Experimental and theoretical relative response factors for a pulsed discharge krypton photoionization detector
被引:10
作者:
Wentworth, WE
Watanesk, S
Helias, N
Swatloski, R
Chen, ECM
Stearns, SD
机构:
[1] UNIV HOUSTON CLEAR LAKE, HOUSTON, TX 77058 USA
[2] VALCO INSTRUMENTS CO INC, HOUSTON, TX 77255 USA
关键词:
detection;
GC;
photoionization detection;
krypton;
benzene;
D O I:
10.1016/0021-9673(96)00473-6
中图分类号:
Q5 [生物化学];
学科分类号:
071010 ;
081704 ;
摘要:
The relative photoionization cross sections (RePIX) have been determined for a series of compounds normalized to benzene. The detectors are doped (5.01% Kr) Kr-PDPID and pure helium(He-PDPID) pulsed discharge helium photoionization detectors. The pure helium detector is used to reference the responses. The magnitudes for the Kr-PDPID varied from about 0.0 to 0.95. In order to understand the relative responses of these compounds in the Kr-PDPID, the SCF-MO energy levels were calculated for these compounds using HyperChem. Calculated vertical ionization potentials are compared with experimental adiabatic ionization potentials.