Acoustophoresis in shallow microchannels

被引:35
作者
Koklu, Mehti [1 ]
Sabuncu, Ahmet Can [1 ]
Beskok, Ali [1 ]
机构
[1] Old Dominion Univ, Dept Mech & Aerosp Engn, Norfolk, VA 23529 USA
关键词
Acoustophoresis; Ultrasonic standing waves; Wall effects on colloidal transport; Sample concentration; MICROFLUIDIC CHANNELS; STANDING WAVES; PARTICLES; SEPARATION; FLOW; MANIPULATION; ULTRASOUND; FORCE; CELLS; FIELD;
D O I
10.1016/j.jcis.2010.08.029
中图分类号
O64 [物理化学(理论化学)、化学物理学];
学科分类号
070304 ; 081704 ;
摘要
Acoustophoretic (AP) motion of spherical polystyrene particles in a steady pressure driven flow is investigated in shallow microchannels, where the channel height is comparable to the particle diameter. Particle trajectories at different ultrasonic actuation amplitudes are extracted by a particle tracking algorithm. Depths of the particles are predicted using the streamwise particle speed that is due to the pressure driven flow. The particle depths are shown to be influenced by the actuation voltage. The particle migration along the channel height is explained using the second-order perturbation theory. The particle equation of motion is employed to extract the AP force. Wall effects are included in the analysis of both particle depth and force predictions. Differences as large as 20% in the AP force magnitude due to the wall corrections are reported. The AP force is also calculated using the theoretical force expression, and compared with the experimental results. The focal length, which is the necessary distance to effectively concentrate particles in a microchannel, is calculated using the analytical solution of the particle equation of motion. The calculated force and the focusing length agree well with the experimental results. The focal length is critical for the design of micro sample concentration devices. (C) 2010 Elsevier Inc. All rights reserved.
引用
收藏
页码:407 / 414
页数:8
相关论文
共 27 条
[1]   Measuring the local pressure amplitude in microchannel acoustophoresis [J].
Barnkob, Rune ;
Augustsson, Per ;
Laurell, Thomas ;
Bruus, Henrik .
LAB ON A CHIP, 2010, 10 (05) :563-570
[2]  
Bruus H., 2007, Theoretical microfluidics
[3]  
Coakley WT, 1997, TRENDS BIOTECHNOL, V15, P506, DOI 10.1016/S0167-7799(97)01122-0
[4]   Methods of digital video microscopy for colloidal studies [J].
Crocker, JC ;
Grier, DG .
JOURNAL OF COLLOID AND INTERFACE SCIENCE, 1996, 179 (01) :298-310
[5]   Continuous inertial focusing, ordering, and separation of particles in microchannels [J].
Di Carlo, Dino ;
Irimia, Daniel ;
Tompkins, Ronald G. ;
Toner, Mehmet .
PROCEEDINGS OF THE NATIONAL ACADEMY OF SCIENCES OF THE UNITED STATES OF AMERICA, 2007, 104 (48) :18892-18897
[6]   Noninvasive acoustic cell trapping in a microfluidic perfusion system for online bioassays [J].
Evander, Mikael ;
Johansson, Linda ;
Lilliehorn, Tobias ;
Piskur, Jure ;
Lindvall, Magnus ;
Johansson, Stefan ;
Almqvist, Monica ;
Laurell, Thomas ;
Nilsson, Johan .
ANALYTICAL CHEMISTRY, 2007, 79 (07) :2984-2991
[7]  
Gor'kov L., 1962, SOV PHYS DOKL, V6, P773, DOI DOI 10.1039/C004504G
[8]   Acoustic resonances in microfluidic chips:: full-image micro-PIV experiments and numerical simulations [J].
Hagsater, S. M. ;
Jensen, T. Glasdam ;
Bruus, H. ;
Kutter, J. P. .
LAB ON A CHIP, 2007, 7 (10) :1336-1344
[9]  
Happel J., 1983, Mechanics of Fluids and Transport Processes
[10]   Force field particle filter, combining ultrasound standing waves and laminar flow [J].
Hawkes, JJ ;
Coakley, WT .