Development of Visual Field Screening Procedures: A Case Study of the Octopus Perimeter

被引:13
|
作者
Turpin, Andrew [1 ]
Myers, Jonathan S. [2 ]
McKendrick, Allison M. [3 ]
机构
[1] Univ Melbourne, Dept Comp & Informat Syst, Melbourne, Vic 3010, Australia
[2] Wills Eye Hosp & Res Inst, Philadelphia, PA USA
[3] Univ Melbourne, Dept Optometry & Vis Sci, Melbourne, Vic, Australia
来源
关键词
visual field; screening; glaucoma; FREQUENCY-DOUBLING TECHNOLOGY; AUTOMATED PERIMETRY; GLAUCOMA; VARIABILITY; SUPRATHRESHOLD; PATTERN; PERFORMANCE; LOCATIONS; DEFECTS; POINTS;
D O I
10.1167/tvst.5.3.3
中图分类号
R77 [眼科学];
学科分类号
100212 ;
摘要
Purpose: We develop a methodology for designing perimetric screening procedures, using Octopus perimeters as a case study. Methods: The process has three stages: analytically determining specificity and number of presentations required for different multisampling suprathreshold schemes at a single location of the visual field, ranking visual field locations by their positive predictive value (PPV) for glaucoma, and determining a pass/fail criteria for the test. For the case study the Octopus G-program visual field test pattern is used, and a dataset of 385 glaucoma and 86 normal patients. Results: Using a 1-of-3 sampling strategy at a level equal to the 95 percentile of normal observers gave the most robust specificity under the influences of false-negative responses using an average of 1.5 presentations per location. The PPV analysis gave 19 locations that completely classified our glaucomatous data. A further 9 points were added to screen for nonglaucomatous loss. The final stage found that insisting that 3 locations are missed for the screening to fail gave a simulated specificity and sensitivity of approximately 95% for unreliable responders. Conclusions: Our method gives a principled approach to choosing between the many parameters of a visual field screening procedure. We have developed a procedure for the Octopus that should terminate in less than 1 minute for normal observers with high specificity and sensitivity to glaucoma.
引用
收藏
页数:9
相关论文
共 50 条
  • [41] A computerized perimeter for assessing modality-specific visual field loss
    Calabro, Finnegan J.
    Vaina, Lucia M.
    2011 ANNUAL INTERNATIONAL CONFERENCE OF THE IEEE ENGINEERING IN MEDICINE AND BIOLOGY SOCIETY (EMBC), 2011, : 2025 - 2028
  • [42] Visual field home-monitoring with a tablet perimeter by glaucoma patients
    Prea, Selwyn Marc
    Kong, George Yu Xiang
    Vingrys, Algis J.
    INVESTIGATIVE OPHTHALMOLOGY & VISUAL SCIENCE, 2022, 63 (07)
  • [43] THE EFFECT OF REFRACTIVE BLUR ON THE VISUAL-FIELD USING THE RING PERIMETER
    HOUSE, PH
    DRANCE, SM
    SCHULZER, M
    WIJSMAN, K
    ACTA OPHTHALMOLOGICA, 1990, 68 (01): : 87 - 90
  • [44] Visual field examination using a video projector: comparison with Humphrey perimeter
    Brouzas, Dimitrios
    Tsapakis, Stylianos
    Nitoda, Eirini
    Moschos, Marilita M.
    CLINICAL OPHTHALMOLOGY, 2014, 8 : 523 - 528
  • [45] Validation of a novel LED-based chromatic visual field perimeter
    Balasubramanian, Giri
    Park, Jason
    Hyde, Robert
    McAnany, J.
    INVESTIGATIVE OPHTHALMOLOGY & VISUAL SCIENCE, 2023, 64 (08)
  • [46] REPEATED AUTOMATIC VISUAL-FIELD SCREENING IN A COHORT STUDY
    BENGTSSON, B
    ACTA OPHTHALMOLOGICA, 1985, 63 (02): : 242 - 244
  • [47] Prevent blindness America Visual Field Screening Study - Reply
    Sponsel, WE
    Ritch, R
    Stamper, R
    Higginbotham, EJ
    Anderson, DR
    Wilson, MR
    Zimmerman, TJ
    AMERICAN JOURNAL OF OPHTHALMOLOGY, 1996, 121 (05) : 590 - 591
  • [48] VISUAL FIELD SCREENING WITH AN OPHTHALMOSCOPE
    TSUJIMOTO, EY
    CALOROSO, E
    AMERICAN JOURNAL OF OPTOMETRY AND ARCHIVES OF AMERICAN ACADEMY OF OPTOMETRY, 1970, 47 (06): : 496 - +
  • [49] OCTOPUS PROGRAM SAPRO AND OCTOPUS PROGRAM-F - 2 NEW PRINCIPLES FOR THE ANALYSIS OF THE VISUAL-FIELD
    FANKHAUSER, F
    HABERLIN, H
    JENNI, A
    ALBRECHT VON GRAEFES ARCHIV FUR KLINISCHE UND EXPERIMENTELLE OPHTHALMOLOGIE, 1981, 216 (02): : 155 - 165
  • [50] Seismic assessment of buildings by rapid visual screening procedures
    Kapetana, P.
    Dritsos, S.
    EARTHQUAKE RESISTANT ENGINEERING STRUCTURES VI, 2007, 93 : 409 - +