SEMICONDUCTOR-DETECTORS [J].
LUKE, PN .
IEEE TRANSACTIONS ON NUCLEAR SCIENCE, 1995, 42 (04) :207-213