摘要
Purpose: To evaluate a prototype fast spin-echo (FSE) triple-echo Dixon (FTED) technique for T2-weighted spine imaging with and without fat suppression compared to conventional T2-weighted fast recovery (FR) FSE and short-tau inversion recovery (STIR) imaging. Materials and Methods: Sixty-one patients were referred for spine magnetic resonance imaging (MRI) including sagittal FTED (time 2:26), STIR (time 2:42), and T2 FRFSE (time 2:55). Two observers compared STIR and VIED water images and T2 FRFSE and FTED T2 images for overall image quality, fat suppression, anatomic sharpness, motion, cerebrospinal fluid (CSF) flow artifact, susceptibility, and disease depiction. Results: On FTED images water and fat separation was perfect in 58 (.95) patients. Compared to STIR, the FTED water images demonstrated less motion in 57 (.93) of 61 patients (P < 0.05), better anatomic sharpness in 51 (.84) and patients (P < 0.05), and less CSF flow artifact in 7 (.11) P < 0.05) patients. There was no difference in fat suppression or chemical shift artifact. T2 FRFSE and FTED T2 images showed equivalent motion, CSF flow, and chemical shift artifact. Lesion depiction was equivalent on FTED water and STIR images and FTED T2 and T2 FRFSE images. Conclusion: FTED efficiently provides both fat-suppressed and nonfat-suppressed T2-weighted spine images with excellent image quality, equal disease depiction, and 56% reduction in scan time compared to conventional STIR and T2 FRFSE.
引用
收藏
页码:390 / 400
页数:11
相关论文
共 37 条
  • [1] Fast Spin-Echo Triple-Echo Dixon: Initial Clinical Experience With A Novel Pulse Sequence for Fat-Suppressed T2-Weighted Abdominal MR Imaging
    Low, Russell N.
    Ma, Jingfei
    Panchal, Neeraj
    JOURNAL OF MAGNETIC RESONANCE IMAGING, 2009, 30 (03) : 569 - 577
  • [2] Improving Fat-Suppressed T2-Weighted Imaging of the Head and Neck with 2 Fast Spin-Echo Dixon Techniques: Initial Experiences
    Ma, J.
    Jackson, E. F.
    Kumar, A. J.
    Ginsberg, L. E.
    AMERICAN JOURNAL OF NEURORADIOLOGY, 2009, 30 (01) : 42 - 45
  • [3] T2-weighted magnetic resonance imaging of the liver: comparison of fat-suppressed GRASE with conventional spin echo, fat-suppressed turbo spin echo, and gradient echo at 1.0 T
    Karantanas, AH
    Papanikolaou, N
    ABDOMINAL IMAGING, 2001, 26 (02): : 139 - 145
  • [4] T1-weighted Fat-suppressed Imaging of the Pelvis with a Dual-Echo Dixon Technique: Initial Clinical Experience
    Beddy, Peter
    Rangarajan, R. Deepa
    Kataoka, Masako
    Moyle, Penelope
    Graves, Martin J.
    Sala, Evis
    RADIOLOGY, 2011, 258 (02) : 583 - 589
  • [5] Imaging of the degenerative spine using a sagittal T2-weighted DIXON turbo spin-echo sequence
    Sollmann, Nico
    Moench, Sebastian
    Riederer, Isabelle
    Zimmer, Claus
    Baum, Thomas
    Kirschke, Jan S.
    EUROPEAN JOURNAL OF RADIOLOGY, 2020, 131
  • [6] T2-WEIGHTED MR IMAGING OF FOCAL HEPATIC-LESIONS - COMPARISON OF VARIOUS RARE AND FAT-SUPPRESSED SPIN-ECHO SEQUENCES
    SEMELKA, RC
    SHOENUT, JP
    KROEKER, RM
    JMRI-JOURNAL OF MAGNETIC RESONANCE IMAGING, 1993, 3 (02): : 323 - 327
  • [7] Fat-suppressed T2-weighted MRI of the liver: Comparison of respiratory-triggered fast spin-echo, breath-hold single-shot fast spin-echo, and breath-hold fast-recovery fast spin-echo sequences
    Katayama, M
    Masui, T
    Kobayashi, S
    Ito, T
    Takahashi, M
    Sakahara, H
    Nozaki, A
    Kabasawa, H
    JOURNAL OF MAGNETIC RESONANCE IMAGING, 2001, 14 (04) : 439 - 449
  • [8] Fat-suppressed fast spin-echo mid-TE (TE[effective]=34) MR images: comparison with fast spin-echo T2-weighted images for the diagnosis of tears and anatomic variants of the glenoid labrum
    Tuite, MJ
    Shinners, TJ
    Hollister, MC
    Orwin, JF
    SKELETAL RADIOLOGY, 1999, 28 (12) : 685 - 690
  • [9] Fat-suppressed fast spin-echo mid-TE (TE[effective]=34) MR images: comparison with fast spin-echo T2-weighted images for the diagnosis of tears and anatomic variants of the glenoid labrum
    Michael J. Tuite
    Theodore J. Shinners
    Mark C. Hollister
    John F. Orwin
    Skeletal Radiology, 1999, 28 : 685 - 690
  • [10] Qualitative and Quantitative Analysis of a Spiral Gradient Echo Sequence for Contrast-Enhanced Fat-Suppressed T1-Weighted Spine Magnetic Resonance Imaging
    Sartoretti, Elisabeth
    Sartoretti, Thomas
    van Smoorenburg, Luuk
    Sartoretti-Schefer, Sabine
    Wyss, Michael
    Binkert, Christoph A.
    INVESTIGATIVE RADIOLOGY, 2021, 56 (08) : 517 - 524