Determination of arsenic and cadmium in crude oil by direct sampling graphite furnace atomic absorption spectrometry

被引:26
|
作者
de Jesus, Alexandre [1 ]
Zmozinski, Ariane Vanessa [1 ]
Ferreira Damin, Isabel Cristina [2 ]
Silva, Marcia Messias [1 ,3 ]
Rodrigues Vale, Maria Goreti [1 ,3 ]
机构
[1] Univ Fed Rio Grande do Sul, Inst Quim, BR-91501970 Porto Alegre, RS, Brazil
[2] Fac Dom Bosco Porto Alegre, BR-90520280 Porto Alegre, RS, Brazil
[3] Univ Fed Bahia, Inst Natl Ciencia & Tecnol CNPq, INCT Energia & Ambiente, BR-40170115 Salvador, BA, Brazil
关键词
Arsenic; Cadmium; Direct sampling; Graphite furnace atomic absorption spectrometry; PETROLEUM; VANADIUM; METALS; COPPER; NICKEL; LEAD; STABILIZATION; NAPHTHA; FILTER;
D O I
10.1016/j.sab.2012.03.010
中图分类号
O433 [光谱学];
学科分类号
0703 ; 070302 ;
摘要
In this work, a direct sampling graphite furnace atomic absorption spectrometry method has been developed for the determination of arsenic and cadmium in crude oil samples. The samples were weighed directly on the solid sampling platforms and introduced into the graphite tube for analysis. The chemical modifier used for both analytes was a mixture of 0.1% Pd + 0.06% Mg + 0.06% Triton X-100. Pyrolysis and atomization curves were obtained for both analytes using standards and samples. Calibration curves with aqueous standards could be used for both analytes. The limits of detection obtained were 5.1 mu g kg(-1) for arsenic and 0.2 mu g kg(-1) for cadmium, calculated for the maximum amount of sample that can be analyzed (8 mg and 10 mg) for arsenic and cadmium, respectively. Relative standard deviations lower than 20% were obtained. For validation purposes, a calibration curve was constructed with the SRM 1634c and aqueous standards for arsenic and the results obtained for several crude oil samples were in agreement according to paired t-test. The result obtained for the determination of arsenic in the SRM against aqueous standards was also in agreement with the certificate value. As there is no crude oil or similar reference material available with a certified value for cadmium, a digestion in an open vessel under reflux using a "cold finger" was adopted for validation purposes. The use of paired t-test showed that the results obtained by direct sampling and digestion were in agreement at a 95% confidence level. Recovery tests were carried out with inorganic and organic standards and the results were between 88% and 109%. The proposed method is simple, fast and reliable, being appropriated for routine analysis. (C) 2012 Elsevier B.V. All rights reserved.
引用
收藏
页码:86 / 91
页数:6
相关论文
共 50 条
  • [1] Comparison of direct sampling and emulsion analysis using a filter furnace for the determination of lead in crude oil by graphite furnace atomic absorption spectrometry
    Damin, Isabel C. F.
    Dessuy, Morgana B.
    Castilhos, Tamara S.
    Silva, Marcia M.
    Vale, Maria Goreti R.
    Welz, Bernhard
    Katskov, Dmitri A.
    SPECTROCHIMICA ACTA PART B-ATOMIC SPECTROSCOPY, 2009, 64 (06) : 530 - 536
  • [2] Direct sampling graphite furnace atomic absorption spectrometry - feasibility of Na and K determination in desalted crude oil
    Seeger, Tassia S.
    Machado, Eduarda Q.
    Flores, Erico M. M.
    Mello, Paola A.
    Duarte, Fabio A.
    SPECTROCHIMICA ACTA PART B-ATOMIC SPECTROSCOPY, 2018, 141 : 28 - 33
  • [3] Magnesium and calcium determination in desalted crude oil by direct sampling graphite furnace atomic absorption spectrometry
    Seeger, Tassia S.
    Muller, Edson I.
    Mesko, Marcia F.
    Duarte, Fabio A.
    FUEL, 2019, 236 : 1483 - 1488
  • [4] Direct determination of nickel in petroleum by solid sampling-graphite furnace atomic absorption spectrometry
    Brandao, Geisamanda Pedrini
    de Campos, Reinaldo Calixto
    Ribeiro de Castro, Eustaquio Vinicius
    de Jesus, Honerio Coutinho
    ANALYTICAL AND BIOANALYTICAL CHEMISTRY, 2006, 386 (7-8) : 2249 - 2253
  • [5] Slurry sampling for direct determination of trace cadmium in amber by graphite furnace atomic absorption spectrometry
    Chen Shi-zhong
    Lu Deng-bo
    SPECTROSCOPY AND SPECTRAL ANALYSIS, 2006, 26 (10) : 1928 - 1930
  • [6] Simultaneous determination of Cr, Fe, Ni and V in crude oil by emulsion sampling graphite furnace atomic absorption spectrometry
    Luz, Maciel Santos
    Oliveira, Pedro Vitoriano
    ANALYTICAL METHODS, 2011, 3 (06) : 1280 - 1283
  • [7] Determination of cadmium, copper and lead in alumina based catalysts by direct solid sampling graphite furnace atomic absorption spectrometry
    Rodrigues, Luiz Frederico
    Paz de Mattos, Julio Cezar
    Dressler, Valderi Luiz
    Pozebon, Dirce
    de Moraes Flores, Erico Marlon
    SPECTROCHIMICA ACTA PART B-ATOMIC SPECTROSCOPY, 2007, 62 (09) : 933 - 938
  • [8] Determination of lead and cadmium in wines by graphite furnace atomic absorption spectrometry
    Kim, M
    FOOD ADDITIVES AND CONTAMINANTS, 2004, 21 (02): : 154 - 157
  • [9] Direct determination of trace arsenic in coal by slurry-sampling graphite-furnace atomic absorption spectrometry
    Chen, SZ
    SPECTROSCOPY AND SPECTRAL ANALYSIS, 2004, 24 (10) : 1267 - 1269
  • [10] Direct analysis of silica by means of solid sampling graphite furnace atomic absorption spectrometry
    Resano, M.
    Mozas, E.
    Crespo, C.
    Perez, J.
    Garcia-Ruiz, E.
    Belarra, M. A.
    SPECTROCHIMICA ACTA PART B-ATOMIC SPECTROSCOPY, 2012, 71-72 : 24 - 30