Determination of arsenic and cadmium in crude oil by direct sampling graphite furnace atomic absorption spectrometry

被引:26
作者
de Jesus, Alexandre [1 ]
Zmozinski, Ariane Vanessa [1 ]
Ferreira Damin, Isabel Cristina [2 ]
Silva, Marcia Messias [1 ,3 ]
Rodrigues Vale, Maria Goreti [1 ,3 ]
机构
[1] Univ Fed Rio Grande do Sul, Inst Quim, BR-91501970 Porto Alegre, RS, Brazil
[2] Fac Dom Bosco Porto Alegre, BR-90520280 Porto Alegre, RS, Brazil
[3] Univ Fed Bahia, Inst Natl Ciencia & Tecnol CNPq, INCT Energia & Ambiente, BR-40170115 Salvador, BA, Brazil
关键词
Arsenic; Cadmium; Direct sampling; Graphite furnace atomic absorption spectrometry; PETROLEUM; VANADIUM; METALS; COPPER; NICKEL; LEAD; STABILIZATION; NAPHTHA; FILTER;
D O I
10.1016/j.sab.2012.03.010
中图分类号
O433 [光谱学];
学科分类号
0703 ; 070302 ;
摘要
In this work, a direct sampling graphite furnace atomic absorption spectrometry method has been developed for the determination of arsenic and cadmium in crude oil samples. The samples were weighed directly on the solid sampling platforms and introduced into the graphite tube for analysis. The chemical modifier used for both analytes was a mixture of 0.1% Pd + 0.06% Mg + 0.06% Triton X-100. Pyrolysis and atomization curves were obtained for both analytes using standards and samples. Calibration curves with aqueous standards could be used for both analytes. The limits of detection obtained were 5.1 mu g kg(-1) for arsenic and 0.2 mu g kg(-1) for cadmium, calculated for the maximum amount of sample that can be analyzed (8 mg and 10 mg) for arsenic and cadmium, respectively. Relative standard deviations lower than 20% were obtained. For validation purposes, a calibration curve was constructed with the SRM 1634c and aqueous standards for arsenic and the results obtained for several crude oil samples were in agreement according to paired t-test. The result obtained for the determination of arsenic in the SRM against aqueous standards was also in agreement with the certificate value. As there is no crude oil or similar reference material available with a certified value for cadmium, a digestion in an open vessel under reflux using a "cold finger" was adopted for validation purposes. The use of paired t-test showed that the results obtained by direct sampling and digestion were in agreement at a 95% confidence level. Recovery tests were carried out with inorganic and organic standards and the results were between 88% and 109%. The proposed method is simple, fast and reliable, being appropriated for routine analysis. (C) 2012 Elsevier B.V. All rights reserved.
引用
收藏
页码:86 / 91
页数:6
相关论文
共 50 条
  • [21] Direct determination of phosphorus in biodiesel samples by graphite furnace atomic absorption spectrometry using a solid sampling accessory
    Lyra, Fernanda Henrique
    Weitzel Dias Carneiro, Maria Tereza
    Brandao, Geisamanda Pedrini
    Pessoa, Helen Moura
    Ribeiro de Castro, Eustaquio Vinicius
    [J]. JOURNAL OF ANALYTICAL ATOMIC SPECTROMETRY, 2009, 24 (09) : 1262 - 1266
  • [22] Direct solid sampling determination of lead in cheese varieties by high resolution continuum source graphite furnace atomic absorption spectrometry
    Tinas, Hande
    Ozbek, Nil
    Akman, Suleyman
    [J]. ANALYTICAL METHODS, 2017, 9 (45) : 6365 - 6370
  • [23] Arsenic and antimony determination in soils and sediments by graphite furnace atomic absorption spectrometry with slurry sampling
    LopezGarcia, I
    SanchezMerlos, M
    HernandezCordoba, M
    [J]. SPECTROCHIMICA ACTA PART B-ATOMIC SPECTROSCOPY, 1997, 52 (04) : 437 - 443
  • [24] Determination of As in Vegetable Oil and Biodiesel by Graphite Furnace Atomic Absorption Spectrometry
    Vieira, Mariana Antunes
    Castro de Oliveira, Ligia Claudia
    Goncalves, Rodrigo Araujo
    de Souza, Vanderlea
    de Campos, Reinaldo Calixto
    [J]. ENERGY & FUELS, 2009, 23 (12) : 5942 - 5946
  • [25] Direct determination of arsenic in petroleum derivatives by graphite furnace atomic absorption spectrometry: A comparison between filter and platform atomizers
    Becker, Emilene
    Rampazzo, Roger T.
    Dessuy, Morgana B.
    Vale, Maria Goreti R.
    da Silva, Marcia M.
    Welz, Bernhard
    Katskov, Dmitri A.
    [J]. SPECTROCHIMICA ACTA PART B-ATOMIC SPECTROSCOPY, 2011, 66 (05) : 345 - 351
  • [26] Determination of cadmium and lead in fresh meat by slurry sampling graphite furnace atomic absorption spectrometry
    Ferreira Damin, Isabel Cristina
    Zmozinski, Ariane Vanessa
    Borges, Aline Rocha
    Rodrigues Vale, Maria Goreti
    da Silva, Marcia Messias
    [J]. ANALYTICAL METHODS, 2011, 3 (06) : 1379 - 1385
  • [27] Determination of copper and mercury in phosphate fertilizers employing direct solid sampling analysis and high resolution continuum source graphite furnace atomic absorption spectrometry
    Souza, Sidnei de Oliveira
    Francois, Luciane Luiza
    Borges, Aline Rocha
    Rodrigues Vale, Maria Goreti
    Oliveira Araujo, Rennan Geovanny
    [J]. SPECTROCHIMICA ACTA PART B-ATOMIC SPECTROSCOPY, 2015, 114 : 58 - 64
  • [28] Determination of cadmium in rice by microwave digestion and graphite furnace atomic absorption spectrometry
    Amakawa, E
    Ogiwara, T
    Ohhashi, N
    Kamata, K
    Suzuki, S
    [J]. JOURNAL OF THE FOOD HYGIENIC SOCIETY OF JAPAN, 2000, 41 (01): : 66 - 69
  • [29] Direct determination of lead and cadmium in soil by slurry-sampling graphite furnace atomic absorption spectrometry using matrix modification technique
    Sun, HW
    Wen, XH
    Liang, SX
    [J]. SPECTROSCOPY AND SPECTRAL ANALYSIS, 2006, 26 (05) : 950 - 954
  • [30] Fast emulsion-based method for simultaneous determination of Co, Cu, Pb and Se in crude oil, gasoline and diesel by graphite furnace atomic absorption spectrometry
    Luz, Maciel S.
    Nascimento, Angerson N.
    Oliveira, Pedro V.