Development of Visual Field Screening Procedures: A Case Study of the Octopus Perimeter

被引:13
|
作者
Turpin, Andrew [1 ]
Myers, Jonathan S. [2 ]
McKendrick, Allison M. [3 ]
机构
[1] Univ Melbourne, Dept Comp & Informat Syst, Melbourne, Vic 3010, Australia
[2] Wills Eye Hosp & Res Inst, Philadelphia, PA USA
[3] Univ Melbourne, Dept Optometry & Vis Sci, Melbourne, Vic, Australia
来源
关键词
visual field; screening; glaucoma; FREQUENCY-DOUBLING TECHNOLOGY; AUTOMATED PERIMETRY; GLAUCOMA; VARIABILITY; SUPRATHRESHOLD; PATTERN; PERFORMANCE; LOCATIONS; DEFECTS; POINTS;
D O I
10.1167/tvst.5.3.3
中图分类号
R77 [眼科学];
学科分类号
100212 ;
摘要
Purpose: We develop a methodology for designing perimetric screening procedures, using Octopus perimeters as a case study. Methods: The process has three stages: analytically determining specificity and number of presentations required for different multisampling suprathreshold schemes at a single location of the visual field, ranking visual field locations by their positive predictive value (PPV) for glaucoma, and determining a pass/fail criteria for the test. For the case study the Octopus G-program visual field test pattern is used, and a dataset of 385 glaucoma and 86 normal patients. Results: Using a 1-of-3 sampling strategy at a level equal to the 95 percentile of normal observers gave the most robust specificity under the influences of false-negative responses using an average of 1.5 presentations per location. The PPV analysis gave 19 locations that completely classified our glaucomatous data. A further 9 points were added to screen for nonglaucomatous loss. The final stage found that insisting that 3 locations are missed for the screening to fail gave a simulated specificity and sensitivity of approximately 95% for unreliable responders. Conclusions: Our method gives a principled approach to choosing between the many parameters of a visual field screening procedure. We have developed a procedure for the Octopus that should terminate in less than 1 minute for normal observers with high specificity and sensitivity to glaucoma.
引用
收藏
页数:9
相关论文
共 50 条
  • [21] Frequency doubling technology and Octopus visual field
    Iester, M
    Mennoud, A
    Schnyder, C
    ACTA OPHTHALMOLOGICA SCANDINAVICA, 2000, 78 : 23 - 24
  • [22] SUBCLINICAL VISUAL-FIELD DEFECTS IN MULTIPLE-SCLEROSIS - FINDINGS WITH THE COMPUTED PERIMETER OCTOPUS AND COMPARISON WITH VISUALLY EVOKED-POTENTIALS
    MEIENBERG, O
    FLAMMER, J
    LUDIN, HP
    ELECTROENCEPHALOGRAPHY AND CLINICAL NEUROPHYSIOLOGY, 1984, 57 (02): : P22 - P23
  • [23] Normal Values for the Full Visual Field, Corrected for Age- and Reaction Time, Using Semiautomated Kinetic Testing on the Octopus 900 Perimeter
    Grobbel, Julia
    Dietzsch, Janko
    Johnson, Chris A.
    Vonthein, Reinhard
    Stingl, Katarina
    Weleber, Richard G.
    Schiefer, Ulrich
    TRANSLATIONAL VISION SCIENCE & TECHNOLOGY, 2016, 5 (02):
  • [24] STATISTICAL-ANALYSIS OF VISUAL-FIELD WITH THE OCTOPUS 2000-R AUTOMATED PERIMETER AND PROGRAM DELTA-2000-R
    PRADINES, F
    DELBOSC, B
    ROYER, J
    JOURNAL FRANCAIS D OPHTALMOLOGIE, 1985, 8 (8-9): : 541 - 548
  • [25] A comparison of visual field testing with a new automated perimeter, the Compass visual field analyser, and the Humphrey visual field analyser
    Fenolland, J. R.
    Bonnel, S.
    Rosenberg, R.
    Sendon, D.
    Ghazal, W.
    Giraud, J. M.
    Renard, J. P.
    ACTA OPHTHALMOLOGICA, 2016, 94
  • [26] VARIATION IN VISUAL-FIELD MEASUREMENTS WITH AN AUTOMATED PERIMETER
    WILENSKY, JT
    JOONDEPH, BC
    AMERICAN JOURNAL OF OPHTHALMOLOGY, 1984, 97 (03) : 328 - 331
  • [27] A CLINICAL COMPARISON OF VISUAL-FIELD TESTING WITH A NEW AUTOMATED PERIMETER, THE HUMPHREY FIELD ANALYZER, AND THE GOLDMANN PERIMETER
    BECK, RW
    BERGSTROM, TJ
    LICHTER, PR
    OPHTHALMOLOGY, 1985, 92 (01) : 77 - 82
  • [28] Computer Perimeter for Screening of Field of Visions and Standardization of this Method
    Gotsiridze, Irina A.
    Gigilashvili, Giorgi T.
    Gurtskaia, Zviad T.
    Invia, Nikoloz S.
    2015 E-HEALTH AND BIOENGINEERING CONFERENCE (EHB), 2015,
  • [29] Occupant Preferences and Satisfaction with the Visual Environment in Perimeter Zone Offices: A Pilot Field Study
    Sadeghi, Seyed Amir
    Karava, Panagiota
    2016 ASHRAE ANNUAL CONFERENCE PAPERS, 2016,
  • [30] STUDY OF PROCEDURES USED FOR SCREENING ELEMENTARY SCHOOL CHILDREN FOR VISUAL DEFECTS - REFERRALS BY SCREENING PROCEDURES VERSUS OPHTHALMOLOGICAL FINDINGS
    CRANE, MM
    SCOBEE, RG
    FOOTE, FM
    GREEN, EL
    AMERICAN JOURNAL OF PUBLIC HEALTH, 1952, 42 (11) : 1430 - 1439